Intelect® Neo Clinical Therapy System FOREWORD PRECAUTIONARY INSTRUCTIONS This manual is intended for users of Intelect® Neo Clinical The precautionary instructions found in this section and Therapy System. It contains general information on throughout this manual are indicated by speci c symbols. operation, precautionary practices, and maintenance.
SETUP Intelect® Neo Clinical Therapy System COMPONENTS Cart with Vacuum Module Throughout these instructions the terms “left” and “right” Cart installed referring to the machine sides are from the perspective of a user standing in front of the unit. The Intelect® Neo Clinical Therapy System allows installation of optional modality modules (except Vacuum module) by the user.
Page 6
SETUP Intelect® Neo Clinical Therapy System COMPONENTS CONTINUED Powercord Leadwires The available leadwires are shown below. If the user orders Stimulation Channel 1/2 module, the box will include the blue and green leadwires. Stimulation Channel 3/4 is the cranberry and orange leadwires. If both modules are ordered, the box contains all four colored leadwires.
SETUP Intelect® Neo Clinical Therapy System HEAD TO CART ASSEMBLY The optional Therapy System Cart, PN 70001, is designed 3. Place the Neo Head on the cart facing toward the for use with the Intelect® Neo Clinical Therapy System drawers. only and allows the user to easily transport the System from patient to patient within the clinic as well as store all necessary accessories, supplies, and applicators used for...
SETUP Intelect® Neo Clinical Therapy System NEO LEG TO CART ASSEMBLY/ADJUSTMENT The Neo Cart is shipped without the legs attached. To 2. There are two Cart height adjustments. Standard install or adjust the leg assemblies onto the Neo Cart, shown on the left and lowered, shown on the right. follow these steps: For initial installation, determine the desired height.
SETUP Intelect® Neo Clinical Therapy System MODULE INSTALLATION All modules (except vacuum module) are installed on 3. Insert a standard slotted screwdriver (not provided) from the left side (when facing the screen) of the Neo into the top slot, pressing down with slight pressure. head unit and are each installed in the same manner.
SETUP Intelect® Neo Clinical Therapy System MODULE INSTALLATION CONTINUED MODULE SPECIFIC INFORMATION Ultrasound Cable Insertion Shown below is the Ultrasound Cable Insertion location. 5. Secure the module with a screw provided at the bottom as shown (using stimulation channel 3/4 as example).
SETUP Intelect® Neo Clinical Therapy System THERAPY SYSTEM START UP Complete the following steps for initial setup of the 3. Select desired function on the Home Screen (shown Intelect® Neo Clinical Therapy System: below). 1. Plug the Power cord into the back of device. Plug the other end of the cord into an electrical outlet.
NOMENCLATURE Intelect® Neo Clinical Therapy System OPERATOR INTERFACE Front Controls The Intelect® Neo Clinical Therapy System Operator Interface contains all the functions and controls necessary for operator access to all operator utilities, modalities, and parameters for modi cation and system set up. 1.
NOMENCLATURE Intelect® Neo Clinical Therapy System MODULE SLOTS ULTRASOUND APPLICATOR 1. Laser 1. Applicator Head The component of the applicator that makes 2. Stimulation (1 & 2) / Stimulation (1&2) + sEMG contact with the patient during Ultrasound or Combination therapy. 3.
NOMENCLATURE Intelect® Neo Clinical Therapy System LASER APPLICATOR NOTE: When inserting the plugs for the Ultrasound and Laser modules, be sure to align the at side of the plug with the at side of the slot and push in gently. This is to avoid bending the pins in the plug.
NOMENCLATURE Intelect® Neo Clinical Therapy System PATIENT REMOTE/LASER INTERRUPT SWITCH PATIENT REMOTE/LASER INTERRUPT SWITCH INSTALLATION The Intelect® Neo Patient Remote/Laser Interrupt Switch To operate the Patient Remote/Laser Interrupt Switch, buttons are described below. By default, the remote is not plug the remote into the device on the Rear Access Panel assigned to any treatment.
NOMENCLATURE Intelect® Neo Clinical Therapy System 2. Press the Remote ON/OFF toggle icon to assign or (")'%'* unassign the remote to the selected treatment. The remote can be assigned to only one treatment • Disconnect the system from the power source before attempting at a time however the remote can be reassigned as any maintenance, installation, removal or replacement procedures needed.
