5 Setups, alignments and adjustments
Subparameter
default 0
compression limit mode
default 1
compression force for force
switch setting 1
default 2
compression force for force
switch setting 2
default 3
compression force for force
switch setting 3
default 4
compression force for force
switch setting 4
default 5
compression force for force
switch setting 5
default 6
compression force for force
switch setting 6
Subparameter 0 determines the compression limit mode. The available
modes are 0 (absolute limit) or 5 (temporary limit). With selection 0, the
compression stops at default 1. With selection 5, compression can be
continued by five individual steps up until default 6.
Range for subparameters 1-6 is 5-25,5 kg (11-56lbs.), lower parameter not
being less than the following parameter.
Subparameter 6 will be the maximum compression force option.
Comments
Adjust the desired parameter 2 and compression forces (switch positions
1-6) according to desires of the customer.
If needed, use customer's own reference when adjusting forces. Adjust
limit values to comply with reference system.
Compression minimum force limit is not recommended to adjust to be
less than 8 kg (16lbs.).
See appendix Software Block diagram.
70
Meaning
GE Healthcare
Manufacturer
default setting
0 = absolute limit
100 (10kg)
120 (12kg)
140 (14kg)
160 (16kg)
180 (18kg)
200 (20kg)
32551-IMG rev 6