Page 1
Service Manual Instrument Spectrum Analyzer ® ® R&S FSP3 R&S FSP30 1164.4391.03 1164.4391.30/.39 R&S ® FSP7 R&S ® FSP31 1164.4391.07 1164.4391.31 ® ® R&S FSP13 R&S FSP40 1164.4391.13 1164.4391.40 Printed in Germany Test and Measurement Division 1164.4556.82-02-...
Page 2
R&S ® is a registered trademark of Rohde & Schwarz GmbH & Co. KG Trade names are trademarks of the owners 1164.4556.82...
Page 3
R&S FSP Tabbed Divider Overview Tabbed Divider Overview Index Safety Instructions Certificate of Quality Spare Parts Express Service List of R&S Representatives Contents of Manuals for Spectrum Analyzer R&S FSP Service and Repair Tabbed Divider Chapter 1: Performance Test Chapter 2: Adjustment Chapter 3: Repair...
Page 5
Before putting the product into operation for the first time, make sure to read the following S a f e t y I n s t r u c t i o n s Rohde & Schwarz makes every effort to keep the safety standard of its products up to date and to offer its customers the highest possible degree of safety.
Page 6
Safety Instructions Observing the safety instructions will help prevent personal injury or damage of any kind caused by dangerous situations. Therefore, carefully read through and adhere to the following safety instructions before putting the product into operation. It is also absolutely essential to observe the additional safety instructions on personal safety that appear in other parts of the documentation.
Page 7
Safety Instructions 4. If products/components are mechanically 10. Intentionally breaking the protective earth and/or thermically processed in a manner connection either in the feed line or in the that goes beyond their intended use, product itself is not permitted. Doing so can hazardous substances (heavy-metal dust result in the danger of an electric shock such as lead, beryllium, nickel) may be...
Page 8
Safety Instructions 19. If a product is to be permanently installed, matching Rohde & Schwarz type (see the connection between the PE terminal on spare parts list). Batteries and storage site and the product's PE conductor must batteries are hazardous waste. Dispose of be made first before any other connection them only in specially marked containers.
Page 9
Por favor lea imprescindiblemente antes de la primera puesta en funcionamiento las siguientes informaciones de seguridad Informaciones de seguridad Es el principio de Rohde & Schwarz de tener a sus productos siempre al día con los estandards de seguridad y de ofrecer a sus clientes el máximo grado de seguridad. Nuestros productos y todos los equipos adicionales son siempre fabricados y examinados según las normas de seguridad vigentes.
Page 10
Informaciones de seguridad Tener en cuenta las informaciones de seguridad sirve para tratar de evitar daños y peligros de toda clase. Es necesario de que se lean las siguientes informaciones de seguridad concienzudamente y se tengan en cuenta debidamente antes de la puesta en funcionamiento del producto. También deberán ser tenidas en cuenta las informaciones para la protección de personas que encontrarán en otro capítulo de esta documentación y que también son obligatorias de seguir.
Page 11
Informaciones de seguridad seguridad (control a primera vista, control de peligro a causa de la radiación conductor protector, medición de resistencia electromagnética. El empresario está de aislamiento, medición de medición de la comprometido a valorar y señalar areas de corriente conductora, control trabajo en las que se corra un riesgo de...
Page 12
Informaciones de seguridad 12. No utilice nunca el producto si está dañado el 20. En caso de que los productos que son cable eléctrico. Asegure a través de las instalados fijamente en un lugar sean sin medidas de protección y de instalación protector implementado, autointerruptor o adecuadas de que el cable de eléctrico no similares objetos de protección, deberá...
Page 13
Informaciones de seguridad 27. Baterías y acumuladores no deben de ser 31. Las asas instaladas en los productos sirven expuestos a temperaturas altas o al fuego. solamente de ayuda para el manejo que Guardar baterías y acumuladores fuera del solamente está previsto para personas. Por alcance de los niños.
Page 15
Safety Instructions WARNING Danger of injuries When removing the rear feet, the unit can slip out of the cabinet. Put the unit onto the front handles, before removing the rear feet and taking off the cabinet. Thus the risk of personal injuries and damages to the unit is avoided. When mounting the cabinet take care not to pen in the fingers.
Page 17
Certified Quality System DIN EN ISO 9001 : 2000 DIN EN 9100 : 2003 DIN EN ISO 14001 : 1996 DQS REG. NO 001954 QM/ST UM QUALITÄTSZERTIFIKAT CERTIFICATE OF QUALITY CERTIFICAT DE QUALITÉ Sehr geehrter Kunde, Dear Customer, Cher Client, Sie haben sich für den Kauf eines you have decided to buy a Rohde &...
Page 19
Spare Parts Express Service Phone: +49 89 4129 - 12465 Fax: +49 89 41 29 - 13306 E-mail:werner.breidling@rsd.rohde-schwarz.com In case of urgent spare parts requirements for this Rohde & Schwarz unit, please contact our spare parts express service. Outside business hours, please leave us a message or send a fax or e-mail.
Page 21
Customer Support Technical support – where and when you need it For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or applications of Rohde &...
Page 23
Address List Headquarters, Plants and Subsidiaries Locations Worldwide Headquarters Please refer to our homepage: www.rohde-schwarz.com ◆ ROHDE&SCHWARZ GmbH & Co. KG Phone +49 (89) 41 29-0 Sales Locations Mühldorfstraße 15 · D-81671 München Fax +49 (89) 41 29-121 64 ◆ Service Locations P.O.Box 80 14 69 ·...
Page 25
Service and Repair R&S FSP Contents of Manuals for Spectrum Analyzer R&S FSP Service Manual - Instrument The service manual - instrument informs on how to check compliance with rated specifications, on instrument function, repair, troubleshooting and fault elimination. It contains all information required for repairing the R&S FSP by the replacement of modules.
Page 26
R&S FSP Manuals Service and Repair Please contact your Rohde & Schwarz support center or our spare parts express service if you need service or repair of your equipment or to order spare parts and modules. The list of the Rohde & Schwarz representatives and the address of our spare parts express service are provided at the beginning of this service manual.
R&S FSP Contents - Performance Test Contents - Chapter 1 "Performance Test" 1 Performance Test....................1.1 Test Instructions..........................1.1 Measurement Equipment and Accessories ...................1.1 Performance Test R&S FSP......................1.4 Checking Reference Frequency Accuracy................1.4 Checking Immunity to Interference ..................1.4 1st IF Image Frequency Rejection .................1.5 2nd IF Image Frequency Rejection................1.5 3rd IF Image Frequency Rejection.................1.5 1st IF Rejection ......................1.6...
Page 28
Contents - Performance Test R&S FSP Performance Test Option External Mixing R&S FSP-B21 ............1.37 Checking LO-Levels .......................1.37 Checking the input LO / IF (2-Port-Mixers) ...............1.38 Checking the input IF (3-Port-Mixers) ..................1.38 Checking the bias supply .......................1.39 Performance Test Report R&S FSP ....................1.40 Performance Test Report Option R&S FSP-B6 ................1.53 Performance Test Report Option R&S FSP-B9 ................1.54 Performance Test Report Option R&S FSP-B15 .................1.55...
R&S FSP Measurement Equipment 1 Performance Test Test Instructions • The rated specifications of the analyzer are tested after a warm-up time of at least 15 minutes and overall calibration. Only in this case can compliance with the specified data be ensured. Start of overall adjustment: [CAL : CAL TOTAL] •...
Page 30
Measurement Equipment R&S FSP Item Type of Specifications recommended Equipment Alternative equipment recommended equipment Signal generator R&S FSP 07: 10 MHz to 7 GHz R&S SMP02 HP 63620 B Third order intercept R&S FSP 13: 10 MHz to 13.6 GHz R&S SMP02 HP 63620 B R&S FSP 30: 10 MHz to 30 GHz...
Page 31
R&S FSP Measurement Equipment Item Type of Specifications recommended Equipment Alternative equipment recommended equipment Voltmeter DC and AC voltages R&S URE 34401A TG modulation (2 units) (Agilent) TV signal source Checking the TV trigger Power sensor Frequency range R&S NRV-Z55 B21: LO output power 7.0 GHz up to 15.5 GHz max.