NOMENCLATURE Intelect® Neo Clinical Therapy System GENERAL TERMINOLOGY The following are de nitions for the terminology used throughout this manual. Study these terms to become familiar with them for ease of system operation, and control functionality of the Intelect® Neo Clinical Therapy System.
NOMENCLATURE Intelect® Neo Clinical Therapy System SCREEN DESCRIPTION Setup: Indicates a treatment for the channel is Each screen contains the following areas: currently being setup but treatment has not yet Title bar begun Located at the top of each screen and lists the current screen and previous screens back to the Home screen.
PRECAUTIONARY INSTRUCTIONS Intelect® Neo Clinical Therapy System !"#$%&' !"#$%&' • Read, understand, and practice the precautionary and operating • Failure to use and maintain the Intelect® Neo System, its modules, and instructions. Know the limitations and hazards associated with using its accessories in accordance with the instructions outlined in this any electrical stimulation, laser device or ultrasound device.
Page 21
PRECAUTIONARY INSTRUCTIONS Intelect® Neo Clinical Therapy System (")'%'* (")'%'* • This device should be used only under the continued supervision of a • Clean applicators after each use, otherwise it can lead to cross physician or licensed practitioner. contamination and infection. •...
Page 22
Therefore, when performing any Improper application may result in patient injury. If any question operational or functional check, always wear Chattanooga arises as to the proper electrode size, consult a licensed practitioner laser protective eyewear.
INDICATIONS Intelect® Neo Clinical Therapy System ELECTROTHERAPY INDICATIONS Indications Contraindications The Intelect® Neo Clinical Therapy System o ers VMS, VMS The Intelect® Neo Clinical Therapy System should NOT be Burst, VMS-FR, Russian, TENS, TENS HAN, High Voltage used under the following conditions: Pulsed Current (HVPC), Interferential, and Premodulated •...
Page 24
INDICATIONS Intelect® Neo Clinical Therapy System ELECTROTHERAPY INDICATIONS CONTINUED Additional Precautions • Use caution for patients with suspected or diagnosed heart problems • Use caution for patients with suspected or diagnosed epilepsy • Use caution in the presence of the following: - When there is a tendency to hemorrhage following acute trauma or fracture - Following recent surgical procedures when...
INDICATIONS Intelect® Neo Clinical Therapy System SEMG & STIM INDICATIONS Additional Precautions Indications • Use caution for patients with suspected or diagnosed • Stroke rehab by muscle re-education heart problems • Relaxation of muscle spasms • Use caution for patients with suspected or diagnosed •...
INDICATIONS Intelect® Neo Clinical Therapy System ULTRASOUND INDICATIONS LASER INDICATIONS Indications Indications Application of therapeutic deep heat for the treatment of To provide topical heating for the following: selected sub-chronic and chronic medical conditions such • Increasing local blood circulation •...
Page 27
INDICATIONS Intelect® Neo Clinical Therapy System Additional Precautions Additional precaution should be used when the Laser is used on patients with the following conditions: • Over an area of the spinal cord following a laminectomy, i.e., when major covering tissues have been removed •...
SPECIFICATIONS Intelect® Neo Clinical Therapy System LASER APPLICATOR TECHNICAL SPECIFICATIONS For all single diode and cluster laser and LED applicators, the expected increase in the measured quantities after manufacture added to the values measured at the time of manufacture is 20%.
Page 31
SPECIFICATIONS Intelect® Neo Clinical Therapy System LASER APPLICATOR TECHNICAL SPECIFICATIONS CONTINUED Cluster Diode Applicators Nominal Ocular Output Divergence Divergence Spot Power Density Treatment Diode Diode Hazard Distance Applicator Power Size (W/cm Area(cm Type Speci cations (NOHD-in (mW0 (rad) (rad) (cm2) meters) 9 Diode Cluster Laser GaAIAs...
Page 32
SPECIFICATIONS Intelect® Neo Clinical Therapy System LASER APPLICATOR TECHNICAL SPECIFICATIONS CONTINUED Cluster Diode Applicators Nominal Ocular Output Power Divergence Divergance Spot Treatment Diode Diode Hazard Distance Applicator Power Density (W/ Size Area(cm Type Speci cations (NOHD-in (mW0 (rad) (rad) (cm2) meters) Six 670 nm (10 mW) LED 19 Diode Cluster Laser...
