Connecting The Tb-4304 - NI PXIe-4304 Calibration Procedure

32 ch, 24-bit, +-42 v, 5 ks/s or 51.2 ks/s simultaneous filtered data acquisition module
Table of Contents

Advertisement

Connecting the TB-4304

The TB-4304 provides connections for the PXIe-4304/4305. Figure 1 shows the pin assignments
of the TB-4304.
Figure 1. TB-4304 Circuit Board Parts Locator Diagram
S/N
158016B-02L
AI7-
AI7+
AI6-
AI6+
AI5-
AI5+
AI4-
AI4+
AI3-
AI3+
AI2+
AI2+
AI1-
AI1+
AI0-
AI0+
Each channel consists of two terminal connections specific to that channel as shown in Table 2.
You can verify or adjust the accuracy for any or all channels depending on the desired test
coverage. Refer to Figure 2 and only connect the input channels required for verification or
adjustment in parallel.
Refer to Table 2 for the analog signal names of the TB-4304.
Signal Name
AIGND
Complete the following steps to connect the TB-4304.
1.
Install the PXIe-4304/4305 and the TB-4304 in the PXI Express chassis according to the
instructions in the NI PXIe-4304/4305 and TB-4304 User Guide and Terminal Block
Specifications.
2.
Configure the PXIe-4139 to voltage output mode and enable remote sensing. Connect the
PXIe-4139 output to the TB-4304 as shown in Figure 2.
4 | ni.com | NI PXIe-4304/4305 Calibration Procedure
NI TB-4304
AI15-
AI15+
AI14-
AI14+
AI13-
AI13+
AI12-
J9
AI12+
AI11-
AI11+
AI10-
AI10+
AI9-
J7
AI9+
AI8-
AI8+
FOR PATENTS: NI.COM/PATENTS
Table 2. TB-4304 Analog Signal Names
AI+
AI-
COPYRIGHT 2015
AI23-
AI23+
AI22-
AI22+
AI21-
AI21+
AI20-
AI20+
AI19-
AI19+
AI18-
AI18+
AI7-
AI7+
AI6-
AI6+
Signal Description
Positive input voltage terminal
Negative input voltage terminal
Analog ground input
CR2
C4
UL94V
DGND
AI31-
PFI0
AI31+
AI30-
AI30+
AI29-
AI29+
AI28-
AI28+
AI17-
J17
AI17+
AI16-
AI16+
AI5-
AI5+
J15
AI4-
AI4+
-0