Prime the pump rinsing system
Follow the instructions below to fill the pump rinsing system with rinsing solution.
Step
1
2
3
4
5
ÄKTA go Operating Instructions 29360951 AC
Action
Unscrew and remove the rinsing solution tube from the holder.
Fill the rinsing solution tube with 50 mL of 20% ethanol or aqueous buffer.
Screw the rinsing solution tube back into the holder.
Insert the inlet tubing into the solution in the rinsing solution tube.
Note:
Make sure that the inlet tubing reaches close to the bottom of the rinsing liquid
tube.
Connect a 25 to 30 mL syringe to the outlet tubing. Draw liquid slowly into
the syringe until the rinsing system tubing is filled.
4.2 Hardware installation
4.2.3 Prepare the pump rinsing system
4 Installation
49