SPECIFICATIONS Intelect® Neo Clinical Therapy System LASER PROTECTIVE EYEWEAR SPECIFICATIONS The graph below illustrates optical density in relation to wavelength. Each unit is shipped with laser protective eyewear that is L3 rated and approved as well as EN207 compliant. Wavelength Useful Range Optical Density 5+.
SPECIFICATIONS Intelect® Neo Clinical Therapy System LASER LABELS This serial decal label is a xed at the rear of the system. The device is labeled with the date of manufacturer. The Laser applicator handle label includes important safety information. The LED applicator handle label includes important safety information.
WAVEFORMS Intelect® Neo Clinical Therapy System TENS- Asymmetrical Biphasic IFC (Interferential) Traditional (4 Pole) The Asymmetrical Biphasic waveform has a short pulse duration. It is capable Interferential Current is a medium frequency waveform. Current is distributed of strong stimulation of the nerve bers in the skin as well as of muscle tissue. through two channels (four electrodes).
Page 36
WAVEFORMS Intelect® Neo Clinical Therapy System VMS™ TENS- Monophasic Rectangular VMS is a symmetrical biphasic waveform with a 100 µsec interphase interval. The Monophasic Rectangular waveform is an interrupted unidirectional Because the pulse is relatively short, the waveform has a low skin load, current with a rectangular pulse shape.
Page 37
WAVEFORMS Intelect® Neo Clinical Therapy System Diadynamic Waveforms Russian The Diadynamic waveforms are recti ed alternating currents. The alternating Russian Current is a sinusoidal waveform, delivered in bursts or series of pulses. current is modi ed (recti ed) to allow the current to ow in one direction only. This method was claimed by its author (Kots) to produce maximal muscle strengthening e ects without signi cant discomfort to the patient.
Page 38
WAVEFORMS Intelect® Neo Clinical Therapy System MONOPHASIC: Monophasic Rectangular Pulsed GALVANIC: Interrupted The Monophasic Rectangular Pulsed waveform is an interrupted unidirectional Galvanic Current is a direct current owing in one direction only. The current current with a rectangular pulse shape. can be continuous or interrupted.
Page 39
WAVEFORMS Intelect® Neo Clinical Therapy System VMS™ FR SURGED: Monophasic Rectangular The VMS-FR version of the VMS waveform is a physiologically based channel The SURGED: Monophasic Rectangular Current is a series of rectangular, interaction in which one channel stimulates the agonist and the other the monophasic pulses.
Page 40
WAVEFORMS Intelect® Neo Clinical Therapy System Iontophoresis Iontophoresis is the introduction of ionizable drugs through intact skin by the administration of continuous, direct electrical current into the tissues of the body. The medicinal substance must be placed under the electrode pole with the same charge.
UTILITIES Intelect® Neo Clinical Therapy System UTILITIES AND OPTIONS The Utilities icon on the Home screen o ers users the opportunity to set the following preferences: 1. Print Screen uncoupled Select the Printer icon in the top right corner to capture screen shots of current screen.
Page 42
UTILITIES Intelect® Neo Clinical Therapy System UTILITIES AND OPTIONS CONTINUED 6. E-Stim Contact Quality 11. Import Protocols from USB flash drive Select the <E-Stim Contact Quality> icon to turn on Select the Import Protocols from USB flash drive or off the electrotherapy contact quality monitoring icon to import all protocols from a valid USB flash function.
PATIENT PREPARATION Intelect® Neo Clinical Therapy System ELECTRODE PLACEMENT ELECTROTHERAPY PATIENT PREPARATION DURA-STICK ELECTRODE INSTRUCTIONS • Examine the skin for any wounds and clean the skin • Apply the electrodes to the treatment area Connecting Lead Wires • Ensure the electrodes are applied securely to the skin 1.
PATIENT PREPARATION Intelect® Neo Clinical Therapy System IONTOPHORESIS PATIENT PREPARATION • Appropriate skin preparation is required for adequate conduction between the skin and the electrode. Clean the skin with warm water. Do not use alcohol • Choose the appropriate size electrode for the body part (the larger the body part the larger the electrode required) •...
PATIENT PREPARATION Intelect® Neo Clinical Therapy System VACUUM ELECTRODE PATIENT PREPARATION DJO Vacuum Electrodes are designed speci cally for use NOTE: Place the Vacuum plugs on all vacuum with the Intelect® Neo Vacuum Module for optimum channels not being used to prevent vacuum leakage e ciency and performance of the Intelect®...
