I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® CONTENTS Foreword..............3 Device Connected to the Mains....13 Intended User Profile...........3 To stop treatment..........13 Intended Environment for Use......3 Caution..............14 Intended Use............3 Warning..............15...
FOREWORD INTENDED ENVIRONMENT FOR USE This guide is intended for users of Intelect RPW 2. ® The device is intended to be operated in a clinical It contains general information on operation, environment, it can be moved from room to room. precautionary practices, and maintenance.
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® PRECAUTIONARY INSTRUCTIONS The precautionary instructions found in this section and throughout this manual are indicated by specific symbols.
HOME SCREEN ICON DESCRIPTIONS DESCRIPTION OF DEVICE MARKINGS Refer to Instructional Manual Falcon Radial Pressure Waves and Quick Start Guide Booklet handpiece Vibration Therapy handpiece Warning, Caution, or Danger V-Actor Electrical Type B Equipment Shortcut Play SPS (Suggested Parameter Setup) Pause Custom Protocols ON/OFF...
Page 6
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® Relative Humidity Range Atmospheric Pressure Range Testing agency CE Mark of Conformity with notified body number Class 1 equipment Radio frequency equipment WEEE Directive conformity...
PAIN IN THE MUSCULOSKELETAL Radial and Ulnar humeral Epicondylitis SYSTEM Tennis elbow, inflammation of tendon attachments on cubital or radial part of elbow joint (humeral) Before a treatment with Radial Pressure Wave, a correct examination and diagnosis should be Achillodynia performed.
ADDITIONAL PRECAUTIONS ADVERSE EFFECTS When administering Radial Pressure Wave treatment, Side effects could occur after a treatment with Radial keep in mind the following: Pressure Wave therapy. The majority will appear after • Radial Pressure Wave treatment should 1-2 days. Do not repeat a treatment until the previous be applied with caution over bone where side effects have diminished.
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® PATIENT APPLIED PARTS Transmitters Material Applied part Type Type B Steel 1.4021 Type B Steel 1.4021 Type B Ceramic Volcera...
PRODUCT DESCRIPTION The Intelect RPW 2 is a compressed air-operated ® ballistic shock wave generator. The shock waves in the device are generated with a precision ballistic mechanism in the handpiece. A projectile is accelerated by compressed air. The motion and weight of the projectile produce kinetic energy.
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® COMPONENTS Cable Holders Throughout these instructions the terms “left” and “right” referring to the device sides are from the perspective of a user standing in front of the device.
IFU DOWNLOAD TO STOP TREATMENT 1. To download the IFU go to the Chattanooga website Press Play/Pause button to pause treatment then press www.chattanoogarehab.com Stop on touch screen. If device is on mains power 2. Complete the registration to be informed of...
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® CAUTION • This device should be operated at 10°C to +40°C and 30% to 75% Relative Humidity (no condensation). The device should be transported and stored at -20°C to +60°C and 10% to 90% Relative Humidity (no condensation).
CAUTION • Danger of injury due to pulse triggering when handpiece is open. • Disconnect the handpiece from the control device before starting any cleaning or maintenance work. • There are no user-serviceable parts inside the device. If a malfunction occurs, discontinue use immediately and contact your local DJO office via www.djoglobal.com for assistance and service information.
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® WARNING » Increase the separation between the equipment » Connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected »...
UNPACKING THE DEVICE » Sealing Set » Guide Tube The device is generally delivered with the packaging » Cleaning Brush material supplied by the manufacturer. As the device weighs approximately 32 kg (72 lbs), it must be 4. Retain the original packaging. It may prove useful unpacked by at least 2 persons.
Page 18
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® 6. The Gel Holder cup can be attached on the any side of the device 7.
DEVICE DESCRIPTION Handle Color Touch Screen Power Button Play / pause button V actor applicator support Rotary Falcon Hand Piece adjusment Knob Support Switch USB Connector ON/OFF Transmitter Tray Holder* * Device delivered with 2 transmitters. Other transmitters displayed are optional. * V-actor applicator is optional.
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® FALCON HANDPIECE SETUP 3. Screw the shaft into the handpiece until finger-tight. Using one hand, press the handpiece 1.
2. The first setup screen will be displayed after this allowing the user to set language and time in the device. 3. If after this step you see the screen below, switch of the device and follow instructions in point 4 below. If POWERING UP THE DEVICE you do not see this screen the device is ready to use.
I N T E L E C T R P W 2 Q U I C K S TA RT G U I D E ® OPERATING DATA AND RATINGS Electrical Class ..............CLASS I Electrical Type (Degree of Protection).......TYPE B Values displayed on the device are with a tolerance Regulatory Risk Class...IIa according to MDD 93/42/EEC of ±10%...
PATIENT PREPARATION Talk to your patient and ensure that they understand the following: Before applying Radial Pressure Wave therapy to the • Treatment should start at minimal Bar level patient, first prepare the patient’s skin. By properly • Bar level should increase slowly, manually or preparing the patient’s skin for therapy, it allows more automatically when using Comfort Mode energy to reach the targeted areas and reduces the...
Page 24
DJO France SAS Centre Européen de Frêt 3 rue de Bethar 64990 Mouguerre • France T: + 33 (0) 5 59 52 86 90 • F: + 33 (0) 5 59 52 86 91 DJOglobal.com DJO, LLC | 2900 Lake Vista Drive | Dallas, TX 75067 | U.S.A. www.DJOglobal.com...