Initialization
The Setup the communications settings option presents the Setup DSX dialog box
(Figure 4-3), which is used to select the desired communications port. In most cases, the
problem is that you have not connected the DSX™ Automated ELISA System to the indicated
communications port on the personal computer. If this is the problem, select the correct port
(at this point, do not worry about the rest of the tabs). In almost all cases, the system should
be connected to COM1.
Figure 4-3 Setup DSX Dialog Box-RS-232 Tab
4.2.5 Incorrect Password
The user is prompted to enter a password whenever the application program is opened (when
the instrument is shipped, the default password is Dynex). The operator can establish a new
password via the Password command on the Tools menu
If the user loses or forgets the password, it will not be possible to employ the program (nor
will it be possible to copy the password or edit it). The program will continued to be stored in
memory.
If necessary, the system can be started by deleting the hidden file Password.Pas from the
Revelation directory and re-starting the program. When the password prompt is presented,
enter Dynex or Thermo (depending on the version). The program will then operate in the
normal fashion, and a new password can be entered.
Note: This information must not be provided to end-users.
DSX™ System Service Manual
4-5