32551-IMG rev 6
#53PAr/b The density display shows b which means the Anode Brake
Mode. Select 0 or 1 using the Den+ and Den- buttons.
0 = Anode Brake off
1 = Anode Brake on
The factory default is to have the anode brake off after an
exposure. This reduces the heat loading of the tube housing
and applies less stress to the bearings.
Press kV+
#53PAr/t The density display shows t. In this mode the software for
Alpha RT is selected.
0 = Not for Alpha RT
1 = Not for Alpha RT
2 = Only AA and manual mode (Alpha RT)
3 = Only A and manual mode (Alpha RT)
4 = AA ->> A ->> manual mode (factory default
Alpha RT)
5 = A ->> AA ->> manual mode (Alpha RT)
Press kV+
#53PAr/I Collimator type (Interlock)
1 for MGB-500 Collimator (with interlock)
0 for MGB-400 Collimator (no interlock)
Press kV+
#53PAr/d C-arm control board selection
0 = C-arm control board (32310)
1 = CD-board (32010)
Press Autoexp to leave the PAr program.
Press kV+
GE Healthcare
5 Setups, alignments and adjustments