Table of contents Dentsply Sirona Operating Instructions Sinius Table of contents General information ....................Dear Customer,.................... Contact information..................Notes on these Operating Instructions............1.3.1 General information on the operating Instructions ......1.3.2 Scope of these Operating Instructions........... Other valid documents .................
Page 3
Dentsply Sirona Table of contents Operating Instructions Sinius Headrest ...................... 3.5.1 Motor-driven headrest..............3.5.2 Double-jointed head support............Foot control ....................Dentist element .................... 3.7.1 Instrument positions............... 3.7.2 EasyTouch user interface .............. 3.7.3 Touchscreen .................. 3.7.4 Fixed keys on the dentist element ..........
Page 4
Table of contents Dentsply Sirona Operating Instructions Sinius 4.4.3 Triggering an immediate movement stop........4.4.4 Armrests..................4.4.5 Vario footrest.................. 4.4.6 Adjusting the motor-driven headrest ..........4.4.6.1 Moving the headrest in/out ........... 4.4.6.2 Inclining the headrest ........... 4.4.7 Adjusting the double articulating headrest ........
Page 5
Dentsply Sirona Table of contents Operating Instructions Sinius 4.5.8 General instrument functions ............4.5.8.1 Opening the sub-screen ..........4.5.8.2 Selecting a coolant ............4.5.8.3 Switching the preselected coolant on/off ...... 4.5.8.4 Setting the ApexLocator ..........4.5.8.5 Switch instrument light on/off........
Page 6
Adjusting the water amount for flushing......... 160 4.7.4 Self-sufficient water supply ............161 Tray......................164 4.8.1 Tray on the support arm (for Sinius) ..........164 4.8.2 Tray on the Sinius CS dentist element........... 165 4.8.3 Tray on the Sinius TS dentist element ........... 166 4.8.4...
Page 7
Dentsply Sirona Table of contents Operating Instructions Sinius 4.12.2.3 Connecting the SiroCam AF / AF+ intraoral camera ..176 4.12.2.4 Operating the SiroCam AF / AF+ intraoral camera ..177 4.12.2.5 Technical camera data ..........182 4.13 Operation with a PC..................183 4.13.1...
Page 8
Table of contents Dentsply Sirona Operating Instructions Sinius 4.14.5.3 Showing/hiding the key for the external HF surgical unit................199 4.14.5.4 Setting the spray temperature ........199 4.14.6 Configuring the network connection..........200 4.14.7 Opening the service function ............200 Care, cleaning and maintenance by the practice team........... 201 Basics ......................
Page 9
Dentsply Sirona Table of contents Operating Instructions Sinius 5.3.4.4 Cleaning/disinfecting the SiroCam AF / AF+ intraoral camera................227 5.3.4.5 Check the flow rate on the Sprayvit M multifunctional syringe ................227 5.3.5 Changing the cotton wool roll on the high-speed handpiece hose and oil collector ................
Page 10
Table of contents Dentsply Sirona Operating Instructions Sinius Malfunctions......................267 Error messages.................... 267 Remote diagnosis ..................269 Spare parts and consumables ................270 Disposal ........................272 Overview of all function keys .................. 273 Index ........................292 63 2 2 7 26 D3561...
These operating instructions are designed to assist you prior to initial use and whenever you require information later on. We wish you a great deal of success and pleasure with Sinius. Your Sinius team Contact information Customer service center...
System versions These operating instructions apply to the following treatment centers: ● Sinius (Dentist element with sliding track) ● Sinius CS (Dentist element with swivel arms) ● Sinius TS (Dentist element as OTP system) Equipment options This document describes the full version of your system. It may therefore cover components that are not included in the system you purchased.
Exclusion of liability In the event that the system owner fails to fulfill its obligation to perform maintenance activities or ignores error messages, Dentsply Sirona or your authorized dealer cannot assume any liability for any damage thus incurred.
1 General information Dentsply Sirona 1.7 Formats and symbols used Operating Instructions Sinius Formats and symbols used The formats and symbols used in this document have the following meaning: Prerequisite Prompts you to do something. 1. First action step 2. Second action step ➢...
Dentsply Sirona 2 Safety instructions Operating Instructions Sinius 2.1 Identification of the danger levels Safety instructions Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions.
2 Safety instructions Dentsply Sirona 2.3 On-site installation Operating Instructions Sinius On-site installation The on-site installation must have been performed according to our requirements. The details are described in the document "Installation Requirements". Installation of the treatment center Installation must be carried out by authorized personnel according to the installation instructions.
Dentsply Sirona 2 Safety instructions Operating Instructions Sinius 2.6 Connection to the public drinking water system Connection to the public drinking water system Treatment center isolated from the public drinking water supply The treatment center, if it is equipped with a disinfection system, complies with the requirements of EN 1717 (free discharge with an isolation distance ≥...
No restrictions apply when using salt blasting devices in connection with Dentsply Sirona treatment centers. However, in such cases, make sure that the system is subsequently flushed with an adequate amount of water.
For reasons of product safety, this product may be operated only with original Dentsply Sirona parts or third-party parts expressly approved by Dentsply Sirona. In the event of changes which were not foreseen, Dentsply Sirona is not liable for resulting damages.
2 Safety instructions Dentsply Sirona 2.17 Electromagnetic compatibility Operating Instructions Sinius Electromagnetic compatibility 2.17 Medical electrical devices are subject to special precautionary measures with regard to electromagnetic compatibility (EMC). They must be installed and operated as specified in the document “Installation Requirements”.
Page 21
Dentsply Sirona 2 Safety instructions Operating Instructions Sinius 2.18 Electrostatic discharge ● discharging the electrostatic charges from your own body through contact with – a metallic unit casing – a larger metallic object – any other metal part grounded with the protective earth ●...
2 Safety instructions Dentsply Sirona 2.19 Dismantling/Installation Operating Instructions Sinius Dismantling/Installation 2.19 When dismantling and reinstalling the unit, proceed according to the installation instructions for new installation in order to guarantee its proper functioning and stability. 63 2 2 7 26 D3561...
Operating Instructions Sinius 3.1 Standards/Approvals Unit description Standards/Approvals The Sinius treatment center complies with the following standards, among others: ● IEC 60601-1 (electrical and mechanical safety plus software reliability) ● IEC 60601-1-2 (electromagnetic compatibility) ● IEC 60601-1-6 / IEC 62366 (usability) ●...
Page 24
3 Unit description Dentsply Sirona 3.1 Standards/Approvals Operating Instructions Sinius This unit meets the requirements of BELGAQUA and may therefore be connected to the public drinking water supply in Belgium. This unit meets the requirements of ATS and may therefore be connected to the public drinking water supply in Australia.
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.2 Technical data Technical data Model designation: Sinius / Sinius CS / Sinius TS Power connection: 100 – 240 V AC ± 10% 50/60 Hz Rated current: 2.2 A at 240 V 4.35 A at 115 V 5.0 A at 100 V also max.
Page 26
3 Unit description Dentsply Sirona 3.2 Technical data Operating Instructions Sinius 1060 Transport and storage con- Temperature: -40°C – +70°C ditions: (-40°F – 158°F) Relative humidity: 10% – 95% Air pressure: 500 hPa – 1060 hPa 1060 Operating conditions: Ambient temperature: 10°C – 40°C (50°F –...
Page 27
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.2 Technical data Foot switch wireless interface The wireless foot control can be equipped with various wireless modules. A label with the name of the wireless module used is on the underside of the foot control.
Dentsply Sirona 3.3 System overview Operating Instructions Sinius System overview The Sinius, Sinius CS, and Sinius TS treatment centers comprise the following main components: Sinius treatment center (sliding track device) Patient chair [→ 31] Motor-driven headrest [→ 32] (shown here) or Double articu- lating headrest ...
Page 29
Motor-driven headrest [→ 32] (shown here) or Double articu- lating headrest [→ 33] Foot control [→ 34] (with cable or wireless link) Dentist element Sinius CS with swivel arms [→ 35] Assistant element [→ 45] Water unit [→ 48] External device connection [→ 50] and power switch 63 ...
Page 30
Motor-driven headrest [→ 32] (shown here) or Double articu- lating headrest [→ 33] Foot control [→ 34] (with cable or wireless link) Sinius TS dentist element as OTP system [→ 35] Assistant element [→ 45] Water unit [→ 48] External device connection [→ 50] and power switch 63 ...
Backrest Seat Armrest Footrest Chair base Flange on the Sinius dentist element sliding track 4-way foot switch Foot control cable port Rotary joint for assistant element The patient chair can come with lounge upholstery featuring enhanced comfort and double seams. The lounge upholstery has no footrest; the surface is padded throughout.
3 Unit description Dentsply Sirona 3.5 Headrest Operating Instructions Sinius Headrest 3.5.1 Motor-driven headrest The headrest allows for the following adjustment options: ● Motor-driven extension/retraction to adapt to the patient's stature ● Motor-driven tilting for maxillary/mandibular treatment ● Manual tilting via quick mechanical adjustment ●...