Performance Test R&S FSP Performance Test R&S FSP Checking Reference Frequency Accuracy Test equipment: Frequency counter (refer to "Measurement Equipment", item 1): Accuracy < 1x10 Frequency range up to 10 MHz Test setup: Connect frequency counter to 10 MHz reference output of the R&S FSP (rear panel).
R&S FSP Performance Test 1st IF Image Frequency Rejection Additional signal generator - Frequency + 6952.8 MHz settings: Additional R&S FSP settings: - [ FREQ : CENTER : {f Refer to "Performance Test Report" table for values of f Measurement: Set marker to peak of signal: [ MKR : PEAK ]...
Performance Test R&S FSP 1st IF Rejection Additional signal generator - Frequency 3476.4 MHz settings: Additional R&S FSP settings: - [ FREQ : CENTER : {f Refer to "Performance Test Report" table for values of f Measurement: Set marker to peak of signal: - [ MKR : PEAK ] Evaluation:...
Performance Test R&S FSP Checking IF Filters Test equipment: Signal generator (refer to "Measurement Equipment", item 3): Frequency 128 MHz Level 0 dBm Test setup: Connect RF output of the signal generator to RF input of the R&S FSP. Checking Bandwidth Switching Level Accuracy Reference measurement (RBW 10 kHz) Signal generator settings: - Frequency:...
Performance Test R&S FSP Checking Noise Display Test equipment: termination (refer to "Measurement Equipment", item 7) Frequency range R&S FSP 3: to 3 GHz R&S FSP 7: to 7 GHz R&S FSP 13: to 13.6 GHz R&S FSP 30: to 30 GHz R&S FSP 40: to 40 GHz Test setup:...
R&S FSP Performance Test Checking Level Accuracy and Frequency Response Test equipment: - Signal generator: R&S FSP 3: Refer to "Measurement Equipment", item 3. R&S FSP 7/13/30:Refer to "Measurement Equipment", items 3 & 4. Frequency range R&S FSP 3: 1 MHz to 3 GHz R&S FSP 7: 1 MHz to 7 GHz R&S FSP 13:...
Page 40
Performance Test R&S FSP Determining the level accuracy at 128 MHz Test setup: Connect power sensor (item 9) to the power meter and execute function ´ZERO´ when there is no signal applied to the power sensor. Connect power sensor to RF output of signal generator. Signal generator settings: - Frequency 128 MHz...
Page 41
R&S FSP Performance Test Checking frequency response Test setup: Connect RF output of signal generator to input of power splitter. Connect output 1 of power splitter to power sensor / power meter. Connect output 2 of divider to RF input of the R&S FSP. Signal generator settings: - Level 0 dBm...
Performance Test R&S FSP Checking Display Linearity Test equipment: - Signal generator (refer to "Measurement Equipment", item 3) Frequency 128 MHz Maximum level 10 dBm - Step attenuator (refer to "Measurement Equipment", item 11) Frequency 128 MHz Attenuation 0 to 100 dB in 1 dB steps Attenuation accuracy <...
Page 43
R&S FSP Performance Test Alternative measurement R&S FSP settings: - [ PRESET ] - [ AMPT : RF ATTEN MANUAL : 10 dB ] - [ AMPT : 0 dBm ] - [ FREQ : CENTER : 128 MHz ] - [ SPAN : 0 Hz ] - [ TRACE : DETECTOR : AV ] - [ SWEEP: SWEEP TIME MANUAL: xx ms ]...
Performance Test R&S FSP Checking RF Attenuator Test equipment: - Signal generator (refer to "Measurement Equipment", item 3) Frequency 128 MHz Maximum level 0 dBm - Step attenuator (refer to "Measurement Equipment", item 11) Frequency 128 MHz Attenuation 0 to 80 dB in 10 dB steps Attenuation accuracy <...
R&S FSP Performance Test Checking Reference Level Switching (IF Gain) Test principle: The IF gain of the R&S FSP can be switched from 0 to 50 dB by changing the reference level at fixed RF attenuation. To prevent the IF gain accuracy from being confused with the log amplifier accuracy, it is determined by comparison using an external precision attenuator.
Page 46
Performance Test R&S FSP Measurement Step attenuator settings: Attenuation Refer to table below for values of a R&S FSP settings: - [ AMPT : {reference level} dBm ] Refer to table below for values of reference level. - [ MKR : PEAK ] Evaluation: The difference between the level of the input signal of the R&S FSP...
R&S FSP Performance Test Performance Test Option TV and RF Trigger: R&S FSP-B6 Checking RF Trigger Test equipment: - Signal generator (refer to "Measurement Equipment", item 3) Frequency 128 MHz Test setup: Connect test output of the signal generator to RF input of the R&S FSP.
Performance Test R&S FSP Performance Test Option Tracking Generator – R&S FSP-B9 Checking Output Level Test equipment: - N cable (refer to "Measurement Equipment", item 14) Frequency up to 3 GHz Maximum attenuation < 0.2 dB Test setup: Connect tracking generator output to RF input of the R&S FSP. R&S FSP settings: - [ PRESET ] - [ MODE NETWORK ]...
R&S FSP Performance Test Checking Frequency Response Test equipment: - N cable (refer to "Measurement Equipment", item 14) - 10 dB attenuator (refer to "Measurement Equipment", item 20) Frequency up to 3 GHz Maximum attenuation < 0.2 dB Test setup: Connect 10 dB attenuator to tracking generator output port.
Performance Test R&S FSP Checking Modulation Checking I/Q Modulation Test equipment: - Arbitrary waveform generator (refer to "Measurement Equipment", item 16) - Spectrum Analyzer (refer to "Measurement Equipment", item 15) - 2 Voltmeters (refer to "Measurement Equipment", item 17) - 2 BNC T pieces Test setup with R&S ADS: Apply the I and Q outputs of ADS to the I and Q inputs of the analyzer.
R&S FSP Performance Test Checking Amplitude Modulation Test equipment: - Arbitrary waveform generator (refer to "Measurement Equipment", item 16) - Spectrum analyzer (refer to "Measurement Equipment", item 15) - 2 voltmeters (refer to "Measurement Equipment", item 17) - 2 BNC T pieces Test setup with R&S ADS: Apply the I and Q outputs of ADS to the I and Q inputs of the analyzer.
Performance Test R&S FSP Checking Frequency Modulation Test equipment: - Arbitrary waveform generator (refer to "Measurement Equipment", item 16) - Spectrum analyzer (refer to "Measurement Equipment", item 15) - 2 voltmeters (refer to "Measurement Equipment", item 17) - 2 BNC T pieces Test setup with R&S ADS: Apply the I and Q outputs of ADS to the I and Q inputs of the analyzer.
Performance Test R&S FSP Performance Test Option Electronic Attenuator- R&S FSP-B25 Checking Noise Display with Preamplifier (B25) Test equipment: termination (refer to "Measurement Equipment", item 7) Frequency range: R&S FSP 3 to 3 GHz R&S FSP 7 to 7 GHz Terminate the RF input of the R&S FSP with 50 .
R&S FSP Performance Test Checking Level Accuracy and Frequency Response with Preamplifier Test equipment: - Signal generator : R&S FSP 3: Refer to "Measurement Equipment", item 3. R&S FSP 7: Refer to "Measurement Equipment", items 3 & 5. Frequency range: R&S FSP 3: 10 MHz to 3 GHz R&S FSP 7:...
Page 58
Performance Test R&S FSP Determining the level accuracy at 128 MHz Test setup: Connect power sensor (item 9) to the power meter and execute function ´ZERO´ when there is no signal applied to the power sensor. Connect power sensor to RF output of signal generator. Signal generator settings: - Frequency 128 MHz...
Page 59
R&S FSP Performance Test Reference measurement: Determine signal level L powermeter Set marker to peak of signal: - [ MKR : PEAK ] The signal level L is displayed by the level reading of marker 1. 128MHz powermeter Measurement Signal generator settings: - Frequency fresp Refer to "Performance Test Report"...