Page 46
PATIENT PREPARATION Intelect® Neo Clinical Therapy System VACUUM ELECTRODE PATIENT PREPARATION CONTINUED 7. Select Vacuum Mode: In continuous mode, the vacuum unit will maintain the desired vacuum pressure. 8. With the Vacuum Electrodes placed face down on a clean, firm surface, increase the intensity of the Vacuum with the Vacuum Intensity Knob until the electrodes are secure on the surface.
PATIENT PREPARATION Intelect® Neo Clinical Therapy System SEMG & STIM PATIENT PREPARATION Install DURA-STICK Electrodes 2. Select “Customize” . 1. Connect a DURA-STICK® 1.25 in (3 cm) disposable 3. Select “Electrode Placement” to view electrode electrode to each lead. These electrodes are designed placement according to body area.
PATIENT PREPARATION Intelect® Neo Clinical Therapy System SEMG & STIM PATIENT PREPARATION CONTINUED 8. Thoroughly cleanse the skin treating area. 3. Review the specific Electrode Placement graphic for positioning of the Reference (green lead) electrode. NOTE: Thorough and proper cleaning of the NOTE: The electrodes may be placed for specific, treatment area to remove any topical medication general and quasi-specific biofeedback muscle or...
PATIENT PREPARATION Intelect® Neo Clinical Therapy System ULTRASOUND PATIENT PREPARATION 1. Examine the skin for any wounds and clean the skin. 2. View the Applicator recommendation in the treatment. 3. Review screen for Ultrasound (as a reference point only) prior to administering treatment. NOTE: Applicators are available in the sizes shown below: Small...
OPERATION Intelect® Neo Clinical Therapy System HOME SCREEN The Intelect® Neo Clinical Therapy System Home screen provides access to all of the system modalities and functions. The Home screen has the following information: 1) Utilities Modality Icons: 2) Electrotherapy 3) Ultrasound 4) Combo 5) sEMG 6) Laser...
OPERATION Intelect® Neo Clinical Therapy System TREATMENT SCREENS The Intelect® Neo Clinical Therapy System Treatment screens for Electrotherapy and Ultrasound, include the following information: 1. Modality Description Icon 8. Therapy Information Window Press the Modality Description Icon to view the text View selected Therapy information such as Waveform, explaining the rationale for the modality associated Electrode Size, CC/CV, Carrier Frequency, Cycle Time,...
OPERATION Intelect® Neo Clinical Therapy System CPS CLINICAL PROTOCOL SETUP The Intelect® Neo Clinical Therapy System has a Clinical 9. Overwrite a previously saved protocol by using the Up Protocol Setup (CPS) icon that is a series of protocol and Down arrows or Previous Page/Next Page icons to presets where the Body Area, Clinical Indication, select the protocol to overwrite.
Page 53
OPERATION Intelect® Neo Clinical Therapy System ELECTROTHERAPY OPERATION CONTINUED - Edit pain scale information by pressing the Edit • To view information explaining the waveform, select Pain Scale icon. Press the Up and Down arrows the Modality Description icon. Press the Up and Down until pain is properly gauged.
OPERATION Intelect® Neo Clinical Therapy System VACUUM OPERATION Vacuum Mode NOTE: Combination Therapy, sEMG and sEMG & Electrical The Intelect® Neo Vacuum allows for continuous and Stimulation Therapy cannot be administered using the pulsed modes of operation, as shown on the Treatment Vacuum Electrode Module.
OPERATION Intelect® Neo Clinical Therapy System SEQUENCING The Intelect® Neo Clinical Therapy System o ers sequencing for special electrotherapy waveform treatment • Save therapy information to USB Flash drive by purposes and stores these protocols in the system inserting a USB Flash drive into the device and memory for recall and use.
OPERATION Intelect® Neo Clinical Therapy System STRENGTH/DURATION S/D CURVE The Strength/Duration (S/D) Curve is an electrodiagnostic method to measure whether a muscle is showing signs of denervation or to assess the re-innervation process. Complete the following steps to create and access customized sequences: 1.
OPERATION Intelect® Neo Clinical Therapy System IONTOPHORESIS OPERATION - Edit pain scale information by pressing the Edit 1. Prepare patient and system for Iontophoresis. Pain Scale icon. Press the Up and Down arrows Refer to the PATIENT PREPARATION section on page until pain is properly gauged.