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.5 Headrest 3.5.2 Double-jointed head support The double articulating headrest is equipped with two rotary joints. They allow the head inclination to be manually adjusted over a wide range for maxillary/mandibular treatment. The headrest extension can be pushed in or pulled out to manually adjust it to the height of the patient.
3 Unit description Dentsply Sirona 3.6 Foot control Operating Instructions Sinius Foot control The foot control enables hand-free control of the treatment instruments. Via the integrated cursor control, virtually all functions of the treatment center can be controlled via the foot control as an alternative to hand control.
This operating manual is valid for the following treatment centers: ● Sinius (Dentist element with sliding track) ● Sinius CS (Dentist element with swivel arms) ● Sinius TS (Dentist element as OTP system) Sinius dentist element The dentist element can be moved the entire length of the patient chair along the sliding track.
Page 36
3 Unit description Dentsply Sirona 3.7 Dentist element Operating Instructions Sinius Additional holder for intraoral camera Support arm, height adjustable by service engineer Slide of sliding track 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Page 37
3.7 Dentist element Sinius CS dentist element The Sinius CS dentist element is attached to the water unit with a flexible support arm. The dentist element is held in place at the adjusted height with a pneumatic locking brake in the support arm. To release the brake simply touch one of the two handles.
Page 38
Operating Instructions Sinius Sinius TS dentist element The Sinius TS dentist element is attached to the water unit with a flexible support arm. The dentist element is held in place at the adjusted height with a pneumatic locking brake in the support arm. To release the brake simply touch one of the two handles in the upper area.
Page 39
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.7 Dentist element Main switch Skid-proof silicone mat Removable NaCI bottle holder Additional holder for intraoral camera Support arm 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
SiroCam AF or handpiece handpiece handpiece SiroCam AF+ intraoral camera SiroSonic TL scaler A maximum of one SiroSonic TL scaler can be connected. Only the Sinius and Sinius TS dentist elements can be equipped with an additional holder. 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Operating Instructions Sinius 3.7 Dentist element 3.7.2 EasyTouch user interface User interface of the Sinius and Sinius TS dentist element User interface of the Sinius CS dentist element Fixed keys for opening the Start, Instrument or Sivision screen (membrane keyboard)
3 Unit description Dentsply Sirona 3.7 Dentist element Operating Instructions Sinius 3.7.3 Touchscreen The touchscreen displays virtual function keys according to the program selected. A list of all function keys is provided in the Appendix of this document, see "Overview of all function keys" [→ 273]..
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.7 Dentist element 3.7.4 Fixed keys on the dentist element For a more detailed description of the fixed key functions, see "Fixed keys on the dentist element" [→ 85]. Main switch Switches the treatment center on/off.
Page 44
3 Unit description Dentsply Sirona 3.7 Dentist element Operating Instructions Sinius Flushing Starts or stops cuspidor flushing. When the Flushing key is pressed (> 2 s), the Flushing Time settings screen appears. Freely selectable function e.g., call key Freely available relay 230 VAC, 6 A (connected by the service engineer).
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.8 Assistant element Assistant element The functional scope of the assistant element is adapted to the dental assistant's field of activity. It can, however, also be positioned so as to enable unassisted treatment by the dentist.
SiroCam AF or SiroCam AF+ intraoral camera Second spray aspirator or saliva ejector for surgical suction Only the Sinius CS and Sinius TS treatment centers can be equipped with an intraoral camera on the assistant element. 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.8 Assistant element 3.8.2 User interface There are three fixed keys located on each side, to the left and right of the assistant element. 3.8.3 Fixed keys on the assistant element For a more detailed description of the fixed key functions, see "Fixed keys on the assistant element" ...
3 Unit description Dentsply Sirona 3.9 Water unit Operating Instructions Sinius Water unit The water unit can be optionally equipped with a disinfection system. In normal operation, this will automatically inoculate the water that comes into contact with the patient (also called treatment water) with an agent to disinfect the water paths.
Page 49
Cover of the storage tank for the disinfectant used to disinfect the water paths or for the stand-alone water supply Mount for support arm of operating light and monitor Mount for Sinius CS and Sinius TS dentist elements or tray support arm Mount for water-carrying instruments of the dentist element...
3 Unit description Dentsply Sirona 3.10 External device connection Operating Instructions Sinius External device connection 3.10 External medical accessories can be connected to the external device connection. They must comply with the requirements of the 93/42/EEC Medical Devices Directive. NOTE...
Page 51
Dentsply Sirona 3 Unit description Operating Instructions Sinius 3.10 External device connection Pressure Flow rate Water 2.2 ± 0.2 bar max. 300 ml/min 4.4 ± 0.5 bar max. 70 Nl/min IMPORTANT The removal of media at the external device connection can reduce the performance of integrated consumers, e.g., the filling quantity of...
4 Operation Dentsply Sirona 4.1 Starting up the treatment center Operating Instructions Sinius Operation Starting up the treatment center 4.1.1 Initial startup Sanitization must be performed prior to initial startup of your treatment center. For sanitizing, the water-carrying lines are filled with the undiluted disinfectant to disinfect the water paths in order to reduce the bacterial load in the water paths.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.1 Starting up the treatment center 4.1.2.2 Main switch Switching the treatment center on The main switch switches the treatment center from Standby mode to operational readiness. Following switch-on, the operating system is booted and an automatic self-test is performed.
Ä The user profile that is displayed in the status bar is active. Tip: The individual user settings can be read out by the service engineer and transferred to other Sinius treatment centers. In this case, the user profiles only have to be set once.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.2 Control concept of touchscreen Control concept of touchscreen 4.2.1 Virtual function keys The touchscreen displays virtual function keys according to the program selected. Required functions can be activated either by touching the function keys with your finger or via the cursor with the foot switch.
4 Operation Dentsply Sirona 4.2 Control concept of touchscreen Operating Instructions Sinius 4.2.2 Start program operating modes After the treatment center is switched on, the Start program automatically appears. The start program can run in two operating modes. In the Standard Start program operating mode, the functions of the patient chair and instruments are each shown on separate screens.
Page 57
The most important function keys for the treatment are therefore always displayed on the touchscreen. There is no switching between the Start and Instrument program. The operation of the Sinius dental treatment center is thus similar to other Dentsply Sirona treatment centers that are operated without a touchscreen.
Page 58
4 Operation Dentsply Sirona 4.2 Control concept of touchscreen Operating Instructions Sinius When an instrument is removed, the EasyMode Start program displays the chosen speed and intensity settings on the upper left of the touchscreen. The value can be set using the quick setting keys (1%, 50%, 100%).
4.2 Control concept of touchscreen The EasyMode Start program is intended for users who would like to retain the familiar operating concept of other Dentsply Sirona treatment centers and who can do without the functions listed above. Program change keys The LED of the relevant program change key lights up in the Standard Start program according to the program selected.
4 Operation Dentsply Sirona 4.2 Control concept of touchscreen Operating Instructions Sinius 4.2.4 Status bar The status bar is located at the bottom edge of the screen. If multiple user profiles A are preselected, the active user profile A to F is displayed along with the current time, including a display with seconds.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.3 Foot control Foot control The treatment center can be operated using a wireless foot control or a foot control with a cable connection. 4.3.1 Wireless foot control Technical data of the wireless module, see Foot control radio interface" ...
4 Operation Dentsply Sirona 4.3 Foot control Operating Instructions Sinius 4.3.2 Operating the foot control The foot control operating elements are assigned different functions, depending on whether the instruments are all in place or an instrument is removed from its holder.
Page 63
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.3 Foot control Ä The chip blower remains switched on as long as the button is pressed. The boost function can be selected in ultrasound mode. When the intraoral camera is removed, it is possible to toggle between single image and quad image in SI Video.
4 Operation Dentsply Sirona 4.3 Foot control Operating Instructions Sinius 4.3.3 Using the cursor control The cursor control cannot be used in the EasyMode Start program operating mode. 4.3.3.1 Functionality Cursor control as an alternative mode of operation The touchscreen and the fixed keys of the dentist element can also be operated hands-free via the foot control.
The Clean and Setup keys and the ON/OFF switch cannot be reached by the cursor control. This adjacent image shows an example of the control panel of the Sinius and Sinius TS dentist element. The cursor path of the Sinius CS dentist element also follows this principle. Orange and blue bars A blue bar indicates which functions are assigned to the left or right button of the foot switch.
Page 66
4 Operation Dentsply Sirona 4.3 Foot control Operating Instructions Sinius ➢ Slide and hold the 4-way foot control plate up or down (auto cursor). Ä The orange cursor slowly moves from one cursor position to the next. Activating a function or fixed key ➢...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.4 Patient chair Patient chair 4.4.1 Safety instructions CAUTION ü û The free space under the patient couch and up to the water unit can be decreased due to chair movements. Parts of the patient's or user's body may be pinched or crushed.
4 Operation Dentsply Sirona 4.4 Patient chair Operating Instructions Sinius 4.4.2 Safety stop The treatment center is equipped with various safety stops to prevent crushing and damages. The cutoff trigger points are shown in the following illustration: Display of triggered safety switches (all shown on one illustration)
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.4 Patient chair 4.4.3 Triggering an immediate movement stop You can stop the movement of the chair to a programmed position as follows: ➢ Press the main switch or the Shock positioning fixed key on the dentist element.