Performance Test R&S FSP Checking Frequency Response with Electronic Attenuator Test equipment: - Signal generator : R&S FSP 3: Refer to "Measurement Equipment", item 3. R&S FSP 7: Refer to "Measurement Equipment", items 3 & 5. Frequency range: R&S FSP 3: 10 MHz to 3 GHz R&S FSP 7: 10 MHz to 7 GHz...
Page 61
R&S FSP Performance Test Test setup: Connect RF output of the signal generator to input of the divider. Connect output 1 of the power splitter to the power sensor / power meter. Connect output 2 of the power splitter to RF input of the R&S FSP. Signal generator settings: - Level 0 dBm...
Performance Test R&S FSP Checking Nonlinearities with Electronic Attenuator Third Order Intercept Test equipment: - 2 signal generators R&S FSP 3: Refer to "Measurement Equipment", items 2 & 3. R&S FSP 7: Refer to "Measurement Equipment", items 2 & 5. Frequency range: R&S FSP 3: 10 MHz to 3 GHz...
R&S FSP Performance Test Checking RF Attenuator (with Option B25) Test equipment: - Signal generator (refer to "Measurement Equipment", item 3) Frequency 128 MHz Maximum level 10 dBm - Step attenuator (refer to "Measurement Equipment", item 11) Frequency 128 MHz Attenuation 0 to 80 dB in 5 dB steps Attenuation accuracy...
Performance Test R&S FSP Checking Electronic Attenuator Accuracy Test equipment: - Signal generator (refer to "Measurement Equipment", item 3) Frequency 128 MHz Maximum level 0 dBm - Step attenuator (refer to "Measurement Equipment", item 11) Frequency 128 MHz Attenuation 0 to 40 dB in 5 dB steps Attenuation accuracy <...
R&S FSP Performance Test Performance Test Option External Mixing R&S FSP-B21 Checking LO-Levels Measurement equipment: - Power sensor (section „Measurement Equipment“, item 19) Frequency range 7.0 MHz to 13.2 GHz Max. input power + 23 dBm 2.5% reffered to measured power Impedance Z = 50 - Power meter (Section „Measurement Equipment“, item 8)
Performance Test Report R&S FSP Performance Test Report R&S FSP Table 1-1 Performance Test Report ROHDE & SCHWARZ Performance Test Report Spectrum Analyzer R&S FSP Version 27-Jul-00 Model (R&S FSP-3/7/13/30/40): Order number: 1093.4495. Serial number: Test person: Date: Signature: Characteristic Included in Min.
Page 69
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Image frequency Page 1.5 rejection, 2nd IF, f R&S FSP 3/7/13/30/40: ________ 100 MHz R&S FSP 7: ________ 3100 MHz ________ 5000 MHz ________ 6999 MHz R&S FSP 13/30/40: ________...
Page 70
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance 3rd order Page 1.7 intercept point, f R&S FSP 3/7/13/3/40: ________ 28 MHz ________ 106 MHz ________ 261 MHz ________ 640 MHz 1000 MHz ________ ________ 1700 MHz...
Page 71
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance FFT bandwidth Page 1.8 Level accuracy 1 Hz -0.2 ________ +0.2 3 Hz -0.2 ________ +0.2 10 Hz -0.2 ________ +0.2 30 Hz -0.2 ________ +0.2 100 Hz...
Page 72
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Noise display Page 1.10 R&S FSP 3: f noise ________ -142 10.99 MHz ________ -142 19.99 MHz ________ -142 49.99 MHz ________ -142 99.99 MHz ________ -142 199.9 MHz...
Page 73
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Noise display Page 1.10 R&S FSP 13/30/40: noise ________ -135 3099 MHz ________ -135 3499 MHz ________ -135 3999 MHz ________ -135 4499 MHz 4999 MHz ________ -135...
Page 74
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Noise display Page 1.11 R&S FSP 40: f noise ________ -120 22999 MHz ________ –120 23999 MHz ________ –120 24999 MHz ________ –120 25999 MHz ________ –120 26999 MHz...
Page 75
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.11 RF attenuation 10 dB R&S FSP 7/13/30/40: fresp ________ 3010 MHz ________ 3500 MHz ________ 4000 MHz 4500 MHz ________ ________ 5000 MHz ________...
Page 76
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.11 RF attenuation 10 dB R&S FSP 30/40: fresp ________ 14000 MHz ________ 15000 MHz ________ 16000 MHz ________ 17000 MHz 18000 MHz ________ ________...
Page 77
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.11 RF attenuation 20 dB fresp 1 MHz -0.5 ________ +0.5 10 MHz -0.5 ________ +0.5 50 MHz -0.5 ________ +0.5 100 MHz -0.5 ________...
Page 78
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Display linearity Page 1.14 RBW 300 Hz 10 dB ________ 10.2 12 dB ________ 14 dB ________ 16 dB ________ 18 dB ________ 20 dB Reference 22 dB -2.2...
Page 79
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Display linearity Page 1.14 RBW 300 kHz 10 dB ________ 10.2 12 dB ________ 14 dB ________ 16 dB ________ 18 dB ________ 20 dB Reference 22 dB -2.2...
Page 80
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Reference level Page 1.17 switching accuracy Reference level 0 dBm +9.8 ________ +10.2 -10 dBm Reference -20 dBm -10.2 ________ -9.8 -30 dBm -20.2 ________ -19.8 -40 dBm...
R&S FSP Performance Test Report Performance Test Report Option R&S FSP-B6 Table 1-1: Performance Test Report Option R&S FSP-B6 ROHDE & SCHWARZ Performance Test Report Option R&S FSP – B6 Version 11-June-02 Serial number: Test person: Date: Signature: Characteristic Included in Min.
Performance Test Report R&S FSP Performance Test Report Option R&S FSP-B9 Table 1-2: Performance Test Report Option R&S FSP-B9 ROHDE & SCHWARZ Performance Test Report Option R&S FSP-B9 Version 27-Jul-00 Serial number: Test person: Date: Signature: Characteristic Included in Min. value Actual value Max.
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.23 Tracking generator Range 2 GHz - 3 GHz Output level: 0 dBm ________ -10 dBm ________ -25 dBm ________ Checking modulation Page 1.24 Tracking generator I/Q modulation...
Performance Test Report R&S FSP Performance Test Report Option R&S FSP-B25 Table 1-3 Performance Test Report Option R&S FSP-B25 ROHDE & SCHWARZ Performance Test Report Option R&S FSP-B25 Version 27-Jul-00 Serial number: Test person: Date: Signature: Characteristic Included in Min. value Actual value Max.
Page 85
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.29 with preamplifier (B25) R&S FSP 3/7: f fresp ________ 1 MHz ________ 10 MHz ________ 50 MHz ________ 100 MHz 200 MHz ________ ________...
Page 86
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.32 with electronic attenuator (B25) = 5 dB R&S FSP 3/7: f fresp ________ 10 MHz ________ 50 MHz ________ 100 MHz ________ 200 MHz ________...
Page 87
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.32 with electronic attenuator (B25) = 10 dB R&S FSP 3/7: f fresp ________ 10 MHz –1 ________ 50 MHz –1 ________ 100 MHz –1...
Page 88
Performance Test Report R&S FSP Characteristic Included in Min. value Actual value Max. value Unit Tolerance Frequency response Page 1.32 with electronic attenuator (B25) = 15 dB R&S FSP 3/7: f fresp ________ 10 MHz ________ 50 MHz ________ 100 MHz ________ 200 MHz ________...
Page 89
R&S FSP Performance Test Report Characteristic Included in Min. value Actual value Max. value Unit Tolerance 3rd order intercept Page 1.34 point with elec. att.,f R&S FSP 3/7: ________ 28 MHz ________ 106 MHz ________ 261 MHz ________ 640 MHz 1000 MHz ________ ________...