Stop button to NOTE: Use only Intelect® Neo Ultrasound Applicators. terminate the treatment. Previous models of Chattanooga Ultrasound Applicators will not work with the Intelect® Neo 4. When treatment has completed, the Treatment Clinical Therapy System.
OPERATION Intelect® Neo Clinical Therapy System COMBINATION OPERATION The Intelect® Neo Clinical Therapy System Combination 6. The Treatment Review Screen below will appear: modality allows the user to select and use ultrasound therapy in combination with electrical muscle stimulation. Combination therapy utilizes the Ultrasound modality in conjunction with High Voltage Pulsed Current (HVPC), IFC (4p), IFC Premodulated (2p), Asymmetrical Biphasic, Symmetrical Biphasic, or VMS™...
Page 60
OPERATION Intelect® Neo Clinical Therapy System Ultrasound Customization Screen • Make the desired changes and press the Back Arrow icon to return to the previous screen, the Home icon to return to the home screen, or the Save icon to save the customized settings.
OPERATION Intelect® Neo Clinical Therapy System SEMG OPERATION The Neo Therapy sEMG + Stim modality utilizes sEMG biofeedback activity coupled with triggered electrical muscle stimulation using selected electrotherapy waveforms for the maximum bene t in muscle retraining. The following option are available: sEMG (Channel 1), sEMG (Channel 2), sEMG (Channels 1 &...
Page 62
OPERATION Intelect® Neo Clinical Therapy System SEMG OPERATION CONTINUED • Use the Intensity dial to set therapy intensity: - Clockwise - Increases Intensity - Counterclockwise - Decreases Intensity 8. Make the desired changes and press the Back Arrow icon to return to the previous screen and begin treatment, the Home icon to return to the home screen, or the Save icon to save the customized settings.
OPERATION Intelect® Neo Clinical Therapy System LASER OPERATION Before applying Laser therapy to the patient, you must • To view information explaining the modality, select rst prepare the patient’s skin as described in the Patient the Modality Description icon. Press the Up and Down Preparation section, page 46.
Page 64
OPERATION Intelect® Neo Clinical Therapy System LASER OPERATION CONTINUED 8. When treatment has completed, the Treatment Summary screen will appear with the following options: • Save therapy information to USB Flash drive by inserting a USB Flash drive into the device and pressing the Save to USB Flash Drive icon.
DATA Intelect® Neo Clinical Therapy System SAVING TO USB FLASH DRIVE/PATIENT DATA CUSTOM PROTOCOLS Patient treatment data can be saved to the USB ash A new custom protocol may be saved at either the drive for viewing/printing on a PC as well as for retrieving Treatment Edit or Treatment Complete screen.
Page 66
DATA Intelect® Neo Clinical Therapy System CUSTOM PROTOCOLS CONTINUED 3. After saving the protocol, you will return to the Complete the following steps to delete a previously saved Treatment Summary screen. Complete one of the protocol: following actions listed on the Treatment Summary screen: 1.
Page 67
DATA Intelect® Neo Clinical Therapy System CUSTOM PROTOCOLS CONTINUED Complete the following steps to assign a Home screen 2. The unit will display a text box asking, “Are you sure shortcut for a customized protocol: you want to de-assign protocol ______ shortcut?” 1.
DATA Intelect® Neo Clinical Therapy System ANATOMICAL LIBRARY The Intelect® Neo Clinical Therapy System contains a 4. View the selected image and if desired, print the unique Anatomical Library designed to aid the operator selected image by pressing the printer icon: in visually understanding and locating speci c muscle groups and commonly found problems associated with pathological conditions as well as providing an...
TROUBLESHOOTING Intelect® Neo Clinical Therapy System TROUBLESHOOTING CODES The Intelect® Neo Clinical Therapy System incorporates information, warning, and error messages to inform the user of problems or potential problems with the system, modality, or accessories. The numbering sequence is: 100-level messages are general use information messages, 200-level messages are warnings, and 300-level messages are errors.
Page 70
TROUBLESHOOTING Intelect® Neo Clinical Therapy System TROUBLESHOOTING CODES CONTINUED Code Type Message Probable Cause Possible Remedies Number 1. If a stim module is installed in the unit, ensure module is securely inserted into the Attempting to select an electrotherapy treatment but no Stim unit Information module is installed in the unit...