4 Operation Dentsply Sirona 4.4 Patient chair Operating Instructions Sinius 4.4.4 Armrests Armrests are available for the patient chair. The right armrest can be swiveled forward to make it easier for patients to get in and out of the chair.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.4 Patient chair 4.4.6 Adjusting the motor-driven headrest The motor-driven headrest can be adjusted via the touchscreen or directly on the headrest. CAUTION Fine objects can enter the mechanism of the motor-driven headrest through the gap.
4 Operation Dentsply Sirona 4.4 Patient chair Operating Instructions Sinius 4.4.6.2 Inclining the headrest The headrest can be tilted either via motor drive or manually (quick mechanical adjustment). Via touchscreen ü The Standard Start program or the Manual Chair Adjustment program is displayed on the touchscreen.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.4 Patient chair 4.4.7 Adjusting the double articulating headrest The double articulating headrest is equipped with two rotary joints. They facilitate manual adjustment of head inclination for maxillary/mandibular treatments. The headrest extension can be pushed in or pulled out manually to adjust it to the height of the patient.
During an automatic program run, such as moving toward the entry/ exit or mouth rinsing positions, the patient may collide with the Sinius CS and Sinius TS dentist elements or the support arms. The patient is at risk of being crushed.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.4 Patient chair Via foot control ü All instruments are in place. ➢ Press the right button of the foot control. Via the assistant element ➢ Press the 0 key on the assistant element briefly (< 2 s).
4 Operation Dentsply Sirona 4.4 Patient chair Operating Instructions Sinius 4.4.8.4 Activating other chair programs ü The Start program is displayed on the touchscreen. ➢ Touch the 1 key or the 2 key briefly (< 2 s). 63 2 2 7 26 D3561...
Moving the patient chair may cause the patient to collide with the Sinius CS and Sinius TS dentist elements or the support arms. The patient is at risk of being crushed. ➢ Before moving the patient chair, position the Sinius CS and Sinius TS dentist elements so that collision with the patient or the patient chair is impossible.
4 Operation Dentsply Sirona 4.4 Patient chair Operating Instructions Sinius 4.4.9.2 Tilting the patient couch Movement of the seat and backrest without any compression or stretching effects for the patient Via touchscreen ü The EasyMode Start program or Manual Chair Adjustment program is displayed on the touchscreen.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.4 Patient chair 4.4.10 Creating chair and shock positioning programs Chair programs The four chair programs preset in the factory: ● Mouth rinsing position S ● Entry/exit position 0 ● 1 and 2 can be individually reprogrammed for each of the six user profiles (A to F).
4 Operation Dentsply Sirona 4.4 Patient chair Operating Instructions Sinius 4.4.11 Setting the active lumbar support The patient chair has the option of lumbar support. Opening Start sub-screen ü The Start program is displayed on the touchscreen. ➢ In the Standard Start program : Touch the Sub-screen key.
2 kg (4.4 lb). A sterilizable silicone mat can be placed over the dentist element. The maximum load on the Sinius CS dentist element tray is 1 kg (2.2 lb). The maximum load of the Sinius TS dentist element with a tray holder is 2 kg (4.4 lb).
Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.2 Height adjustment The height of the Sinius (sliding track device) dentist element can be adjusted to achieve an ergonomic instrument height. Please contact your service engineer. 4.5.3 Positioning the dentist element...
Page 83
Moving the patient chair may cause the patient to collide with the Sinius CS dentist element or its support arm. The patient is at risk of being crushed. ➢ Before moving the patient chair, position the Sinius CS dentist element so as to make collision with the patient or the patient chair impossible.
Page 84
Moving the patient chair may cause the patient to collide with the Sinius TS dentist element or its support arm. The patient is at risk of being crushed. ➢ Before moving the patient chair, position the Sinius TS dentist element so as to make collision with the patient or the patient chair impossible.
Operating Instructions Sinius 4.5 Dentist element 4.5.4 Fixed keys on the dentist element Arrangement of fixed keys on the Sinius and Sinius TS dentist elements Arrangement of fixed keys on the Sinius CS dentist element 4.5.4.1 Main switch The treatment center is switched on/off with the main switch.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius It is not possible: ● to switch to the Instrument program or the sub-screen of an instrument when no instrument has been removed ● To switch to the Sivision program when the PC connection is turned...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element Starting the timer 1. Press the Timer fixed key briefly. Ä The most recently set timer is started immediately. The set and elapsed time are displayed in the status bar. The Timer function program is also displayed.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Filling the tumbler 1. Place the tumbler under the tumbler filler. 2. Press the Tumbler Filling fixed key. Ä The tumbler is filled with water for the preset time. Pressing the Tumbler Filling fixed key again stops the filling function immediately.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element Switching the flushing on/off ➢ Press the Flushing fixed key. Ä The LED in the key lights up during the flushing function. The flushing function is activated for the preset flushing time.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.5 Quick setting keys and function levels Depending on which instrument is taken from the holder, the corresponding instrument program or respective functions appear in the EasyMode Start program. Instrument programs in the Standard Start program operating mode...
Page 91
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element The settings can be recalled by touching the respective quick setting key. Function levels When using function levels, you have two "storage locations" (E1/E2) at your disposal for saving settings or recalling them at the push of a button.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.6 Saving instrument settings With static quick setting keys You can determine whether the Memory key should be displayed in the instrument programs with the quick setting keys, see "Preselecting how instrument settings are to be saved" ...
Page 93
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 3. Repeat this process for other quick setting keys if necessary. Ä The settings can be recalled by touching the respective quick setting key. With function levels The settings which have been made can be saved to and recalled from two function levels (E1, E2).
If more than one instrument is removed, the Instrument program or the speed or intensity values of the instrument removed first are displayed. When using the Sinius or Sinius TS dentist element, make sure that the instruments are always placed in the correct instrument holders. If any instruments are placed in the wrong holders, the wrong instrument dialog will be opened when they are removed from the holders.
Page 95
4 Operation Operating Instructions Sinius 4.5 Dentist element Ball stopper for the Sinius and Sinius TS dentist element A ball stopper can be ordered as an accessory for a free instrument holder. Insert the ball stopper into an unassigned instrument holder. This prevents accidental deposit of an instrument in this holder.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.8 General instrument functions Settings for the coolant, instrument light, and foot switch can be made in the sub-screen of the instrument removed from the holder. The sub-screens vary according to the instrument removed. Functions not available for the respective instrument are not displayed in the sub- screen.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element Via touchscreen The preselected coolant in the EasyMode Start program operating mode can only be turned on/off with the foot control buttons, see below. ü An instrument is taken from the holder and the Instrument program is displayed on the touchscreen.
The spray amount is preset at the factory. However, it can be adjusted using the regulating screw on the dentist element. This setting is then valid for all burr drives. Sinius and Sinius TS dentist elements 1. Set the spray amount by turning the screw at the bottom of the dentist element.
Page 99
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 3. Set the spray amount by turning the screw. 4. Check the set spray amount with a burr drive. Correct the setting if necessary. 5. First insert the rear edge of the instrument holder into the groove on the dentist element.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.8.8 Preparing for use of NaCl saline solution The peristaltic pump prepares a sterile saline solution instead of spray water for cooling. The peristaltic pump hose is a disposable item. To reorder it, see "Spare parts and consumables" ...
Page 101
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 3. Open cover A. Lay the silicone hose B without prestress, with the thickened part wrapped around the pump wheel. See the adjacent drawing. Close cover A. IMPORTANT Direction of flow of the peristaltic pump The shorter end of the hose with the cannula must be located at the top of the pump;...
Operating Instructions Sinius The NaCl pump has two setting for peristaltic pump hoses of different diameters. For the Dentsply Sirona peristaltic hoses, the wide setting must be selected. This ensures an optimal supply rate. The NaCl pump is therefore delivered with the wide default factory setting.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.9 ApexLocator The ApexLocator can be used to measure the working length of the root canal file in endodontic treatments using electrical impedance. The ApexLocator can be used as follows: ●...
Apex adapter, mucosal electrode and file clamp The mucosal electrode and the file clamp are connected to the socket of the dentist element using the apex adapter. The socket on the Sinius and Sinius TS dentist elements is located underneath the dentist element;...
Page 105
4 Operation Operating Instructions Sinius 4.5 Dentist element The Dentsply Sirona Endo 6:1 contra-angle handpiece (from SN 6407 onwards / July 2010) or Endo 6 L handpiece is required for use in endodontic treatments with the ApexLocator. When using the ApexLocator in the motor program (without electronic torque limitation), the Dentsply Sirona SiroNiTi Apex handpiece is required.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.9.2 Distance display The measured root canal depth is shown in the distance display on the touchscreen. A bar with 11 display levels shows the distance from the root canal file to the physiological apex (apical constriction). The root canal is divided into four colored sections in the distance display.
Page 107
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element Green region The tip of the root canal file has reached the physiological apex. Red region The tip of the root canal file has pierced the physiological apex. The instrument overshoot is displayed.