Page 90
Performance Test Report R&S FSP Performance Test Report Option R&S FSP-B21 Tabelle 1-3: Performance Test Report Option B21 Characteristic Included in Min. value Actual value Max. value Unit Tolerance LO-level Page 1.36 Center _________ 140,5 GHz 14,5 16,5 _________ 150,5 GHz 14,5 16,5 _________...
Page 91
R&S FSP Contents - Adjustment Contents - Chapter 2 "Adjustment" 2 Adjustment......................2.1 Service Menu ............................2.1 Entering the Password ....................2.2 Alignment Functions ......................2.2 Manual Adjustment ..........................2.3 Test Instructions ........................2.3 Measuring Equipment and Accessories...................2.3 Adjusting the level measurement accuracy................2.4 Adjusting the frequency accuracy ....................2.6 Adjustment of Module Data ......................2.8 Frequency response correction......................2.8 1164.4556.82...
Page 92
Contents - Adjustment R&S FSP This page is left blank intentionally. 1164.4556.82 I-2.2...
R&S FSP Service Menu 2 Adjustment The following chapter describes the adjustment of the reference sources as well as the software- controlled adjustment of individual module data following module replacement. The R&S FSP permits the following manual adjustments: • Adjustment of the 10 MHz reference oscillator which determines the frequency accuracy of the R&S FSP.
Service Menu R&S FSP Entering the Password SETUP SERVICE submenu: The ENTER PASSWORD softkey allows the entry of a password. ENTER PASSWORD The R&S FSP contains a variety of service functions which, if incorrectly used, can impair the functionality of the analyzer. These functions are normally not accessible and are only usable after the entry of a password.
R&S FSP Manual Adjustment Manual Adjustment In the following, the measuring instruments and auxiliary means required for the manual adjustment of the R&S FSP, the appropriate preparations of the instrument as well as the individual adjustments will be explained. Test Instructions •...
Manual Adjustment R&S FSP Adjusting the level measurement accuracy Test equipment: - Signal generator (Section "Measurement Equipment", item 2): frequency 128 MHz level -30 dBm - power meter (Section "Measurement Equipment", item 3) - power sensor (Section "Measurement Equipment", item 4) frequency 128 MHz maximum power...
Page 97
R&S FSP Manual Adjustment CAUTION The following re-alignment changes the level of the internal calibration source. Since this adjustment influences the level measurement accuracy of the R&S FSP, it is strongly recommended to perform this adjustment only if the level is not within the tolerance. R&S FSP settings: - [ SETUP : SERVICE : ENTER PASSWORD : 894129 ENTER ] - [ SETUP : SERVICE : CAL SIGNAL POWER ]...
Manual Adjustment R&S FSP Adjusting the frequency accuracy Preparation: The measurement can be performed either with a signal generator at connector RF INPUT (front of R&S FSP) at 1 GHz or at connector EXT REF OUT (rear of R&S FSP) at 10 MHz using a frequency counter For the adjustment, the R&S FSP must be set to internal reference.
Page 99
R&S FSP Manual Adjustment Preparation for adjustment with frequency counter: Test equipment: Frequency counter (Section "Measurement Equipment", item 1): error < 1x10 frequency range up to 10 MHz Test setup: connect frequency counter to 10-MHz reference output of the R&S FSP (rear panel) R&S FSP settings: - [SETUP : REFERENCE INT / EXT ]...
Adjustment of Module Data R&S FSP Adjustment of Module Data All boards of the R&S FSP contain EEPROMS for storage of board data. In addition to some standard information such as module name, serial number, hardware status and date of manufacture, these stored data items contain important pieces of information within value tables from module pre-testing, e.g.
Page 101
R&S FSP Contents - Repair Contents - Chapter 3 "Repair" 3 Repair ........................3.1 Instrument Design and Function Description ................3.1 Block Diagram ..........................3.1 Description of Block Diagram....................3.2 Attenuator........................3.2 Electronic Attenuator (Option R&S FSP-B25)..............3.2 RF to IF Conversion for Frequencies < 3 GHz – Frontend ..........3.3 RF to IF Conversion for Frequencies >...
R&S FSP Instrument Design and Function Description 3 Repair This chapter describes the design of the R&S FSP, simple measures for repair and troubleshooting and, in particular, the replacement of modules. For troubleshooting and diagnostics, a selftest is available for polling diagnostic voltages of the modules and detecting limit violations.
Instrument Design and Function Description R&S FSP Description of Block Diagram The R&S FSP is a triple-conversion superjet receiver (double-conversion for receive frequencies >3 GHz) for the frequency range from 9 kHz to several GHz, depending on the instrument model. The signals are processed by one RF board (two for models >...
R&S FSP Instrument Design and Function Description RF to IF Conversion for Frequencies < 3 GHz – Frontend The frontend converts the receive frequencies in the range 9 kHz to 3 GHz to the low 20.4 MHz IF. The frontend also includes the required local oscillators and associated frequency processing circuits. The unit is made up of two PCBs accommodated in a common housing.
Instrument Design and Function Description R&S FSP RF to IF Conversion for Frequencies > 3 GHz - Microwave Converter The high frequency models of the R&S FSP (frequency range > 3 GHz) also comprise a microwave converter board. This module converts the analyzer input signals in the range above 3 GHz to the second IF of 404.4 MHz.
R&S FSP Instrument Design and Function Description IF Filter – Module The frontend in the signal path is followed by the IF filter module (analog IF): X132 X146 IF-FILTER BOARD Signal Path and selftest dither source Selfttest detector selfttest detector Selftest detektor Selftest detektor RBW 100kHz...
Instrument Design and Function Description R&S FSP Reference Frequency 10 MHz - IF Filter Module The reference frequency is generated by a TCXO, and the frequency is adjustable by means of a D/A converter. As another choice, an OCXO or an external reference input (REF IN) can be switched on. In any case, the active reference will be routed to the reference output.
R&S FSP Instrument Design and Function Description Processing of Measured Data - Detector Board The following sections explain the signal and data paths required for result processing in the various operating modes: Sweep Control Magnitude DCON ZF / Video Control Digital Down Phase...
Instrument Design and Function Description R&S FSP FFT Bandwidth In the FFT bandwidth mode, the synthesizer is set to the desired frequency via the DCON which also generates the tuning voltages for the analog hardware. The signal path is routed via ADC, Corr RAM and DDC.
R&S FSP Instrument Design and Function Description Tracking Generator (Option R&S FSP-B9) The tracking generator generates a signal exactly at the input frequency of the R&S FSP during normal operation. For frequency-converting measurements, it is possible to set a constant frequency offset of ±150 MHz between the receive frequency of the R&S FSP and the output signal of the tracking generator.
Instrument Design and Function Description R&S FSP Front Panel The front panel consists of an aluminum case panel (part of the R&S FSP frame) and a mounting plate which accommodates the LCD, the backlight inverter, the keyboard mat with the membrane and the spinwheel.
R&S FSP Instrument Design and Function Description Power Supply Module The power supply module provides all currents necessary for the operation of the R&S FSP. It can be switched off by means of the power switch on the rear panel. The power supply module is a primary clocked switching power supply with Power Factor Correction (PFC) and Standby circuit (+12 V Standby).On the secondary side, it generates DC voltages (+3.3 V;...
Module Replacement R&S FSP Module Replacement This section describes the service concept and contains the spare parts list and the basic documents for the overall the R&S FSP instrument. Replacement of modules is described in detail in Section 3 under "Module Replacement".
R&S FSP Module Replacement Overview of the Modules Table 3-1 Overview - module replacement Module Required tests and adjustments after replacement Function tests and Adjustment Other system error correction Front module controller SYSTEM MESSAGES/ DOS/BIOS update SELFTEST / CAL Lithium battery SYSTEM MESSAGES/ Cold boot SELFTEST / CAL...
Module Replacement R&S FSP Performing a Cold Boot Connect power cable, switch on power switch. Simultaneously while switching on with the ON button, press and hold down the decimal-point key until the computer responds with a beep. Windows XP subsequently boots the instrument firmware. 1164.4556.82 3.14...
