Druck DPI 705E User Manual

Druck DPI 705E User Manual

Digital pressure indicator
Hide thumbs Also See for DPI 705E:

Advertisement

DPI 705E/705E-IS
Digital Pressure Indicator
User Manual
Druck.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DPI 705E and is the answer not in the manual?

Questions and answers

James Longmuir
March 3, 2025

My Druck gauge went into alarm and I can’t stop it now. I’ve tried to switch alarm perimeters off but still it flashes with alarm icon showing. I’ve switched it off and on and still in alarm

1 comments:
Mr. Anderson
March 3, 2025

To stop the alarm on your Druck DPI 705E gauge:

1. Ensure the sensor reading screen is displayed.
2. Press the designated keys simultaneously.
3. The current alarm state will be shown as “ ” (On) or “ ” (Off).
4. Ensure the alarm is set to OFF to stop it.

This answer is automatically generated

Summary of Contents for Druck DPI 705E

  • Page 1 DPI 705E/705E-IS Digital Pressure Indicator User Manual Druck.com...
  • Page 3 DPI 705E instrument. The RTD-INTERFACE is a remote interface for use with the DPI 705E or the DPI 620G to allow connection of a PT 100 RTD probe to the instrument for the purpose of measuring temperature.
  • Page 4 Commercial and Intrinsically Safe Product versions The DPI 705E is available as either non-Intrinsically Safe (blue in color) or Intrinsically safe (yellow in color) versions, with case coloring used as well as labeling to indicate the version. The remote pressure sensor PM 700E and the remote RTD interface RTD-INTERFACE are available as either non-Intrinsically Safe or Intrinsically safe versions.
  • Page 5 Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | iii...
  • Page 6 External Sensors The DPI 705E supports two types of external sensor, both of which connect via a cable into the right- hand side of the unit. External sensor connector The sensor must be connected via supplied connection cable (it will not engage correctly if directly plugged in).
  • Page 7 Also check (Press Simultaneously) sensor full-scale values, system date/time, software version and other General settings. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | v...
  • Page 8 15. Sensor Type Absolute/Gauge Indicator 16. Minimum Symbol Indicator 17. Negative Value Indicator 18. Maximum Symbol Indicator Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. vi | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 9: Table Of Contents

    3.2.7 Time 3.2.8 Custom Unit 3.2.9 Software Version Accessories Error Codes & Diagnostics Error Codes Diagnostics Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | vii...
  • Page 10 Approved Service Centers Return Goods/Material Procedure Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. viii | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 11: Basic Functionality

    The positive full-scale of the sensor is briefly displayed. The number of days remaining to the next calibration due date are then briefly displayed. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 1...
  • Page 12: Pressure Measurement

    The sensor reading screen is then displayed as below. This is the default state or main screen of the instrument. 1.2 Pressure Measurement The DPI 705E can be used to measure pressure from an external pressure source - it does not generate pressure itself. To measure pressure: Connect an external pressure source to be measured via the P1 pressure port at the top of the instrument.
  • Page 13: Zero

    RTD-Interface. To perform a zero: Open all pressure ports to atmospheric pressure. Press the keys together. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 3...
  • Page 14: Tare

    The filter symbol will disappear. 1.7 UNITS To change the pressure units of the display: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 4 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 15: Backlight

    The backlight should now remain off (regardless of key press). To return the backlight to default mode short press the backlight key. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 5...
  • Page 16: Advanced Functionality

    The minimum reading is the lowest read by the instrument since power up. The value will be reset each time the device is turned off. To view / reset the minimum pressure: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 6 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 17: Leak Test

    Leak test can also be used with the RTD-INTERFACE to record temperature change over time. To perform a leak test, follow the sequence of steps below: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 7...
  • Page 18: Alarm

    The alarm feature allows an alarm to be set based on certain conditions. For example, pressure value above or below user set limits. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 8 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 19: Set High Alarm

    To set a high alarm: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 9...
  • Page 20 When all digits have been processed, press the key to save the value. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 10 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 21: Set Low Alarm

    The dash above the arrow indicates that this is a negative value. To change this value, press the key. The 'current digit' will flash briefly. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 11...
  • Page 22 If this happens press the key and enter the value again. Press the key to go to the main screen. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 12 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 23: Settings

    Factory Reset Custom Unit Setup* Software Version *Not Available with RTD-INTERFACE There are two categories of Settings within DPI 705E, General Settings and Advanced Settings. To access these settings: Ensure the sensor reading screen is shown as per below. Press the keys simultaneously.
  • Page 24: Advanced Settings