Page 108
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Dental unit plugin "ApexLocator distance display" The distance display on the touchscreen can be additionally shown on the Sivision monitor. This requires that the Sidexis 4 and dental unit plugins are installed on the PC. In Sidexis 4, the distance is displayed/ hidden on the touchscreen by touching the distance display.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.9.3 Beeps In addition to the graphical distance display on the touchscreen, the position of the file in the root canal can also be indicated as an acoustic signal. Apex acoustic signals The ApexLocator always plays the following acoustic signals: ●...
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.9.4 Performing manual measurements with the file clamp For an endodontic examination, manual measurements can be made using the file clamp and a root canal file. ü The treatment center is prepared for manual measurements using the file clamp, see section "Preparing to use the ApexLocator" ...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.10 Electric motor IMPORTANT Also observe the operating instructions for the different motors. 4.5.10.1 Motor and coupling versions Different brushless motors are available depending on the type of use and the coupling of the handpiece.
Page 112
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius BL ISO E/C motors The green coded instrument hose (GN) must be used for the BL ISO E/ C motors. BL ISO E BL ISO C The BL ISO E/C motors are directly equipped with an ISO coupling. T1 Line handpieces, for example, can thus be used without an adapter.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.10.2 Setting the speed In the Standard Start program operating mode, settings can be made using the static quick setting keys (with the key values 0.09 or 0.2, 20, 40), with programmable quick setting keys (with changeable key values), or via function levels (E1, E2).
Page 114
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Setting intermediate speed values with static quick setting keys ü The electric motor is removed from the holder. ü The Motor program with static quick setting keys or the EasyMode Start program is displayed on the touchscreen.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.10.3 Setting the direction of rotation The direction of rotation can be changed only when the motor is at a standstill. In the EasyMode Start program operating mode, the direction of rotation can only be set by using the foot control.
Page 116
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Ä In the EasyMode Start program : If counterclockwise rotation is selected, an orange counterclockwise arrow appears. The counterclockwise arrow disappears when clockwise rotation is selected. Tip: After starting the electric motor with the foot switch, an audible warning signal sounds 6 times if counterclockwise rotation is activated.
Standard Start program operating mode. IMPORTANT SiroNiTi Apex Only Dentsply Sirona SiroNiTi Apex must be used to perform apex measurement using a torque-limiting handpiece! The electrical conductivity can be ensured only using this handpiece. The SiroNiTi handpiece can also be used to perform endodontic treatments without the ApexLocator.
Page 118
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Ä If the key is highlighted orange, the motor stops automatically when the physiological apex is reached. The – and + keys and the AutoReverse key are displayed. 4. Use the – and + keys to set the required apex distance from 1.5 to 0.
Page 119
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element acoustic signals vary depending on the measured distance to the preset motor stop position. For more information on the acoustic signals during apex measurement, see the section "Acoustic signals" [→ 109].
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.11 Highspeed handpiece 4.5.11.1 Using the highspeed handpiece IMPORTANT Also observe the operating instructions for the different high-speed handpieces. The high-speed handpiece hose is equipped with a standardized coupling according to ISO 9168.
Page 121
4.5 Dentist element Preselecting the instrument light operating voltage The original Dentsply Sirona halogen and LED lamps are usually operated at 3.6 V. Voltages over 3.8 V destroy the lamp. The operating voltage can be adjusted for lamps made by other manufacturers.
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.12 Sprayvit M multifunctional syringe The multifunctional syringe is used to clean the treatment areas and blow them dry. It supplies air and preheated water media. The heating cartridge for the water is located in the handpiece.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.12.2 Using the Sprayvit M multifunctional syringe The following instructions apply to the standard version (water on the right) of the Sprayvit M multifunctional syringe. A Sprayvit M with inverted media (water on the left) is available as an option.
Page 124
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Ä If the key is highlighted orange, the instrument light is turned on at the Sprayvit M multifunctional syringe if this is the only instrument to have been removed. Activating/deactivating and setting the water temperature in the...
CAUTION Ultrasonic tips from other manufacturers do not guarantee safe operation. ➢ Use only ultrasonic tips from Dentsply Sirona. 4.5.13.2 Setting the intensity In the Standard Start program mode, settings can be made using the...
Page 126
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Selecting the intensity with the quick setting keys ü The SiroSonic TL scaler is removed. ü Either the Ultrasonic program with static or programmable quick setting keys or the EasyMode Start program is displayed on the touchscreen.
Page 127
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element Setting intermediate intensity values with programmable quick setting keys ü The SiroSonic TL scaler is removed. ü The Ultrasonic program with programmable quick setting keys is displayed on the touchscreen.
Page 128
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Switching on the endodontics function The intensity of the endodontics function is limited for safety reasons, e.g., in order to prevent broken needles. If the instrument programs are displayed in the Standard Start program operating mode with static quick setting keys, they are...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.14 Implantology and endodontic treatment The treatment functions support implantology and endodontic treatments. The speed and torque of the rotating instrument can be precisely adjusted and saved for later reference, if desired. This is possible for every work step for implantology treatments.
Page 130
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Ä The Start program is displayed or, if an instrument has already been taken from the holder, the system will switch to the Instrument program . The Treatment key is marked orange. The bur drive that corresponds to the treatment is displayed in the status bar with an orange dot.
Page 131
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element Switching the treatment function off When the treatment function is activated, the Implantology program or the Endodontics program is displayed on the touchscreen instead of the Motor program when the bur drive assigned to the selected treatment is removed from its holder.
Operating Instructions Sinius 4.5.14.2 Implantology CAUTION Only the Dentsply Sirona contra-angle handpiece Implant 20:1 may be used for the Implantology function. Instruments from other manufacturers can lead to malfunctions. Third- party instruments may be improperly calibrated for implantology. ➢ Use only the Dentsply Sirona Implant 20:1 contra-angle handpiece from Sirona for implantology.
Page 133
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element IMPORTANT Torque adjustment The maximum adjustable torque depends on the system motor and the speed settings. CAUTION Improperly selected speeds and torque values endanger the patient. Treatment errors, e.g. jaw damage, may result from incorrect settings.
Page 134
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.14.2.5 Selecting a work step Individual settings can be made and saved for each implantology work step, e.g. predrilling, final drilling, tapping, etc., see "Saving settings" [→ 136]. At the end of each work step, the required settings can be accessed immediately by selecting the next step.
Page 135
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 4.5.14.2.6 Functions in the sub-screen Opening the implantology sub-screen ü The Implantology program is displayed on the touchscreen. ➢ Touch the Sub-screen key. Ä The Implantology sub-screen is displayed. Preselecting NaCl rinsing, setting the flow rate, and activating the...
Page 136
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Activating NaCl rinsing via button on foot control ü The NaCl rinsing key is displayed in the Implantology program . ➢ Press the right button of the foot switch. Ä The NaCl rinsing function remains active as long as the button is pressed.
Page 137
The dentist element and tray can be covered with a surgical drape to allow for sterile operation. For the Sinius with sliding track and Sinius TS a rectangle must be cut out of the drape to allow for operation and visibility of the EasyTouch.
Instruments from other manufacturers can lead to malfunctions. Third- party instruments may be improperly calibrated for endodontics. ➢ For endodontics, use Dentsply Sirona Endo 6:1 (SN 6407 and above / July 2010) or Endo 6 L contra-angle handpieces. ➢ Check whether the transmission ratio on the touchscreen agrees with the value specified on the contra-angle handpiece being used.
Page 139
This includes a measurement of motor current at different speeds to assess the properties of the system. CAUTION To ensure proper calibration, use only Dentsply Sirona instruments. ü The Endodontics program is displayed on the touchscreen. 1. Attach the contra-angle handpiece that you would like to use for the treatment to the electric motor.
Page 140
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.14.3.4 Using reciprocal rotating files For endodontic treatment with reciprocating rotating files, the sequence of the files and the preset values recommended by the manufacturer cannot be changed. The keys for setting the speed and torque and the counterclockwise key are hidden.
Page 141
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element For manual measurement using the file clamp, you can change directly from the Endodontics program to the Start sub-screen. ü Endodontics treatment is activated. 1. Return the endo handpiece to the instrument holder.
Page 142
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.14.3.6 Functions in the sub-screens Opening the endodontics sub-screen Settings, such as changing the preselected speed and torque values [→ 142], or removing a file from the sequence [→ 145], are file-specific.
Page 143
Instruments from other manufacturers can lead to malfunctions. Third- party instruments may be improperly calibrated for endodontics. ➢ For endodontics, use Dentsply Sirona Endo 6:1 (SN 6407 and above / July 2010) or Endo 6 L contra-angle handpieces. ➢ Check whether the transmission ratio on the touchscreen agrees with the value specified on the contra-angle handpiece being used.
Page 144
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius Turning on/off the automatic motor stop of the ApexLocator If your treatment center is equipped with the ApexLocator option, you can set the motor to stop at the physiological apex. The motor stop can be combined with the AutoReverse function, see “Switching the...
Page 145
Ä The selected file is removed from the sequence. Switch instrument light on/off The Dentsply Sirona Endo 6L handpiece is equipped with a light guide. ➢ Switch the instrument light on or off with the Instrument light key. Ä If the key is highlighted orange, the instrument light can be activated using the foot pedal.