R&S FSP Module Replacement Replacing the Front Module Controller A90 (See chapter 5, spare parts list, item 577, and explosion drawings 1164.4391, 1164.4404.) The front module controller is mounted behind the front panel. Opening the Instrument and Removing the Front Panel Switch off the instrument and pull the mains plug.
Module Replacement R&S FSP Removing the Front Module Controller Unscrew the ten screws with washers of the front module controller board and take out the front module controller as follows (see Fig. 3-17) Note: The force to disconnect the controller connectors of the motherboard are very high. The pulling of the controller to the front side will be done with the help of the slits at the bottom side of the mounting plate.
R&S FSP Module Replacement Installing the New Front Module Controller and Completing the Instruments Carefully plug the new front module controller to the motherboard and fasten it using the ten screws with washers (590). Carefully insert the cable connectors to the controller board, taking care not to reverse the polarities. Fig.
Module Replacement R&S FSP Replacing the Lithium Battery on the Front Module Controller (See chapter 5, spare parts list, item 775, and explosion drawings 1164.4391, 1164.4404.) The lithium battery is accommodated on the front module controller board behind the front panel. CAUTION Lithium batteries must not be exposed to high temperatures or fire.
R&S FSP Module Replacement Removing the Lithium Battery Carefully push contact springs of the battery holder up and remove battery. Fig. 3-9 Position of lithium battery and jumper 12 on the front module controller 1164.4556.82 3.19...
Page 124
Module Replacement R&S FSP Installing the New Battery and Completing the Instrument Carefully push contact springs of the battery holder up and insert battery into the holder beneath the springs. Note: The plus pole (+) of the battery points upwards. WARNING Do not short-circuit the battery! Rotate the front panel into the instrument and screw with four countersunk screws (610) in the front...
R&S FSP Module Replacement Replacing Hard Disk A60 (See chapter 5, spare parts list, item 710, and explosion drawings 1164.4391 and 1164.4404.) The hard disk is incorporated between the front module and the analog boards. The spare part is already formatted for the R&S FSP and contains the complete software. Opening the Instrument and Removing the Hard Disk Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing the LCD and DC/AC Converter in the Front Module (See chapter 5, spare parts list, items 600 and 870, and explosion drawings 1164.4391, 1164.4404, 1093.4708.) The LCD is accommodated on the mounting plate together with the associated DC/AC converter. It is connected to the front module controller via cables, which can also be replaced individually.
R&S FSP Module Replacement Installing the New LCD and DC/AC Converter and Completing the Instrument Install the new LCD or new DC/AC converter in reverse order, fasten all screws, and connect and mount the corresponding cables. When mounting the display cable (947), make sure it is free of mechanical stress. Observe the following order when installing parts: 1) Before installing the display, connect the display cable (947) to the adapter board.
Module Replacement R&S FSP Replacing the Keyboard Membrane or Mat on the Front Module (See chapter 5, spare parts list, items 630 and 640, and explosion drawings 1164.4391, 1164.4404.) The keyboard membrane is the contact film for the rubber keys (mat) behind the labelling panel and the keyboard frame.
R&S FSP Module Replacement Installing the New Membrane and Completing the Instrument Put the new mat (630) into the keyboard frame (620). Note: The pins of the mat must be pressed in the wholes at the keyboard frame. Locate the new keyboard membrane (640) on the back of the mat (630) . Note: Push the foil cable of the membrane through the slit in the mounting panel.
Module Replacement R&S FSP Replacing the Labelling Panel on the Front Module (See chapter 5, spare parts list, items 270 and 280, and explosion drawing 1093.4495.) The labelling panel is the outer front panel which carries the labelling of all the parts on the front side of the R&S FSP.
R&S FSP Module Replacement Replacing the Floppy Disk Drive A30 (See chapter 5, spare parts list, item 670, and explosion drawings 1164.4391, 1164.4404.) Opening the Instrument and Removing the Floppy Disk Drive Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing the Power Supply A20 (See chapter 5, spare parts list, item 550, and explosion drawing 1164.4404). The power supply is fitted at the rear of the instrument frame of the R&S FSP. Removing the Power Supply Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
R&S FSP Module Replacement Replacing the Fan (See chapter 5, spare parts list, item 15, and explosion drawings 1164.4391.) The fan is fitted at the right side of the frame. Opening the Instrument and Removing the Fan Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing the RF Input Connector (Cable W1) (See chapter 5, spare parts list, items 295, 310, 315, 322, and explosion drawing 1164.4391.) The RF input connector is fitted at the right lower side of the front panel. Depending on the frequency range two different models of the connector are available.
R&S FSP Module Replacement Replacing the Motherboard A10 (See chapter 5, spare parts list, item 515, and explosion drawings 1164.4391, 1164.4404.) The motherboard is fitted from the bottom side. Opening the Instrument and Removing the Motherboard Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Page 136
Module Replacement R&S FSP Installing the New Motherboard and Completing the Instrument Insert the motherboard into the instrument in the reverse order. Note: Please move the motherboard carefully to the rear panel without causing damage to the parts on the board. Connect all cables (see label on cable for position).
R&S FSP Module Replacement Replacing the Module RF Attenuator A40 (See chapter 5, spare parts list, items 20, 30, 35, and explosion drawing 1164.4391.) The RF attenuator is fitted at the bottom side behind the RF input connector. There are different models of the attenuator for instruments up to 7 GHz (20), 30 GHz (30) and 40 GHz (35).
Module Replacement R&S FSP Replacing the Key/Probe Module A80 and Vol./Phone Board A191 (Option R&S FSP-B3) (See chapter 5, spare parts list, items 50 and 1040, and explosion drawings 1164.4391, 1129.6450.) The boards are fitted behind the front panel connectors Keyboard, Probe, Headphone and the volume control knob.
R&S FSP Module Replacement Installing the New Modules and Completing the Instrument Mount the new Key-Probe board (50) with three countersunk screws (55) to mounting plate (60.) Mount the new Vol./Phone Board (1040) with the nut of the volume control and one countersunk screw (1055) to the mounting plate (1050).
Module Replacement R&S FSP Replacing the RF Frontend A100 (See chapter 5, spare parts list, item 100, and explosion drawing 1164.4391.) The RF Frontend is installed in the middle of the instrument . Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
R&S FSP Module Replacement Replacing the Detector A120 or A140 (Option R&S FSP-B70) (See chapter 5, spare parts list, item 1100 or 1900, and explosion drawing 1164.4391.) The board is installed in the front part of the instrument. Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing the IF Filter A130 (See chapter 5, spare parts list, item 120, and explosion drawing 1164.4391.) The board is installed in the middle of the instrument. Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
R&S FSP Module Replacement Replacing the MW Converter Unit A160 (See chapter 5, spare parts list, items 130 to 141, and explosion drawing 1164.4391.) The board is installed in the middle of the instrument. Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing Diplexer A162; R&S FSP7 (See chapter 5, spare parts list, item 2030, and drawing 1130.2544.) The diplexer is installed on the 8 GHz converter unit (1130.2544).. Removing Diplexer on MW Converter 1130.2544 Remove 8 GHz converter unit (130). Note: See description "Replacing 8 GHz Converter Unit A160"...
R&S FSP Module Replacement Replacing 8 GHz YIG Unit A161; R&S FSP7 (See chapter 5, spare parts list, items 2005, 2010, and drawing 1130.2544.) The board is installed on the 8 GHz converter unit (1130.2544, 1130.2396. Note: Only the YIG units (with YIG filter, cable, sheet metal part) are replaced. The interface is thus always the same when using other YIG filters.
Page 146
Module Replacement R&S FSP Completing the Instrument and Frequency Response Correction Put 8 GHz converter unit (130) back into the instrument. Note: See description "Replacing MW Converter Unit A160". Execute frequency response correction software R&S FSP-FRQ.EXE. The software is located on the Gloris server. Test setup according to menu item "Schematic"...
R&S FSP Module Replacement Replacing Diplexer (30 GHz) A230; R&S FSP13/30 (See chapter 5, spare parts list, item 150, and drawing 1164,4391.01.) In R&S FSP13/30, the module is behind the attenuator on the right-hand side. Removing Diplexer Disconnect cable X71 from motherboard (515). Disconnect cables W2 and W30 from diplexer (150).