Page 71
TROUBLESHOOTING Intelect® Neo Clinical Therapy System TROUBLESHOOTING CODES CONTINUED Code Type Message Probable Cause Possible Remedies Number Warning Vacuum reservoir is full Cannot start a vacuum treatment when the vacuum reservoir is full. Empty the vacuum reservoir prior to selecting a vacuum treatment. Any 200-level Warning Immediately stop all use of the system and contact the dealer or DJO for service.
ACCESSORIES Intelect® Neo Clinical Therapy System REPLACEMENT ACCESSORIES The following provides users of the Intelect® Neo Clinical Therapy System the necessary information to order replacement accessories used with the system. This list of replacement accessories is designed for use with the Intelect® Neo Clinical Therapy System.
Page 73
ACCESSORIES Intelect® Neo Clinical Therapy System REPLACEMENT ACCESSORIES CONTINUED GENERAL ACCESSORIES Model Number Description 70000 Electrotherapy Module Channels 1/2 70001 Neo Cart 70002 Ultrasound Module 70003 Electrotherapy Module Channels 3/4 70004 Electrotherapy Module Channels 1/2 + sEMG 70005 Laser Module 70006 Vacuum Module 70008...
Page 74
ACCESSORIES Intelect® Neo Clinical Therapy System REPLACEMENT ACCESSORIES CONTINUED IONTOPHORESIS ELECTRODES DUPEL B.L.U.E. Model Number Description 199332-001 Small (1.5cc ll) 199334-001 Butter y (2.0cc ll) 199335-001 Standard (2.5cc ll) 199336-001 Large (4.0cc ll) 1336 Black Snap Adapters (2 pin to snap adapters) ULTRASOUND APPLICATORS AND GEL Model Number Description...
MAINTENANCE Intelect® Neo Clinical Therapy System FUSE INFORMATION 3.15 A 250V AC INSTRUCTION FOR SOFTWARE UPGRADE 1. Obtain a USB flash-drive with upgrade file in root directory. 2. Power on Unit with flash-drive installed in USB port. Allow initialization to complete. 3.
MAINTENANCE Intelect® Neo Clinical Therapy System CLEANING THE INTELECT® NEO CLINICAL THERAPY SYSTEM Reservoir Draining With the system disconnected from the power source, • When draining the reservoir, wear surgical type gloves. clean the system with a clean, lint-free cloth moistened To drain the reservoir cup, rotate clockwise direction, with water and mild antibacterial soap.
WARRANTY Intelect® Neo Clinical Therapy System WARRANTY REPAIR/OUT OF WARRANTY REPAIR SERVICE When the Intelect® Neo Clinical Therapy System or any of the accessory modules require service, contact the selling dealer or DJO Service Department. All Therapy System and accessory modules returned to the factory for service must include the following: 1.
WARRANTY Intelect® Neo Clinical Therapy System WARRANTY DJO, LLC (“Company”) warrants that the Intelect® Neo Clinical Therapy System, COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR Channel 1/2 Electrotherapy Module, Channel 1/2 Electrotherapy+EMG Module, CONSEQUENTIAL DAMAGES. Channel 3/4 Electrotherapy Module, Laser Module, Ultrasound Module, and Some states do not allow the exclusion or limitation of incidental or Vacuum Module (“Products”) are free of defects in material and workmanship.
APPENDIX Intelect® Neo Clinical Therapy System OVERVIEW OF LASER THERAPY TREATMENT TIPS Contact Laser therapy results in energy absorbed into the patient’s tissue from light, triggering biological changes at a To obtain the most e ective results, the applicator should cellular level, resulting in the following: be in contact with the patient’s skin.
APPENDIX Intelect® Neo Clinical Therapy System COMMON TERMS Applicator – Hand held assembly that delivers Laser Treatment Time – Measured in seconds, it is the energy and includes laser head, diode, and related suggested time per laser point that therapy is given electronics Wavelength –...
The Intelect Neo Clinical Therapy System is intended for use in the electromagnetic environment speci ed below. The customer or the user of the Intelect Neo Clinical Therapy System should assure that it is used in such an environment Emissions Tests...
Page 82
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the Intelect Neo Clinical Therapy System is used exceeds the applicable RF compliance level above, theIntelect Neo Clinical Therapy System should be observed to verify normal operation.
Page 83
The Intelect Neo Clinical Therapy System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Intelect Neo Clinical Therapy System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Intelect Neo Clinical Therapy System as recommended below, according to the maximum output power of the communications equipment.