Page 146
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 4.5.14.3.7 Endodontic treatment administration Up to three endodontic treatments can be added to the treatment list of the Treatment selection screen. The following functions are for managing the endodontic treatment list: ●...
Page 147
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.5 Dentist element 2. Touch the Copy endodontic treatment key. Ä A keyboard is displayed. The name of the endodontic treatment to be copied is displayed in the text box. 3. Enter the name of the copy. Confirm your entry with the OK key.
Page 148
4 Operation Dentsply Sirona 4.5 Dentist element Operating Instructions Sinius 1. Touch the button of an endodontic treatment that you would like to delete. Ä The selected button is highlighted orange. 2. Touch the Delete endodontic treatment key for more than 2 seconds.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.6 Assistant element Assistant element 4.6.1 Maximum load capacity The maximum load of the assistant element is 1.5 kg (3.3 lbs). A skid- proof silicone mat can also be used. CAUTION To prevent injuries caused by falling objects, never place anything on the support arm of the assistant element.
4 Operation Dentsply Sirona 4.6 Assistant element Operating Instructions Sinius 4.6.3.1 Tumbler filling If the your treatment center is equipped with the Tumbler filling option with automatic sensor control, see “Tumbler filling with automatic sensor control” [→ 159]. 1. Place the tumbler under the tumbler filler.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.6 Assistant element 4.6.3.5 Hash key The Hash key on the assistant element can be configured in the setup program. The key allows the user to switch the X-ray viewer or the white screen function on the Sivision monitor on or off or alternatively to activate the bell or hash relay, see “Setting function of the Hash key on...
4 Operation Dentsply Sirona 4.6 Assistant element Operating Instructions Sinius 4.6.4 Suction handpieces The assistant element can be equipped with a maximum of three suction handpieces. CAUTION The tip attaches itself to the oral mucosa. The patient's oral mucosa can be irritated by the vacuum.
Page 153
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.6 Assistant element Saliva ejector A curved tip that can be placed in the corner of the mouth is provided for saliva ejection. Surgical suction device The surgical cannula can be inserted directly into the saliva ejector handpiece.
4 Operation Dentsply Sirona 4.6 Assistant element Operating Instructions Sinius 4.6.5 Sprayvit M multifunctional syringe The function is described in the chapter on the dentist element, see "Sprayvit M multifunctional syringe" [→ 122]. 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.6 Assistant element 4.6.6 Mini LED curing light The curing light is used to cure composite material with short-wave light. 4.6.6.1 Safety instructions CAUTION LED beam exit aperture The curing light contains powerful LEDs. In accordance with IEC 62471, they are classified in risk group 2 with the light bar removed.
4 Operation Dentsply Sirona 4.6 Assistant element Operating Instructions Sinius 4.6.6.3 Connecting the Mini L.E.D. The Mini L.E.D. is connected to the assistant element in holder 1. 1. Connect the supply cable A to the treatment center. 2. Screw hose coupling B onto the Mini L.E.D. C.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.6 Assistant element Quick curing mode In the quick curing mode, the Mini L.E.D. operates at full power for 10 seconds. In this mode, the light intensity output is as follows: Sec. ²...
4 Operation Dentsply Sirona 4.6 Assistant element Operating Instructions Sinius 4.6.6.6 Technical data General Technical Data on Mini L.E.D. Model: Mini L.E.D. Weight of handpiece without 105 g hose: Dimensions: Dia. 23 mm x 240 m m Current consumption of hand- 5 V DC / 0.65 A...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.7 Water unit Water unit 4.7.1 Swiveling the cuspidor bowl The cuspidor on the water unit can be manually swiveled approx. 110 mm toward the patient chair. IMPORTANT Chair movements with cuspidor swiveled in With the cuspidor bowl swiveled in, the chair cannot be moved.
4 Operation Dentsply Sirona 4.7 Water unit Operating Instructions Sinius 4.7.3 Adjusting the water amount for flushing The water quantity and water pressure for flushing can be set using a valve. 1. Open the maintenance flap A on the base of the water unit.
These microorganisms could increase the risk of damage to one’s health. ➢ Dentsply Sirona recommends never operating the treatment center without the agent for disinfecting the water paths. ➢ Mix fresh water for the self-sufficient water supply every day. At the end of the work day, the remaining water must be flushed out of the disinfectant tank using the tumbler filling function.
Page 162
4 Operation Dentsply Sirona 4.7 Water unit Operating Instructions Sinius Opening Start sub-screen ü The Start program is displayed on the touchscreen. ➢ In the Standard Start program : Touch the Sub-screen key. In the EasyMode Start program : Press the Chair program change key.
Page 163
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.7 Water unit Ä In the Start sub-screen, the Aqua key is highlighted in orange. The disinfection system is now changed to the self-sufficient water supply. If the treatment center is changed to self-sufficient water supply, the Sanitization key is hidden in the Start sub-screen.
4.8.1 Tray on the support arm (for Sinius) Maximum load capacity Sinius treatment centers as a sliding track device can be equipped with a tray on a support arm. The maximum load of the tray is 1 kg (2.2 lbs). Removing the tray 1.
Tray on the Sinius CS dentist element Maximum load capacity Sinius CS treatment centers can be equipped with a tray that is mounted underneath the dentist element. The tray holder is available in two sizes for one or two standard trays.
4.8.3 Tray on the Sinius TS dentist element Maximum load capacity Sinius TS treatment centers can be equipped with a tray mounted to the left of the dentist element. The maximum load of the tray is 1 kg (2.2 lbs). With the Sinius TS, the total maximum load on the tray and dentist element should not exceed 2 kg (4.4 lb).
0.2-liter cups. max. 1 kg For Sinius with sliding track and Sinius TS, the cup holder may not be 2.2 lbs attached to the front edge of the tray. The maximum load of the cup holder is 0.5 kg (1.1 lbs). The load on the tray and the cup holder may not exceed a total of 1 kg (2.2 lbs).
Operating Instructions Sinius Panoramic X-ray image viewer The Sinius TS treatment center can be equipped with an X-ray viewer on the support arm of the dentist element arm. An X-ray viewer cannot be attached to the Sinius and Sinius CS treatment centers.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.9 Panoramic X-ray image viewer 4.9.2 Attaching the anti-glare film An anti-glare film is supplied with the X-ray viewer for viewing intraoral dental images. 1. Remove the three retaining clips A. 2. Use two of the retaining clips to clamp the anti-glare film B to the upper edge of the X-ray viewer on the left and right.
4 Operation Dentsply Sirona 4.10 Operating light Operating Instructions Sinius Operating light 4.10 The operating light is mounted on a multi-joint support arm. It can be easily adjusted to the operating field using the handles. Brakes in the support arm hold the operating light in the position to which it has been adjusted.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.10 Operating light 2. Use the – and + keys to set the desired brightness of the operating light (7 levels). The – key decreases the brightness and the + key increases it.
The Sinius, Sinius CS, and Sinius TS treatment centers can be equipped with the Heliodent Plus X-ray tube unit. However, for the Sinius CS, this is possible only if the dentist element is mounted in the standard position at the water unit, see section "Positioning the dentist element" ...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.12 Sivision Digital video system Sivision Digital video system 4.12 The Sivision Digital video system enables intraoral and extraoral images to be made. The SiroCam AF / AF+ intraoral camera generates digital image data that can be transmitted via a USB 2.0 port (high-speed universal serial bus) to a connected PC and stored there.
4 Operation Dentsply Sirona 4.12 Sivision Digital video system Operating Instructions Sinius 4.12.1 Sivision monitor 22" DC monitor (REF 64 9 7 4 52 D3655) The monitor is equipped with loudspeakers and a glass screen to facilitate cleaning. For details, see "Operating Instructions for the 22 inch DC monitor".
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.12 Sivision Digital video system 4.12.2 SiroCam AF / AF+ intraoral camera 4.12.2.1 Safety instructions The SiroCam AF / AF+ intraoral camera is a sensitive optical instrument and must therefore always be handled with care.
Connecting the intraoral camera to the assistant element (only with Sinius CS and Sinius TS) If the treatment center is equipped with the Sinius CS or Sinius TS dentist element, the intraoral camera can be operated at the first instrument position of the assistant element. The camera connection is located underneath the assistant element instead of the connection for the curing light.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.12 Sivision Digital video system Removing the intraoral camera The plug of the intraoral camera is secured against unintentional removal. ➢ Grasp the plug by its locking device A and pull this out without tilting, if possible.
Page 178
4 Operation Dentsply Sirona 4.12 Sivision Digital video system Operating Instructions Sinius 4.12.2.4.2 Using the camera with SI Video Using SI Video, up to four still images can be generated with the SiroCam AF / AF+ intraoral camera. These can be displayed on the Sivision monitor either as a sequence of single images or simultaneously as a quad image.
Page 179
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.12 Sivision Digital video system Switching between quad image and single image When switching from quad to single image, the previously marked quad image is displayed as a single image. Conversely, the displayed single image is marked in the quad image.