Module Replacement R&S FSP Replacing 13/30 GHz YIG Unit A161; R&S FSP13/30 (See chapter 5, spare parts list, items 2115, 2120, 2125, 2130, and drawing 1093.8249.) The board is installed on the 13/30 GHz converter unit (1093.8249). Note: Only the YIG units (with YIG filter, cable, sheet metal part) are replaced. The interface is thus always the same when using other YIG filters.
Page 149
R&S FSP Module Replacement Completing the Instrument and Frequency Response Correction Put the 13/30 GHz MW converter unit (135/140) back into the instrument. Note: See description "Replacing MW Converter Unit A160". Execute frequency response correction software R&S FSP-FRQ.EXE. The software is located on the Gloris server. Test setup according to menu item "Schematic"...
Page 150
Module Replacement R&S FSP Temperature Adjustment and Test of the YIG Filter A digital temperature adjustment is to be performed for the following MW converters: MW converter unit 13/30: 1093.8249 Mod: XX Rev: 10 SubRev: All other MW converters should undergo analog adjustment. Analog adjustment: Connect the RF output of the signal generator to the RF input of the R&S FSP.
R&S FSP Module Replacement Replacing RF Extension 13/30 A162; R&S FSP13/30 (See chapter 5, spare parts list, items 2140, 2145, and drawing 1093.8249.) The board is installed on the 13/30 GHz converter unit. Removing RF Extension 13 Remove 13 GHz MW converter unit (135). Note: See description "Replacing MW Converter Unit A160."...
Page 152
Module Replacement R&S FSP Completing the Instrument and Frequency Response Correction Put the 13/30 GHz MW converter unit (135/140) back into the instrument. Note: See description "Replacing MW Converter Unit A160". Execute frequency response correction software R&S FSP-FRQ.EXE. The software is located on the Gloris server. Test setup according to menu item "Schematic"...
R&S FSP Module Replacement Replacing Diplexer A163; R&S FSP40 (See chapter 5, spare parts list, item 2225, and drawing 1093.8584.) The board is installed on the 40 GHz converter unit (141). Removing the Diplexers on the 40 GHz MW Converter Remove 8 GHz converter unit (130).
Module Replacement R&S FSP Replacing 40 GHz YIG Unit A161; R&S FSP40 (See chapter 5, spare parts list, items 2215, 2217 and drawing 1093.8584.) The board is installed on the 40 GHz converter unit (141). Note: Only the YIG units (with YIG filter, cable, sheet metal part) are replaced. The interface is thus always the same when using other YIG filters.
Page 155
R&S FSP Module Replacement Temperature Adjustment and Test of the YIG Filter The MW converter 40 should undergo digital temperature compensation. Digital adjustment: Place R&S FSP40 into a climatic chamber for 2 hours at 50°C. Execute frequency response correction software R&S FSP-FRQ.EXE. The software is located on the Gloris server.
Module Replacement R&S FSP Replacing RF Extension 40 A162; R&S FSP40 (See chapter 5, spare parts list, item 2235, and drawing 1093.8584.) The board is installed on the 40 GHz converter unit. Removing RF Extension 40 Remove 40 GHz MW converter unit (141). Note: See description "Replacing MW Converter Unit A160".
R&S FSP Module Replacement Replacing AF Demodulator A190 (Option R&S FSP-B3) (See chapter 5, spare parts list, item 1000, and drawing 1164.4391.) The board is fitted the front part of the instrument. Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing the OCXO A200 (Option R&S FSP-B4) (See chapter 5, spare parts list, item 1100, and drawings 1164.4391 and 1129.6791.) The board is fitted in the front part behind the front module . Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
R&S FSP Module Replacement Replacing Trigger A230 (Option R&S FSP-B6) (See chapter 5, spare parts list, item 1800, and drawings 1164.4391 and 1129.8613.) The board is installed at the rear of the instrument in front of the power supply. Opening the Instrument and Removing the Board Switch off unit, disconnect it from power supply, unscrew 4 rear-panel feet (450) and pull off enclosure (410) towards rear.
Module Replacement R&S FSP Replacing the Tracking Generator A170 (Option R&S FSP-B9) (See chapter 5, spare parts list, item 1500, and drawings 1164.4391 and 1129.7069.) The board is fitted in the back part of the instrument in front of the power supply. Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
R&S FSP Module Replacement Replacing the External Generator Control A210 (Option R&S FSP-B10) (See chapter 5, spare parts list, item 1640, and drawings 1164,4391 and 1129.7317.) The board is fitted in the front part of the instrument . Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Page 162
Module Replacement R&S FSP Activate the generator as follows: Press PRESET Set FREQ – CENTER – 100 MHz Set SPAN 10 MHz Press NETWORK hotkey Press EXT SOURCE softkey Press GENERATOR softkey Select generator type Set IEC/IEEE? to TTL Set GPIB address to address of generator Select remote mode Press FREQUENCY SWEEP softkey: Set state to ON...
R&S FSP Module Replacement Replacing Wideband Calibrator A190 (Option R&S FSP-B15) (See chapter 5, spare parts list, item 1160, and drawing 1155.1012.) The board is installed in the front part of the instrument. Opening the Instrument and Removing the Board Switch off unit, disconnect it from power supply, unscrew 4 rear-panel feet (450) and pull off enclosure (410) towards rear.
Module Replacement R&S FSP Replacing the LAN Adapter (Option R&S FSP-B16) (See chapter 5, spare parts list, item 1270, and drawings 1164.4391 and 1129.8071.) The adapter is located at the rear of the instrument. Opening the Instrument and Removing the Board Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
R&S FSP Module Replacement Replacing Compact Flash Board S-ATA A60 (Option R&S FSP-B18) (See chapter 5, spare parts list, item 4900) Opening the Instrument and Removing the Compact Flash Board Switch unit off, disconnect it from AC supply, unscrew 4 rear-panel feet (450), and pull off enclosure (410) toward the rear.
Page 166
Module Replacement R&S FSP Installing the New Module and Completing the Instrument Mount new Compact Flash Board (4900) with 4 screws and washers (4930) to CF Card Holder (4920). Carefully insert CF Card Holder (4920) with Board (4900) from the top and fix it to adapter plate (4940) by means of 3 screws and washers (4950).
R&S FSP Module Replacement Replacing Flash Disk A60 (Option R&S FSP-B20) (See chapter 5, spare parts list, item 1702, and drawings 1164,4391 and 1155.1741.) The flash disk is located in the instrument between the controller panel and the modules. The spare part is supplied with pre-installed software.
Module Replacement R&S FSP Replacing External Mixer Board A180 (Option B21) (see chapter 5, spare parts list, item 2900, and illustrations 1164.4391 and 1157.1764.) The board is installed in rear part of the instrument. Opening Instrument and Removing Board Switch off instrument, disconnect from mains, unscrew 4 rear panel feet (450) and pull off enclosure (410) toward rear.
R&S FSP Module Replacement Replacing the Electronic Attenuator A50 (Option R&S FSP-B25) (See chapter 5, spare parts list, item 1400, and drawings 1164.4391 and 1129.7800.) The electronic attenuator is fitted behind the RF attenuator under the fan. Opening the Instrument and Removing the Electronic Attenuator Switch off the instrument, pull the mains plug, unscrew the 4 rear-panel feet (450) and push the tube (410) backwards and take off.
Module Replacement R&S FSP Replacing the DC/DC Converter Option R&S FSP-B30) (See chapter 5, spare parts list, item 3000.) The DC/DC converter (3000) is mounted on the rear panel of the instrument. See also DC Power Supply R&S FSP-B30 operating manual. CAUTION The DC/DC converter in an unmounted state is a module in "open frame"...
R&S FSP Module Replacement Replacing the Battery Charger Module (Option R&S FSP-B31/32) Opening the Battery Pack and Removing the Module Remove the battery pack from the instrument. Undo the three screws (1) on the bottom. Pull out the cover (2) to the rear. Unplug all cables.
