5. Error Codes & Diagnostics

5.1 Error Codes

Error Code
Meaning
Err.01
Entered PIN is not correct
Err.02
PIN mismatch during PIN setting –
second PIN does not match first PIN
Err.03
Invalid values entered (e.g calibration
attempt beyond sensor range)
Err.04
User entered value is not valid (e.g date
dd.mm = 31.02, or a user scaling factor
= 0.0)
Err.05
Invalid calibration point applied during
user calibration
Err.06
User attempt to zero a sensor failed
Err.07
Calibration attempted without valid
date set
Err.21
Remote Sensor Error (pressure or RTD
Err.23
sensor)
Err.24
Err.40-63
Internal error
Err.64
User calibration rejected
Err.80
Critical Software Fault
Solution/Action
Press the
(OK) key and enter correct pin, or
press the
(Back) key to exit the function.
Press the
(OK) key and enter correct pin, or
press the
(Back) key to exit the function.
Press the
(OK) key and use valid calibration
points, or press the
function.
Press the
(OK) key and perform valid
calibration, or press the
the function.
Press the
(OK) key and enter valid data
range, or press the
function.
Press the
(OK) key and enter valid data
range, or press the
function.
Set the date/time (this is internally battery-
backed)
Disconnect and reconnect remote sensor – if
error persists then sensor is either incompatible
type with DPI 705E / DPI 705-IS or is faulty.
Internal fault - return unit to service center.
Press the
(OK) key or the
to return to measurement screen.
Power off and back on. If problem persists return
unit to service center.
Copyright 2019 Baker Hughes, a GE company, LLC ("BHGE") — All rights reserved.
137M7122 DPI 705E / DPI 705E IS User Manual–English | 37
(Back) key to exit the
(Back) key to exit
(Back) key to exit the
(Back) key to exit the
(Back) key

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DPI 705E and is the answer not in the manual?

Questions and answers

James Longmuir
March 3, 2025

My Druck gauge went into alarm and I can’t stop it now. I’ve tried to switch alarm perimeters off but still it flashes with alarm icon showing. I’ve switched it off and on and still in alarm

1 comments:
Mr. Anderson
March 3, 2025

To stop the alarm on your Druck DPI 705E gauge:

1. Ensure the sensor reading screen is displayed.
2. Press the designated keys simultaneously.
3. The current alarm state will be shown as “ ” (On) or “ ” (Off).
4. Ensure the alarm is set to OFF to stop it.

This answer is automatically generated

This manual is also suitable for:

Dpi 705e-is

Table of Contents

Save PDF