Page 180
4 Operation Dentsply Sirona 4.12 Sivision Digital video system Operating Instructions Sinius 4.12.2.4.3 Using the camera with Sidexis Besides displaying and processing X-ray images, Sidexis 4 and Sidexis XG can be used additionally as a video application for the SiroCam AF / AF+ intraoral camera. Sidexis displays live and still images in separate windows.
Page 181
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.12 Sivision Digital video system Switching the SiroCam AF / AF+ intraoral camera on/off ü The PC is in operation and the Sivision Connect or Siucom Plus PC application is started. ➢ Take the SiroCam AF / AF+ intraoral camera from the holder.
4 Operation Dentsply Sirona 4.12 Sivision Digital video system Operating Instructions Sinius 4.12.2.5 Technical camera data General technical data Model: Intraoral camera SiroCam AF SiroCam AF+ Weight of handpiece without ca- approx. 80 g approx. 85 g ble: Dimensions: diam. 27.5 x 207 mm Working temperature: +10 to +40 °C...
For details, please refer to the "Installation and configuration of Siucom Plus / Sivision Connect" manual. CAUTION Integrating the Sinius into an IT network that includes other devices can lead to previously unknown risks. The following changes to the IT network can lead to new risks: •...
4 Operation Dentsply Sirona 4.13 Operation with a PC Operating Instructions Sinius 4.13.1 Sivision program Various PC applications of the PC can be started and operated in the Sivision program. The PC applications to be operable via the treatment center can be set in Siucom Plus.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.13 Operation with a PC The treatment center starts the PC applications automatically. An orange square located in front of the respective PC application on the touchscreen indicates whether the application has been started and is ready for operation on the PC.
4 Operation Dentsply Sirona 4.13 Operation with a PC Operating Instructions Sinius 4.13.1.4 Communication with Sidexis The Sidexis 4 and Sidexis XG PC applications can display X-ray and intraoral camera images on the Sivision monitor. They can save images taken with the SiroCam AF / AF+ intraoral camera in the patient database.
Page 187
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.13 Operation with a PC Contrast optimization filter This image filter analyses and optimizes the current grayscale distribution of an image. In this way, for instance, details within a very low-contrast, “faint” image can be made visible.
4 Operation Dentsply Sirona 4.13 Operation with a PC Operating Instructions Sinius Cancel/confirm entry Accept an order Accepts an order that was placed and is waiting in Sidexis, e.g. for creating an intraoral image with the X-ray unit of the treatment center or a video recording with the intraoral camera.
Type BF CAUTION Only type BF medical devices and applied parts according to IEC 60601-1, e.g. the Dentsply Sirona Xios XG USB module intraoral X- ray system, may be connected to the USB interface. CAUTION USB devices with their own voltage supply (e.g., via a power supply unit) may lead to increased leakage current.
4 Operation Dentsply Sirona 4.14 Configuration of the treatment center (setup) Operating Instructions Sinius Configuration of the treatment center (setup) 4.14 Various treatment center functions can be individually configured via the Setup settings. The treatment center can therefore be adapted to match each user's personal method of treatment.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.14 Configuration of the treatment center (setup) 4.14.2 Configuring the EasyTouch user interface ➢ Touch the EasyTouch user interface key in the setup program. Ä The sub-screen opens. 4.14.2.1 Switching the key sound on/off The program can be set to issue an acoustic signal that sounds when the operator touches a key on the touchscreen or a fixed key.
4 Operation Dentsply Sirona 4.14 Configuration of the treatment center (setup) Operating Instructions Sinius 4.14.3 Setting the date and time ➢ Touch the Date and time key in the setup program. Ä The sub-dialog opens. Setting the date The date is displayed in the format day/month/year.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.14 Configuration of the treatment center (setup) 4.14.4 Configuring control options Operation of the entire treatment center operation can be configured in this dialog. ➢ Touch the Control options key in the setup program.
4 Operation Dentsply Sirona 4.14 Configuration of the treatment center (setup) Operating Instructions Sinius ➢ Change to the next setup dialog page. 4.14.4.4 Switching on/off the foot control function for intraoral camera focus The foot control can be set to focus the SiroCam AF / AF+ intraoral camera: ●...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.14 Configuration of the treatment center (setup) If you deposit the spray aspirator in its holder while the suction flow is interrupted, the suction flow is automatically restarted when you pick it up again.
Switching the central supply for chemical suction hose cleaning off and on For clinical use, Sinius treatment centers can be equipped with a central cleaning agent supply for chemical suction hose cleaning. The CDS 60 system of Dürr Dental is intended for this purpose.
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.14 Configuration of the treatment center (setup) 4.14.4.12 Adjusting the temperature of the tumbler heater The water temperature for the tumbler filling can be adjusted. ➢ Use the – and + keys to set the intensity of water heating using tumbling heater.
4 Operation Dentsply Sirona 4.14 Configuration of the treatment center (setup) Operating Instructions Sinius 4.14.5 Configure instruments ➢ Touch the Instruments key in the setup program. Ä The sub-screen opens. 4.14.5.1 Preselecting how instrument settings are to be saved The settings in the instrument programs can be made either via static quick setting keys (with the key values 0.09 or 2, 20, 40 or 1, 50, 100),...
Dentsply Sirona 4 Operation Operating Instructions Sinius 4.14 Configuration of the treatment center (setup) 4.14.5.3 Showing/hiding the key for the external HF surgical unit External HF surgical devices may interfere with the treatment center and Sivision monitor. Therefore, the External HF surgical devices key may be displayed on the Start sub-screen.
Ä The sub-dialog Network connection opens. It shows the currently used network configuration. Call in your data processing specialist for network configuration. Network configuration is described in detail in the "Sinius / Sinius CS / Sinius TS installation instructions". 4.14.7 Opening the service function The Service domain is intended to be used only by service engineers.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.1 Basics Care, cleaning and maintenance by the practice team Basics Handling mainly involves the following steps: ● Cleaning ● Disinfection ● Sterilization if possible It is recommended to clean the unit as soon as possible after use.
Page 202
Check the message system of the amalgam separator [→ 245] Change the battery of the wireless foot control [→ 262] For a quick overview of the work involved, see the “Maintenance, cleaning, and disinfection schedule” for the Sinius treatment center. 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.1 Basics 5.1.2 Care, cleaning, and disinfecting agents NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents approved by Dentsply Sirona! A continuously updated list of approved agents can be downloaded from the Internet on the online portal for technical documents.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.1 Basics Operating Instructions Sinius 4. Immerse the tester in the filled container for 30 seconds. Ä The cardboard disk with the culture medium will now absorb 1 ml of the water sample.
134°C 274°F Sterilization may be conducted for components that are marked 3 min accordingly. Dentsply Sirona recommends sterilization in a steam sterilizer (autoclave) according to ISO 13060, Class B (e.g., DAC PREMIUM / DAC PROFESSIONAL). The sterilization must be carried out with a multiple fractionated suction system (Class B sterilizer).
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.2 Surfaces Operating Instructions Sinius Surfaces 5.2.1 Clean/disinfect surfaces The surfaces can be disinfected by spraying and wiping with surface disinfectants. NOTE Drugs have a chemical reaction with the surface of the unit.
NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. 1. Actuate the Clean fixed key on the dentist element. Ä A display stating that the touchscreen and fixed keys are deactivated appears on the touchscreen.
NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. The handles on the dentist and assistant element can be spray, wipe and thermally disinfected.
Page 209
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.2 Surfaces Assistant element ➢ Press the button provided at the end of the handle. Ä This will release the locking mechanism and the handle can be detached.
6. Thermally disinfect the tray. To insert the tray, simply guide it into the mount. The mechanism locks automatically. Tray on the Sinius CS and Sinius TS dentist element 1. Hold the tray tightly. 2. Open the lock by swiveling the lever downwards.
The disposable cup must be replaced after each patient. NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Operating Instructions Sinius 5.2.6 Care for, clean, and disinfect upholstery Special care, cleaning, and disinfecting agents are recommended by Dentsply Sirona for the care, cleaning, and disinfection of the upholstery. NOTE Approved care, cleaning, and disinfecting agents Use only agents approved by Dentsply Sirona for the upholstery, see"Care, cleaning, and disinfecting agents" ...
5.2.8 Thermally disinfecting the instrument holder of the dentist element and sterilizing the silicone mat Sinius and Sinius TS dentist elements The instrument holder can be removed for cleaning and thermal disinfection. The removable silicone mat on the dentist element can be sterilized.
Page 214
Ball stopper (for Sinius and Sinius TS dentist elements) If a ball stopper is inserted in an unused instrument holder, it can simply be pushed out of the holder from the rear to facilitate cleaning or thermal disinfection.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.2 Surfaces 5.2.9 Thermally disinfecting the instrument holder of the assistant element and sterilizing the silicone mat The instrument holder can be removed for cleaning and thermal disinfection.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Operating Instructions Sinius Instruments and instrument hoses 5.3.1 Rinse water paths Microorganisms can grow in the water paths of the treatment center. Use a large amount of water for rinsing the lines prior to starting patient appointments.