    ” (Calibration Point 1) near zero and “ ” (Calibration Point 2) near the full-scale value of the sensor. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 14 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 25 If the PIN is correct “ ” (Adjust) is shown on the screen as per below. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 15...
  • Page 26: Calibrate Sensor

    Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 16 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 27 “ ” (Advanced) screen. Press the key again to return to the sensor reading screen. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 17...
  • Page 28: Set Calibration Date

    (6 months) and so on. The calibration interval determines the next required calibration date of the sensor. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 18 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 29 Press the key to return to the Advanced Settings screen. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 19...
  • Page 30: Change User Pin

    If the pin is accepted, the user will be redirected back to the “ ” (Pin) screen as shown above. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 20 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 31: Units Lock

    4 times to navigate to the “ ” (Units) screen as shown below. Press the key to enter the Units setting screen. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 21...
  • Page 32: Lock Display

    Enter Pin. The steps to enter Pin are indicated in above Section 3.1.1.1. If the Pin is correct, the display will show “ ” (Adjust). Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 22 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 33: Factory Reset

    Advanced Settings. Enter Pin. The steps to enter Pin are indicated in above Section 3.1.1.1. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 23...
  • Page 34: General Settings

    Calibration Due Date Displays next calibration date of the sensor Audio Turn audio on or off Auto Shutdown Enable or disable auto-shutdown Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 24 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 35: Full-Scale Value

    3.2.2 Calibration Due Date The calibration due date is the date when the sensor is next due for calibration. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 25...
  • Page 36: Audio

    The display will show “ ” (Advanced) which indicates that the unit is now in the Settings option. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 26 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 37: Auto Shutdown

    3.2.4 Auto shutdown This feature will allow control of the auto shutdown behavior of the instrument. By default, the DPI 705E is configured to shut down after 10 minutes of no activity (i.e. no key presses). To change this setting: Ensure sensor reading screen is shown per below.
  • Page 38: Battery Type

    The latter two are both Nickel based chemistry and have a single setting type “ ” (Battery.Nickel). Note: The Intrinsically Safe DPI 705E-IS can only be used with the first two battery technologies listed (Al and LiFeS) - refer to safety manual. To change the battery type: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) —...
  • Page 39: Date

    3.2.6 Date This setting changes the system date. To change the date: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 29...
  • Page 40 Note: Only the last 2 digits of the year can be changed. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 30 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 41: Time

    The display will show “ ” (Advanced) which indicates that the unit is now in the Settings option. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 31...
  • Page 42: Custom Unit

    The scaling factor will only be applied when the custom unit is selected during pressure measurement. To set a custom scaling factor: Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 32 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 43 • The Default scaling factor for whichever units the device is currently set to is 1.0000 Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 33...
  • Page 44: Software Version

    To see the application software version: Ensure the sensor reading screen is shown as per below. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 34 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 45 Press the key again to return to the sensor reading screen. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 35...
  • Page 46: Accessories

    4. Accessories The DPI 705E / DPI 705E-IS is compatible with the following accessories. All accessories are compatible with either the Intrinsically Safe or Non-Intrinsically Safe versions of the product unless otherwise specified. Part Code Description IO-RTD-M12CON Field wireable M12 connector 4-pin to fit RTD-INTERFACE IO-RTD-M12EXT M12 male to female extension lead 2 m (6.6 ft) 4-wire...
  • Page 47: Error Codes & Diagnostics

    Critical Software Fault Power off and back on. If problem persists return unit to service center. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 37...
  • Page 48: Diagnostics

    6. Approved Service Centers 6.1 Return Goods/Material Procedure If the unit requires calibration or is unserviceable, return it to the nearest Druck Service Centre listed Druck.com/essential Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved.
  • Page 49 Serial number. • Details of defect/work to be undertaken. • Calibration traceability requirements. • Operating conditions. Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 137M7122 DPI 705E / DPI 705E IS User Manual–English | 39...
  • Page 50 Copyright 2019 Baker Hughes, a GE company, LLC (“BHGE”) — All rights reserved. 40 | 137M7122 DPI 705E / DPI 705E IS User Manual–English...
  • Page 52 BHGE. Baker Hughes reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your BHGE representative for the most current information. Druck and Druck Logo are registered trademarks of BHGE in the United States and other countries.