Vertical installation in a drop line ......17 4.6.3 Horizontal installation in case of measurement of liquids ..............17 4.6.4 Horizontal installation in case of measurement of gases ..............17 CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 3...
Page 4
(parameterization) ..........48 9.2.3 Selecting and changing parameters ....49 Overview of parameters on the configuration level50 Appendix ............... 54 10.1 Approvals and certifications ....... 54 10.2 Installation diagram 3KXF002126G0009 .... 59 4 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
The operators must strictly observe the applicable national regulations with regards to installation, function tests, repairs, and maintenance of electrical products. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 5...
18 Maximum power consumption | 19 Serial number of sensor | "IMPORTANT (NOTE)" does not indicate a 20 CE mark | 21 Ex approval cFMus | 22 Ex approval ATEX / IECEx dangerous or harmful situation. 6 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
(3.28 ft) must be maintained. For installation on steel parts (e.g., steel brackets), a minimum spacing of 100 mm (4") must be maintained. (These values have been calculated on the basis of IEC 801-2 and IEC TC77B.) CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 7...
If it is not possible to dispose of old equipment properly, ABB Service can accept and dispose of returns for a fee. 1.14.2 RoHS Directive 2002/95/EC...
(-58 … 392 °F) (-58 … 320 °F) (-58 … 392 °F) 1) Stated measuring accuracy in % of rate (% o. r.) 2) Measuring accuracy following on-site calibration under operating conditions CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 9...
Page 10
Local display / totalization Field optimization for flow and density Degree of protection acc. to EN 60529 Integral mount design: IP 65/IP 67, NEMA 4X Remote mount design: IP 67, NEMA 4X 10 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
— Zone 1, 21 — Zone 0 IMPORTANT (NOTE) Detials can be found in chapter „Ex relevant specifications acc. to ATEX / IECEx / NEPSI“ or in the respective certificate. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 11...
— Class I Div. 1 — Zone 1, 21 — Zone 0, 20 IMPORTANT (NOTE) Details can be found in chapter „Ex relevant specifications acc. to cFMus“ or in the respective certificate. 12 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
— Do not expose the transmitter and sensor to direct sunlight. Provide appropriate sun protection as necessary. — When installing the transmitter in a control cabinet, make sure adequate cooling is provided. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 13...
4.3.2 Transmitter in remote mount design (option R1 or R2) IP 65 / 67, NEMA 4X Fig. 3: Dimensions in mm (inch) Change from one to tw o c olumns 14 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Check that the gasket is properly seated when closing the housing cover. DANGER – Risk of explosion! Violation of hazardous area protection. Do not disconnect the transmitter from the sensor. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 15...
— Check that the flowmeter sensor does not come into contact with other objects. Do not attach the flowmeter sensor to the enclosure. Fig. 6: Vertical installation, self-draining 16 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Installing the flowmeter sensor at the lowest point of a pipeline leads to increased inaccuracies due to the accumulation of fluid or the formation of condensate in the meter tube. 18 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
EHEDG- compliant parts. Note the information in the latest version of the following document: EHEDG Position Paper: "Hygienic process connections to use with hygienic components and equipment". CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 19...
The power supply is connected to terminal L (phase), N (neutral), or 1+, 2-, and PE, as stated on the name plate. Connect the transmitter and flowmeter sensor to functional ground. 20 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
1 Cover for connection area | 2 Pin assignment | 3 Cable entry points | 4 Terminals for power supply | 5 Terminals for signal inputs and signal outputs | 6 Terminal cover CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 21...
Before closing the housing cover, check the gasket (o-ring) for any damage and replace if necessary. Check that the gasket is properly seated when closing the housing cover. 22 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
250 … 560 Ω (in hazardous area: maximum 300 Ω) Cable Design Two-wire cable AWG 24, twisted Maximum length 1500 m (4921 ft) See the interface description for detailed information. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 23...
— Input "Off": 0 V ≤ UKL ≤ 2 V Equipotential bonding "PA" When the FCT300 transmitter is connected to the FCB3xx / FCH3xx flowmeter sensor, the transmitter must also be connected to "PA". 24 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
"B" Input for external totalizer reset or external output zero return | I Internal | E External Pulse output Fig. 23: Active / passive pulse output "A" Active | "B" Passive (optocoupler) | I Internal | E External CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 25...
The following flowmeter sensor / transmitter combinations are permitted: — FCB330, FCH330 flowmeter sensor with FCT330 transmitter — FCB350, FCH350 flowmeter sensor with FCT350 transmitter Change from one to tw o c olumns CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 27...
STEP key to call up "System Zero adj. Function automatic?" device. and select ENTER to start the adjustment. You can choose between slow or fast adjustment. Slow adjustment generally provides a more accurate zero point. 28 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Pulse output 51 / 52 active (not for hazardous area design) IMPORTANT (NOTE) Please note that in case of Zone 1 or Div. 1 approved meters, this jumper will be in position "B" (active) although the pulse output is passive! CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 29...
(PA), even in the event of a fault. — In case of Dust-Ex application, the maximum surface temperature is 85 °C (185 °F). — The process temperature of connected piping can exceed 85 °C (185 °F). 30 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
10 times higher figures (3.0 kPa (300 mmWS)) If the flowmeter is to be insulated, follow the instructions in the may be used. "Mounting / Installation positions / Installation with option TE1 "extended tower length"" section. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 31...