Page 217
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.3 Instruments and instrument hoses Ä The Start sub-dialog is displayed. 2. Touch the Purge function key. Ä The Purge screen is displayed on the touchscreen.
Page 218
ü The Remove instruments display appears on the touchscreen. 1. Remove the instruments to be purged from the holder. With the Sinius CS: Move the swivel arm of the instruments to be purged into the operating position so that the weight of the instrument hoses keeps them in this position.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.3 Instruments and instrument hoses 5.3.3 Automatically purge water paths (autopurge function) The AutoPurge function enables automatic purging of all water-carrying instruments in the dentist element, of the Sprayvit M multifunctional syringe in the dentist and assistant elements, and of the suction hoses and the tumbler filling.
Page 220
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Operating Instructions Sinius 2. Touch the Autopurge function key. Ä The Autopurge screen is displayed on the touchscreen. Setting the instrument purge time and starting the AutoPurge...
Page 221
Sprayvit M: valve lever in the up position, locking knob in the down position). Also for Sinius CS: Move the swivel arm of the instruments to be purged into the operating position so that the weight of the instrument hoses keeps them in this position.
Page 222
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Operating Instructions Sinius IMPORTANT Pinching the instrument hoses Be careful not to pinch the instrument hoses when inserting the instruments. If the hoses are pinched, the water flow will be obstructed during purging.
Page 223
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.3 Instruments and instrument hoses Following the purging process and suction hose cleaning, the display Deposit instruments appears. Displays if the purging process is not completed The AutoPurge program can determine whether the purge process for all instruments and the tumbler filling has been completed.
Page 224
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Operating Instructions Sinius ➢ Remove the instruments from the water unit and place them back in their holders. Ä When all of the instruments have been deposited, the Deposit instruments display disappears.
NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. Disconnect the mucosal electrode from the connection cable. The metal hook can be sterilized, and the connection cable can be disinfected by wiping.
NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. 1. Remove the connection cable of the Mini L.E.D. by turning the 134°C 274°F...
It can be wipe disinfected. NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. NOTE The lens window is sensitive to scratches.
This oil is collected in a drip container by the cotton wool roll. Sinius and Sinius TS dentist elements 1. Slide the drip container down and remove the cotton wool roll.
Page 229
Operating Instructions Sinius 5.3 Instruments and instrument hoses Oil collector The Sinius CS dentist element is also equipped with an oil collector. It is located below the instrument connections and collects any liquids that leak out. 1. Remove the instrument holder from the dentist element, as described above.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.4 Vacuum system Operating Instructions Sinius Vacuum system 5.4.1 Cleaning the suction hoses The suction system is exposed daily to secretions, saliva, and blood that contain bacteria. Therefore, for reasons of hygiene, the used suction hoses must be cleaned after each patient, in particular after every treatment involving blood.
Page 231
➢ Do not fill the agent for disinfecting the water paths into the cleaning agent tank for chemical suction hose cleaning! Use only "Agents for the suction lines" approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. 1. Open the maintenance flap on the base of the water unit. The cleaning agent tank for chemical suction hose cleaning is located on the left side.
NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. NOTE Domestic cleaning agents foam up. Foaming cleaning agents often cause foam and water to be sucked into the dry suction system.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.4 Vacuum system Cleaning procedure 1. Pour 1 liter of cleaning solution into the cuspidor. 2. Remove the suction hoses from their holders and attach them to the side of the cleaning adapter as simultaneously as possible.
➢ Never use petroleum jelly or similar lubricants. ➢ Use only lubricants approved by Dentsply Sirona. 1. Pull the handpiece of the spray aspirator, the saliva ejector and, if installed, the surgical suction device off of their suction hoses.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.4 Vacuum system 5.4.4 Cleaning and disinfecting the suction hoses The hoses of the spray aspirator and the saliva ejector as well as the connection hose to the water unit can be removed for rinsing under running water.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.4 Vacuum system Operating Instructions Sinius 4. Take the collectors C out of the suction hoses. Collect the amalgam residues in a glass filled with water. 5. Take off the suction cannulae and rinse out the suction hoses with clean water.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.5 Components of the water unit Components of the water unit 5.5.1 Clean the gold trap The gold trap retains larger solid particles to prevent them from being washed down the drain in the cuspidor bowl.
This can cause damage to the suction machine. ➢ Use only care, cleaning, and disinfecting agents approved by Dentsply Sirona, see"Care, cleaning, and disinfecting agents" [→ 203]. The cuspidor bowl is attached to the water unit via a bayonet catch and can be removed for thorough cleaning.
Use the same agent as for the suction lines. NOTE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents" [→ 203]. NOTE Domestic cleaning agents foam up.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Operating Instructions Sinius 5.5.4 Refill disinfectant for the water paths The water unit can be optionally equipped with a disinfection system. In normal operation, this will automatically inoculate the water that comes in contact with the patient (also called treatment water) with an agent to disinfect the water paths.
NOTE Approved care, cleaning, and disinfecting agents Use only the agent for disinfecting the water paths that has been approved by Dentsply Sirona for the disinfection system; see "Care, cleaning, and disinfecting agents" [→ 203]. 5.5.5 Change the water and air filters If you notice any changes in media flow rates, check the water and air filters for permeability.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Operating Instructions Sinius 5.5.6 Changing the amalgam rotor Amalgam residues and other solid particles are trapped in the amalgam rotor according to the centrifugal principle.
Page 243
6. Place the container in the special packaging and ship it for disposal or authorize a certified waste management company. Installing the amalgam rotor Use only original Dentsply Sirona accessories. Never use a used or recycled amalgam rotor. To reorder the amalgam rotor, see "Spare parts and consumables" ...
Page 244
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Operating Instructions Sinius Amalgam separator operator's log In Germany, users are obligated by law to keep an operation log for the amalgam separator. This log is included with the treatment center.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.5 Components of the water unit 5.5.7 Check the message system of the amalgam separator An electronic control monitors the function of the amalgam separator. It detects any mechanical blocking or failure of the drive motor.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Operating Instructions Sinius 5.5.8 Emptying the sediment container In addition to other solid particles, a large proportion of the amalgam residues are trapped in the sediment container by gravitational force.
Page 247
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.5 Components of the water unit 2. Hold the sediment container B so that the latching noses of the bayonet catch are positioned transverse to the water unit. Rotate the sediment container B clockwise.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Operating Instructions Sinius 5.5.9 Cleaning the filter insert of the standard wet suction The automatic separator and amalgam separator or sediment container are not installed in the water unit for standard wet suction. The separation of air and water and amalgam separation are performed centrally in this case.
Page 249
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.5 Components of the water unit unit is completely empty. When the key turns gray the extraction process is finished. Removing and cleaning the filter insert...
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitation Operating Instructions Sinius Sanitation Sanitation effectively combats the reproduction of microorganisms in the water paths. If the disinfection system is set for operation with public drinking water supply, the sanitization must be guided using the sanitization screen, see "Interactively guided sanitation of the treatment center" ...
Page 251
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.6 Sanitation 2. Set all bur drives and the SiroSonic TL scaler to the maximum water flow rate. max. 3. Put all instruments and suction hoses in their holders.
Page 252
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitation Operating Instructions Sinius ● Crossed-out solid gray circle Instrument excluded from sanitation, after a failed flow rate check or if an instrument was returned during the reaction time ●...
Page 253
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.6 Sanitation Sanitization phase 2 – Checking the water flow First, the treatment center checks whether there is water flow through the instruments. ü Sanitization phase 2 is highlighted on the touchscreen.
Page 254
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitation Operating Instructions Sinius Sanitization phase 4 – Allow 24 hours reaction time In order to effectively combat germs, you must let the agent for disinfecting the water paths react for at least 24 hours and not longer than 3 days (maximum sanitation time).
Page 255
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.6 Sanitation Sanitization phase 5 – Rinsing water paths with water The disinfectant for the water paths is rinsed purged from the instrument hoses, Sprayvit M hoses, and tumbler filling unit with water. This takes several minutes.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitation Operating Instructions Sinius 5.6.2 Display sanitation report Sanitation processes with interactive guidance, with a reaction time of at least 24 hours, are logged by the treatment center. The corresponding log reports can be displayed on the touchscreen.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.6 Sanitation 5.6.3 Sanitizing the treatment center manually IMPORTANT The sequence for manual sanitization described below is feasible only if the treatment center is operated with a self-sufficient water supply.
Page 258
(in for Sprayvit M: valve lever up, locking button down). Additionally for Sinius CS: Move the swivel arm of the Sprayvit M into the working position so the weight of the instrument tube holds it in this position.
Page 259
Activate each instrument for approximately 15 seconds. Then return the instruments to their holders. Also with the Sinius CS: Move the swivel arm of the instruments to the working position. 4. Empty the rinsing tumbler and return the empty tumbler below the tumbler outlet.
Page 260
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitation Operating Instructions Sinius Purging water paths with water The disinfectant for the water paths is rinsed out of the instrument hoses, Sprayvit M hoses, and tumbler filling unit with water.
