Troubleshooting Problems with Boot Process
•
Error: R&S FSP does not start the measurement application.
Following switch-on, the R&S FSP first boots the computer BIOS. After successful initialization of the
computer the Windows XP operating system starts up. Subsequently, the test application is loaded as
start-up program. Simultaneously, selftests are performed at various locations and error messages, if
any, are output. It is advisable for troubleshooting to connect a keyboard to the keyboard socket.
Normal action
Start R&S FSP
Subsequent to switching on the R&S FSP, the following BIOS
message is displayed:
Phoenix/AwardBIOS v6.00PG, An Energy Star Ally
Copyright (C) 1984-2003, Phoenix Technolgies LTD
R&S ANALYZER BIOS V5.1-16-3
Main Processor : Intel(R) Celeron(R) M processor 800MHz (100x8.0)
Press DEL to enter SETUP
03/14/2006-i855E-W83727HF-6A692EF0C-00
The computer starts the hardware test. The test results are
displayed on the screen. If errors occur during the boot procedure,
these messages may indicate defects.
Phoenix/AwardBIOS v6.00PG, An Energy Star Ally
Copyright (C) 1984-2003, Phoenix Technolgies LTD
R&S ANALYZER BIOS V5.1-16-3
Main Processor : Intel(R) Celeron(R) M processor 800MHz (100x8.0)
Memory Testing : 491520K OK + 32M shared memory
Memory Frequency For DDR333
03/14/2006-i855E-W83727HF-6A692EF0C-00
The memory test issues the memory capacity of the front module
controller. The basic version of the R&S FSP provides 512 Mbytes.
Subsequently, BIOS starts the hardware check.
1164.4556.82
Error and error cause
3.71
Troubleshooting
E-1
Need help?
Do you have a question about the FSP3 and is the answer not in the manual?
Questions and answers