Important (NOTE) For more details on hazardous areas, protection type and instrument Model No see installation diagram FCB 3KXF002126G0009 (see chapter „Appendix“). Change from one to tw o c olumns 32 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
10 mm (0.39 inch) protruding from the end (if necessary, shorten the protective tubing). 4. Press on the ring-type terminal (2) supplied. 5. Connect to the grounding connection (3). CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 33...
Seal all conduits within 18 inches (457.20 mm). Limits Flange or tube Material All materials of the existing model The flow meters from ABB are designed for the world wide number industrial market and are suitable to measure among others Sizes DN 20 ...
All inputs and outputs are electrically isolated from one other and from the power supply. Only current outputs 1 and 2 in zone 1 design are not electrically isolated from one another. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 35...
The switching output and the pulse output (terminals 41 / 42 and 51 / 52) can be wired internally as a NAMUR contact for the purpose of connecting a NAMUR amplifier. Change from two to one c olumn 36 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
-40 ... 60 °C (-40 ... 140 °F) (only for integral mount design devices) amb, optional -50 ... 200 °C (-58 ... 392 °F) medium Protection class IP 65, IP 67, and NEMA 4X CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 37...
Specific marking according to ATEX, IECEx and NEPSI applies depending on the design of the flowmeter sensor (integral or remote mount design). IMPORTANT (NOTE) ABB reserves the right to make changes to the Ex-marking. The exact marking can be found at the name plate of the meter. Modell FCx3xx-A2A… or FCx3xx-S2A… (remote mount design in Zone 2) Approval...
IMPORTANT (NOTE) ABB reserves the right to make changes to the Ex-marking. The exact marking can be found at the name plate of the meter. Model FCT3xx-Y0… (Transmitter outside the hazardous area, Sensor in Zone 0, 1 or 2)
Terminals 51 / 52 Switching output Terminals 41 / 42 Switching input Terminals 81 / 82 All inputs and outputs are electrically isolated from one other and from the power supply. 40 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
PE and the equipotential bonding PA. outputs. The rated voltage of the non-intrinsically-safe circuits is = 60 V. Change from two to one c olumn CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 41...
In the case of the remote mount design, the signal cable between the flowmeter sensor and the transmitter must measure at least 5 m (16.4 ft) in length. Install conduit seals within 18” (45 cm). 42 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Specific marking according to FM applies depending on the design of the flowmeter sensor (integral or remote mount design). IMPORTANT (NOTE) ABB reserves the right to make changes to the Ex-marking. The exact marking can be found at the name plate of the meter. Model FCx3xx-F2A… (remote mount design in Zone 2, Div 2)
Page 44
XP: CL I, Div 1, GPS BCD DIP: CL II, Div 1, GPS EFG DIP: CL III, Div 1, 2 Ex d ia ib IIC T6 or Ex d ia IIC T6 44 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
IMPORTANT (NOTE) ABB reserves the right to make changes to the Ex-marking. The exact marking can be found at the name plate of the meter. Model FCT3xx-Y0… (Transmitter in General Purpose and Sensor in Zone 2, Div 2 or Zone 0, Div 1)
Page 46
XP: CL I, Div 1, GPS BCD DIP: CL II, Div 1, GPS EFG DIP CL III, Div 1, 2 Ex d ib [ia] IIC T6 or Ex d [ia] IIC T6 46 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
The device configuration can be changed on this level. IMPORTANT (NOTE) The values entered are checked for plausibility; if they are not plausible, a corresponding message appears on the LCD display. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 47...
The LCD display now indicates the first menu item on the configuration level. 8. Use to select a menu. 9. Use (ENTER function) to confirm your selection. 48 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
1. Select the parameters you want to set in the menu. 2. Use (ENTER function) to call up the parameter for editing. 3. Use oder to select the required value. 4. Use (ENTER function) to confirm your selection. CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 49...
Totalizer Unit Mass Prog. Unit Qm Unit name Units factor Prog. Unit Qv Unit name Units factor Temp. unit Unit concentration Flowmeter primary Meter tube QmMax meter tube Order no. 50 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Page 51
Totalizer mass Overflow > F Totalizer < R Totalizer volume Overflow < R Totalizer net. mass Totalizer Reset Pulse Output Output of QvMax pulse Qm% max pulse Pulse Pulse width CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 51...
Page 52
Concentr. → l = 100 % Iout for Alarm Low Alarm High Alarm Switch contacts Contact input Contact output Label TAG number Descriptor Date Unit number Interface Communication Unit address 52 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Page 53
Change from one to tw o c olumns ™ Hastelloy C-22 is a Haynes International trademark ™ Hastelloy C-276 is a Haynes International trademark Change from two to one c olumn CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 53...
— IEC standards — NEPSI (China) — cFMus Approvals for Canada and United States IMPORTANT (NOTE) All documentation, declarations of conformity, and certificates are available in ABB's download area. www.abb.com/flow 54 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350...
Page 64
Oldends Lane, Stonehouse notice. With regard to purchase orders, the agreed Gloucestershire, GL10 3TA particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. Tel:...