Module Replacement R&S FSP Replacing the Batteries (Option R&S FSP-B31/32) Opening the Battery Pack and Removing the Batteries Remove the battery pack from the instrument. Undo the three screws (1) on the bottom. Pull out the cover (2) to the rear. Disconnect the batteries (3).
R&S FSP Troubleshooting Troubleshooting Malfunctions may have simple causes but also may be caused by faulty components. These troubleshooting instructions allow for locating the error causes down to board level and make the instrument ready for use again by means of board replacement. We recommend that the instrument be shipped to our experts in the service centers (cf.
Troubleshooting R&S FSP Troubleshooting Start-up Problems • Error: R&S FSP cannot be switched on. Action Possible error causes and further steps Check power-on switch on the rear Power switch OFF: Switch on power supply.. Check yellow LED (Stand-by). LED remains dark: Measure voltage at X20.D24 (power supply unit) Rated value: +12 V ±...
R&S FSP Troubleshooting Troubleshooting Problems with Boot Process • Error: R&S FSP does not start the measurement application. Following switch-on, the R&S FSP first boots the computer BIOS. After successful initialization of the computer the Windows XP operating system starts up. Subsequently, the test application is loaded as start-up program.
Page 176
Troubleshooting R&S FSP Normal action Error and error cause This procedure may be interrupted using the "BREAK" key on If this hard disk entry is missing, the connected external keyboard, any other key continues the the hard disk may be faulty boot process.
Page 177
R&S FSP Troubleshooting Normal action Error and error cause After this test, the BIOS has been loaded and the operating system The message „No System Disk is started. or Disk error ...“ at this point indicates that the contents of the hard disk correct.
Page 178
Troubleshooting R&S FSP Normal action Error and error cause If the operating system on the hard disk has been destroyed and cannot be loaded correctly, Windows XP reacts by displaying a "bluescreen". This bluescreen contains all essential information on the internal states of the computer which are displayed as follows (by way of example): Windows XP and the instrument...
Page 179
R&S FSP Troubleshooting Normal action Error and error cause If a "bluescreen" is displayed with loading, a cold start may be necessary. Proceed as follows in such a case: Cold boot (see section "Performing a Cold Boot" in this chapter) Firmware update from the backup partition, if a cold boot does not succeed.
Page 180
Troubleshooting R&S FSP Normal action Error and error cause When the program has been loaded, the measurement hardware is initialized first. A timer which is controlled by a 32-MHz clock signal If an error occurred on the is set on the detector board. This test reveals proper functioning of detector board or the clock is the detector board and the clock oscillator in the R&S FSP missing, the following message...
R&S FSP Troubleshooting Troubleshooting - Loading Module EEPROMs • Error: Data of modules cannot be read. Normal action Error and error cause On booting the instrument all calibration data required must be written into the RAM of the computer. The calibration data of a module is read from the EEPROM in the case of a cold start, otherwise from the associated binary file.
Page 182
Troubleshooting R&S FSP Normal action Error and error cause Upon successful reading of the module header from the EEPROM the contents of the module header is entered into the array for storing the module information. The module header read from the EEPROM is compared with the module header of the associated binary file.
R&S FSP Troubleshooting Troubleshooting via Selftest The selftest is provided for identification of instrument errors and tolerance violations which can not be corrected with self-calibration of the instrument. All signal paths are connected and the signal is traced via test points. The selftest checks all possible hardware settings which are used for the self-calibration with regard to sufficient setting range including reserves.
Page 184
Troubleshooting R&S FSP If the result is ***FAILED***, the service technician should pay attention not only to the entries marked with FAILED** but also to those marked with SKIPPED. These entries indicate that a test item was not performed because the preconditions were not observed.
Page 185
R&S FSP Troubleshooting Selftest Design and Error Messages Overview The calibration source on the IF-filter module is used as signal source for testing the signal path. 1. Measurement of the operating voltages a) of the power supply b) regulated operating voltages on the detector and IF-filter modules 2.
Page 186
Troubleshooting R&S FSP Operating Voltages When an operating voltage fails, the selftest is aborted to avoid subsequent error. The test is not aborted under service level 1. All subsequent errors are then listed in the result record. Errors which occurred independent of the voltage failure can then be detected, still. The error message indicates the error source (power supply, IF filter, detector board) and the voltage which failed.
Page 187
R&S FSP Troubleshooting Temperature Measurement on IF Filter Normal action Error and error cause The temperature is measured, first. If the temperature violates the permitted range from 0° to 70°C, the following warning is output: WARNING! IF-FILTER: Operating Temperature xx °C out of range Check the temperature data to plausibility.
Page 188
Troubleshooting R&S FSP Checking the 4-fold D/A Converter on the Detector Board. Normal action Error and error cause Detector Board Pretune-DAC Test The 4-fold D/A-converter on the detector board is checked. One D/A converter controls the frequency- dependant IF gain on the IF filter for frequency response correction.
Page 189
R&S FSP Troubleshooting Testing the Reference Signals on Standard Frontend (Modules 1 and 2) and on the IF Filter Module Errors detected do not cause abortion of the test. However, the subsequent checking of the signal path in the IF filter and detector boards is skipped, since there is no local oscillator provided to convert the 128-MHz calibration signal to the 20.4-MHz IF required for the tests.
Page 190
Troubleshooting R&S FSP Normal action Error and error cause Calibration signal level: FATAL ERROR! IF Board: CAL Signal level out of range for selftest Selftest aborted. The selftest of the signal path can not be performed with failure of the calibration signal. Switch on measurement via the INPUT CAL softkey in the SETUP - SERVICE menu.
Page 191
R&S FSP Troubleshooting Testing the Signal Paths Overview of the test sequence • The internal source for the test signal (CAL signal) has already been tested with testing the reference signals. • Testing all attenuators. • Testing the output level of the standard frontend using a level detector at the IF-filter input. •...
Page 192
Troubleshooting R&S FSP Signal Paths on the IF Filter Board The RF attenuator is used to provide an appropriate test level. Therefore, a fault in the RF attenuator will entail subsequent errors with the IF filter test. Normal action Error and error cause Input Level of IF Filter / Calibration Amplifier (CAL-Amps 1 + 2) Selftest detector A (see block diagram)
Page 193
R&S FSP Troubleshooting Normal action Error and error cause If the level adheres to the tolerance, a defective CAL_Amp may have caused the error. The setting ranges of the CAL_Amps are tested during the selftest, later. The selftest passes under service level 1. Note, whether the result file contains any error messages concerning the CAL_Amps.
Page 194
Troubleshooting R&S FSP Normal action Error and error cause Step Gain (IF Amplifier) Selftest Detector C (see block diagram) The 10-dB Step Gain (Step Gain Coarse) and the 0.1 dB Step Gain (Step Gain Fine) are tested. The input level is attenuated in steps of 10 dB by the RF attenuator and amplified by means of the Step Gain by the same amount, simultaneously.
R&S FSP Troubleshooting Troubleshooting Frontend 1 Depending on the kind of error a few measurements should be performed on the frontend prior to replacing the module: • IP3 too high • Signal level too low • LO feedthrough too high •...
Troubleshooting R&S FSP Troubleshooting – MW Converter The most frequent error occurring on MW converters is that the signal level displayed on the spectrum analyzer >3 GHz is too low or missing. • Level conditions on the MW converter Action Error and error cause Apply a high-frequency signal (f >3 GHz) at the If the level is missing or too low, a module is...
Page 197
R&S FSP Contents - Software Update/Installing Options Contents - Chapter 4 "Software Update/Installing Options" 4 Software Update / Installing Options..............4.1 Installation of new R&S FSP Software ...................4.1 Restoring the Operating-System Installation ................4.2 Installing the Options........................4.3 1164.4556.82 I-4.1...
Page 198
Contents - Software Update/Installing Options R&S FSP This page is left blank intentionally. 1164.4556.82 I-4.2...
R&S FSP Installation of new R&S FSP Software Software Update / Installing Options This chapter contains information on updating software, restoring the operating-system installation and installing options to the R&S FSP. Additional manuals obtained together with a software/firmware update or with subsequently acquired options can be filed here. Installation of new R&S FSP Software The installation of a new firmware version can be performed using the built-in diskette drive.
