15.1 Sub-tank Section
3. Remove the screw and then the solution
thermometer.
Installation
Installation is essentially in the reverse order of removal.
15.1.5 Solution Level Sensor (FS700 to FS705) Replacement
Removal
1. Remove:
• Upper connector cover
• PS1/PS2/PS3: Processor inner cover
.
(See Subsection 15.1.3)
2. Open the five clamps and disconnect the sensor
connector.
3. Remove the screw and then the sensor.
Installation
Installation is essentially in the reverse order of removal.
Distributed by minilablaser.com
.
(See Subsection 15.1.2)
15-5
Screw
Solution Thermometer (TS700)
Clamps (5)
Sensor Connector (FS704)
Sensor Connector (FS703)
Sensor Connector (FS702)
Sensor Connector (FS701)
Sensor Connector (FS700)
Sensor (FS700)
LII646
Sensor Connector
(FS705)
LII647
Screw
LII648
15