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.6 Sanitation 5.6.4 Biofilm removal by the service technician If the microbiological test of the water from the treatment center does not correspond with the hygiene requirements in spite of regular sanitization of the water paths and/or regular purge/autopurge cycles, the biofilm must be removed using special chemicals.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.7 Foot control and connection box Operating Instructions Sinius Foot control and connection box 5.7.1 Changing the battery of the wireless foot control The wireless foot switch is powered by a battery. An empty battery is detected and displayed by the system.
Page 263
Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Operating Instructions Sinius 5.7 Foot control and connection box Replacing the battery (with battery compartment) 1. If the adjustable feet D sit close to the battery cover E, slide this small part outward.
5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.7 Foot control and connection box Operating Instructions Sinius 5.7.2 Changing the fuse of the external device connection The inlet connector remains live even when the power switch is turned off.
Dentsply Sirona 6 Maintenance by the service engineer Operating Instructions Sinius 6.1 Inspection and maintenance Maintenance by the service engineer Inspection and maintenance In order to ensure the operational safety and reliability of your treatment center and to avoid damage due to natural wear, annual inspection and maintenance must be performed on your treatment center.
6 Maintenance by the service engineer Dentsply Sirona 6.3 Maintenance Manual Operating Instructions Sinius Maintenance Manual Keep this Maintenance Manual near your treatment center. Any inspection and maintenance work as well as all safety tests are documented by the service engineer in the Maintenance Manual.
Dentsply Sirona 7 Malfunctions Operating Instructions Sinius 7.1 Error messages Malfunctions Error messages Error states of the treatment center that cannot be immediately recognized by the user but must nevertheless be corrected are displayed on the touchscreen. If an error state exists, this is indicated by a warning triangle A in the status line of the touchscreen.
Page 268
7 Malfunctions Dentsply Sirona 7.1 Error messages Operating Instructions Sinius Code Anomaly Description Measure The flushing tank is empty It was detected that the flushing Change the water filter, see “Chang- tank had not been filled. Flushing of ing the water and air filters” [→ 241].
Dentsply Sirona 7 Malfunctions Operating Instructions Sinius 7.2 Remote diagnosis Remote diagnosis Functional description With remote diagnosis, you enable the staff of your dental depot or our Customer Service Center, see "Contact data" [→ 11], to connect to the PC of your treatment center. The contents of your PC monitor are then transmitted to the computer of the service specialist and remote access to your PC is enabled.
REF 62 25 903 (5 pcs) Fuses Fuse for the external device connection 100 V – 240 V~ (T 6.3 A, 250 V~) REF 10 77 452 Other Ball stopper For sealing the instrument holder of the Sinius TS and Sinius dentist elements REF 58 99 575 63 2 2 7 26 D3561 D3561.201.01.16.02 08.2018...
Page 271
Dentsply Sirona 8 Spare parts and consumables Operating Instructions Sinius Silicone isolation sleeve for endodontics with the ApexLocator REF 63 24 631 (5 pcs) Separate motor holder for implantology motor REF 59 99 821 Battery for wireless foot control 1x type alkaline baby (C or LR14) with 1.5 V (commercially available) Always use high-quality batteries! 63 ...
9 Disposal Dentsply Sirona 9.1 Battery disposal Operating Instructions Sinius Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.1 Fixed keys Overview of all function keys The following contains a brief description of the fixed keys on the dentist and assistant elements and the function keys on the touchscreen and to provide the reader with a quick overview of the significance of the symbols on the keys.
Page 274
10 Overview of all function keys Dentsply Sirona 10.1 Fixed keys Operating Instructions Sinius Tumbler filling Starts or stops the tumbler filling function. When the Tumbler filling key is pressed (> 2 s), the filling time and water heating settings screen appears.
Page 275
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.1 Fixed keys 10.1.2 Assistant element Tumbler filling On/Off Flushing the cuspidor On/Off Operating light / Composite function Switches the operating light on, to the composite function, or off.
Page 276
10 Overview of all function keys Dentsply Sirona 10.2 Start program Operating Instructions Sinius Start program 10.2 Chair program S Mouth rinsing position with last-position memory function (programmable) Chair program 0 Entry/exit position (programmable) Chair programs 1 and 2 (programmable)
Page 277
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.2 Start program Stand-alone water supply The disinfection system switches to self-sufficient water supply Extracting residual water Extraction of residual water by standard wet suction before cleaning the filter...
Page 278
10 Overview of all function keys Dentsply Sirona 10.3 Instrument program Operating Instructions Sinius Instrument program 10.3 Static speed quick setting keys Setting preset or intermediate speeds BL, BL ISO C and BL Implant motors: min. 90 rpm, max. 40,000 rpm BL ISO E motor: min.
Page 279
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.3 Instrument program Activate preselected coolant The coolants to be available for selection in the Instrument program can be set in the sub-screen of each individual instrument, see below.
Page 280
10 Overview of all function keys Dentsply Sirona 10.4 Treatment program Operating Instructions Sinius Apex distance acoustic signals Switching the apex distance signal tones on or off. The intervals between the signal tones vary according to the measured distance to the apex.
Page 281
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.4 Treatment program AutoReverse function When the set torque value is reached, the bur drive automatically switches to counterclockwise. If your treatment center is equipped with the ApexLocator option, the bur drive stops automatically when the set motor stop position is reached.
Page 282
10 Overview of all function keys Dentsply Sirona 10.5 Other dialogs Operating Instructions Sinius Other dialogs 10.5 10.5.1 Timer screen Timer keys Up to four timers can be set. The maximum time setting is 9 minutes and 30 seconds. Time loop If the key is highlighted orange, the countdown will automatically be restarted when the set time has expired.
Page 283
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.5 Other dialogs 10.5.5 Operating light setting program Sensor control for switching the operating light on/off If the key is highlighted orange, the operating light can be switched on/ off or changed to the Composite function by a hand movement.
Page 284
10 Overview of all function keys Dentsply Sirona 10.6 Sivision program Operating Instructions Sinius Sivision program 10.6 10.6.1 Media Player Start Media Player Media Player is started on the external PC. The keys corresponding to the Media Player PC application are displayed on the right side of the...
Page 285
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.6 Sivision program 10.6.4 Sidexis Starting Sidexis Sidexis is started on the PC. The keys for the Sidexis PC application are displayed on the right side of the touchscreen: Next image The next image window is activated.
Page 286
10 Overview of all function keys Dentsply Sirona 10.6 Sivision program Operating Instructions Sinius Sharpen image Contrasts between neighboring pixels are increased. This function helps to accentuate edges or contours. The impression of a sharper image is created. Invert image This function inverts the brightness values of the image pixels, thus enabling a positive or negative display of the image.
Page 287
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.6 Sivision program 10.6.5 Video plugin Start the video plugin Sidexis 4 and the video plugin are started on the PC. The fixed keys for the video plugin are displayed on the right side of the touchscreen.
Page 288
10 Overview of all function keys Dentsply Sirona 10.7 Setup program Operating Instructions Sinius Setup program 10.7 10.7.1 User interface Configuring the user interface Opens the User Interface setup program. Key tone A setting can be made to activate or deactivate an acoustic signal that sounds when the operator touches a key on the touchscreen.
Page 289
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.7 Setup program Start program operating modes The Start program can run in two operating modes. In the Standard Start program operating mode, the functions of the patient chair and instruments are each shown on separate screens.
Page 290
Central supply for chemical suction hose cleaning For clinical use, Sinius treatment centers can be equipped with a central cleaning agent supply for chemical suction hose cleaning. The function can be switched on and off.
Page 291
Dentsply Sirona 10 Overview of all function keys Operating Instructions Sinius 10.7 Setup program One of the following presettings can be selected: ● Field 1: Quick setting keys with SaveMode ● Field 2: Quick setting keys with DropMode ● Field 3: Function levels...
Index Dentsply Sirona Operating Instructions Sinius Index Assistant element Numeric Instrument positions, 46 4-way foot control, 194 Maximum load capacity, 149 4-way foot control plate, 62, 65 Mini L.E.D., 226 Mini LED, 155 Accessories, 270 Overview, 45 Acoustic signals, 109 Sprayvit M, 122, 227 Adapter, 232 Suction handpieces, 152, 234 Additional devices, 50 Use of blasting devices, 18...
Page 293
Dentsply Sirona Index Operating Instructions Sinius Camera, 20, 173, 175, 227 Default settings, 190 Use with SI video, 178 Dental unit plugins Use with Sidexis, 180 ApexLocator distance display, 108 Canadian Standard Association (CSA), 23 Time synchronization , 192 Capacitive sensor, 82, 84 Video plugin, 188 Care, cleaning, and disinfecting agents, 203 Dentist element CDS 60, 196...
Page 294
Filter Function levels, 90, 198 Air and water, 241 Gold trap, 237 Standard wet suction, 248 Gold trap, 237 Suction hoses, 235 Guide roller on the Sinius CS dentist element, 94 Firmware, 12, 190 Guide rollers, 94 Fixed key Bell, 195 Handles on dentist element, 82, 84 Chair programs, 150 Handles on the dentist element, 208 Clean, 89...
Need help?
Do you have a question about the Sinius and is the answer not in the manual?
Questions and answers
How to programme the water to off for the slow handpiece so I don’t have to manually turn it off each time I use it.