Restoring the Operating-System Installation R&S FSP Restoring the Operating-System Installation If the operating system can no longer be booted, it is possible to boot the analyzer in the boot menu from the backup partition, and to restore the operating-system installation. The function is not available for instruments equipped with option B20.
Page 201
R&S FSP Installing the Options Installing the Options The following options are available with the R&S FSP: AF Demodulator R&S FSP-B3 1129.6491.02 OCXO Reference R&S FSP-B4 1129.6740.02 TV Trigger / RF Power Trigger R&S FSP-B6 1129.8594.02 Tracking Generator R&S FSP-B9 1129.6991.02 External Generator Control R&S FSP-B10...
Installing the Options R&S FSP Switch on R&S FSP (cold boot). Install additional software, if supplied, according to the instructions enclosed with the option. If an adjustment is required for this option, the appropriate hints are to be found in the installation instructions for the option.
R&S FSP Contents - Documents Contents - Chapter 5 "Documents" 5 Documents......................5.1 Shipping of Instrument and Ordering of Spare Parts ..............5.1 Shipping of Instrument ......................5.1 Shipping of a Module........................5.1 Ordering Spare Parts .......................5.2 Refurbished Modules .......................5.2 Return of Defective Replaced Modules..................5.2 Spare Parts............................5.3 Available Power Cables......................5.3 Spare Parts List ..........................5.5...
Page 204
Contents - Documents R&S FSP This page is left blank intentionally. 1164.4556.82 I-5.2...
R&S FSP Shipping of Instrument and Ordering of Spare Parts Documents This chapter provides information on the ordering of spare parts and contains the spare parts list and the documents for the complete R&S FSP unit. Shipping of Instrument and Ordering of Spare Parts Please contact your Rohde &...
Shipping of Instrument and Ordering of Spare Parts R&S FSP Ordering Spare Parts To deliver replacement parts promptly and correctly we need the following indications: • Stock number (see component lists in this chapter) • Designation • Component number according to component list •...
R&S FSP Spare Part Spare Parts The stock numbers necessary for ordering replacement parts and modules can be found in the component lists further down. IMPORTANT NOTE When replacing a module please note the safety instructions and the repair instructions given in chapter 3 and at the beginning of this service manual When shipping a module be careful to provide for sufficient mechanical and antistatical protection.
Page 208
Spare Parts R&S FSP This page is left blank intentionally. 1164.4556.82...
Spare Parts List Mechanical Drawings 1164.4556.82...
Page 210
List of R&S FSP parts including spare parts R&S FSP This page is left blank intentionally. 1164.4556.82...
Page 211
R&S FSP List of R&S FSP parts including spare parts List of R&S FSP parts including spare parts The R&S FSP is constructed in accordance with R&S design 2000. Overall dimension: W x H x L, 372.75 x 176.50 x 395.00 Rack mount: 4E 7/8 T350 Accessories:...
Page 212
List of R&S FSP parts including spare parts R&S FSP Electrical Recommended Position Designation Stock No. Number designation spare parts COVER RD15.9 0009.9200.00 FRONTEND 1093.5491.07 A100 DETECTOR BOARD 1093.6998.07 A120 IF-FILTER 1093.7242.04 A130 CONVERTER UNIT (8 GHZ) 1130.2544.02 A160 MW CONVERTER UNIT (13 GHZ) 1164.4891.13 A160 MW-CONVERTER UNIT (MOD31)
Page 213
R&S FSP List of R&S FSP parts including spare parts Electrical Recommended Position Designation Stock No. Number designation spare parts LABEL 32 1164.4691.00 FRONT COVER DESCR. 40 GHz 1164.4504.00 FRONT COVER DESCR. MOD31 1164.4510.00 FRONT COVER DESCR. MOD38 1164.4527.00 RF CABLE W1 7 GHZ 1093.4872.00 CABLE CLAMP RD4 BR6 0080.3660.00...
Page 214
List of R&S FSP parts including spare parts R&S FSP Electrical Recommended Position Designation Stock No. Number designation spare parts Drawing 1164.4404.01 (Digital Unit) INSTRUMENT FRAME 1093.4543.00 PROTECTION 0001.9447.00 0.06 M DIN965-M2.5X6-A4-PA 1148.3288.00 MOTHERBOARD 1164.5281.02 SCREW WITH WASHER ASSEMBLY M2.5X6 1148.3059.00 LOCKING BOLT M3 0009.6501.00...
Page 215
R&S FSP List of R&S FSP parts including spare parts Electrical Recommended Position Designation Stock No. Number designation spare parts CF-CARD HOLDER CLIP 1130.1260.00 SATA DATA CABLE 265 1091.3440.00 SATA POWER CABLE 265 1091.3427.00 DISK BRACKET 1093.4837.00 DIN965/ISR-M2.5X6-A4-PA 1148.3288.00 DIN965-M3X5-A4-PA 1148.2775.00 LITHIUM BATTERIE CR2032 0858.2049.00...
Page 216
List of R&S FSP parts including spare parts R&S FSP Electrical Recommended Position Designation Stock No. Number designation spare parts 2010 8 GHZ YIG UNIT (FILTRONIC) 1130.2944.03 A161 2025 DIN965/ISR-M2.5X6-A4-PA 1148.3288.00 2030 8 GHZ DIPLEXER 1132.6501.02 A162 2035 DIN6900-M2.5X5 -A2 0071.6830.00 Drawing 1093.8584.01 (UNIT MW CONVERTER 40 GHz) 2210...
Page 217
R&S FSP List of R&S FSP parts including spare parts Electrical Recommended Position Designation Stock No. Number designation spare parts 4032 RF EXTENSION 13 GHz 1108.6505.13 A162 4040 SCREW WITH WASHER M2.5X6-A2 1148.3059.00 4050 RELAIS HOLDER 1164.4933.00 4060 KOAX-REL.3xK-BU 40 GHz 1104.2386.00 4070 RELAIS CABLE...
Page 218
List of R&S FSP parts including spare parts R&S FSP Electrical Recommended Position Designation Stock No. Number designation spare parts 1055 SCREW WITH WASHER ASSEMBLY M2.5X6 1148.3059.00 1060 DIN965/ISR-M2.5X6-A4-PA 1148.3288.00 1070 ROTARY KNOB RD9.5ACHS-RD4T-GR 0852.1111.00 1086 GROUND SPRING 1142.8242.00 Drawing 1129.6791.00 (Option R&S FSP-B4) 1100 OCXO 1093.7871.06...
Page 219
R&S FSP List of R&S FSP parts including spare parts Electrical Recommended Position Designation Stock No. Number designation spare parts 1622 LOCKING BOLT M3 0009.6501.00 1624 DIN137-A3-A2 0005.0296.00 1626 DIN934-M3-A4 0016.4398.00 1640 IEC BUS CONTROLLER 1164.5200.02 A210 1650 ADAPTER PLATE FOR R&S FSP-B10 1164.5275.00 1665 SCREW WITH WASHER M2.5X6-A2...
Page 220
List of R&S FSP parts including spare parts R&S FSP Electrical Recommended Position Designation Stock No. Number designation spare parts Drawing 1155.1764.00 (Option R&S FSP-B21 1155.1758.02) 2900 EXT MIXER AMP 1157.1126.02 A180 2910 RF CABLE W36 1155.1770.00 2920 RF CABLE W37 1155.1787.00 2930 RF CABLE W38...
Page 221
R&S FSP List of R&S FSP parts including spare parts Electrical Recommended Position Designation Stock No. Number designation spare parts Drawing chapter 3 (Option R&S FSP-B31/B32 1155.1258.02/1155.1506.02) 3030 BATTERY CHARGER 1155.1358.02 3040 13.2 V 7.6 AH AKKU PACK 1102.5607.00 G1, G2 3050 TABLE POWER SUPPLY 70 W 4055.3471.00...
Page 222
List of R&S FSP parts including spare parts R&S FSP 1164.4556.82 5.18...