Page 1
Dialysis System SW 8.09.13 Training module 7 – Report and Service Environment Ref: Operator’s M anual S W 8 .09.13 - Chapter 1 5 – Report E nvironment - Service M an c hapters 1 .2, 1 .3 a nd 1 0...
Page 2
Report and Service E nvironment Important Notice: Important : This p resentation d oes n ot r elieve a ny u ser Please note t hat the S W change t o 8.09.13 o f of the ...
Page 3
Report Screen Usability Event List • Automatic treatment records at regular intervals (default 3 0 min) • Manual snapshot - New R eading b utton - Present Time button (available i n a ll 5 M ain screens) Press the New Reading button to access ...
Page 4
Snapshot and Event reporting Press the A uto-chart time button to change the records time interval...
Page 5
Hygiene sub-screen • Ultrafilters history table is always available • Ultrafilters change procedure button is available i n t he standard environment • Default values for Ultrafilters history counters correspond to the maximum authorized ...
Page 6
Button Bar Usability LCD test Screen cleaning using alcohol solution only The test should be performed if the Operator observes or suspects that data are not correctly displayed on the Touch Screen ► Follow the instructions appearing on the screen (click = ...
Page 8
Artis Service mode – Two levels Two Service Levels with different a ccess codes Service Level 1: Service Level 2: • Technicians a ccess • Access code 169000 • Preset environment • Date and time change ...
Page 9
Artis Service mode – NavPad Preset Screen + Service code Service Overview Screen Exit Service Mode Fluid S creen Blood S creen Fluid c ircuit schematics a nd Blood circuit s chematics a nd sensors/actuators ...
Page 10
Artis Service mode – Preset Screen All presets can be viewed by scrolling the different select pads Service level 1 Possible to set/change: • Time date • Auto Start Service level 2 Possible to set/change: • Installed features •...
Page 11
Artis Service mode – Preset transfer Save to … Load from … By the use of a USB key presets can be: • Downloaded from one machine (Import) Import Export • Uploaded to another (Export)
Page 12
Artis Service mode - Calibration The Calibration button available in the button bar gives access to the Calibration Component Map from which individual calibrations and tests can be performed...
Page 13
Artis Service mode - Blackbox In the Blackbox, A rtis stores the latest 1 to max 24 sessions • According to the principle “ first in- first out” Example of information stored: • Treatment data • Alarms •...
Page 14
Blackbox files The Blackbox files can be downloaded t o an USB key via the Report screen, requiring Service L evel 2 access code - During preparation until A uto-Prime is booked or when no process is running The downloaded ...
Page 15
Blackbox downloading procedure Files Target + Service code Export Export All Export Blackbox F ile Close C urrent Session ► Place a USB s tick in t he Connectivity Panel U SB p ort ►...
Page 16
Hands on Module 7 - Event list - New r eading/Snapshot - Time interval - Hygiene s creen - LDC t est - Screen Cleaning - Access Service Level 1 and 2 - Preset process -...
Page 17
For the safe and proper use of the Artis Dialysis System, refer to Warnings, Precautions and complete Instructions for Use in the Operator’s Manual available by contacting customer support. Gambro is an indirect subsidiary o f Baxter International Inc. ...