10 PROGRAMMING

Note.
The following programming pages apply to both display units.
UPPER DISPLAY
Range 1
Secure Parameters
Upper Display
Operating Page
xxx.x
xxx.x
%H2 IN AIR
Alarm 1 Setpoint
Alarm 2 Setpoint
Lower Display
NOT IN USE
Secure Parameters
Fig. 10.1 Overall Programming Chart for 6553 Display Unit (Range 1) – AK101 and AK104
Access to
SECURITY CODE
Language Page
English
Note. All parameter values shown on the upper
display line are the default settings.
Set Up Outputs Page
SET UP OUTPUTS
A1 Action EA
EB
A1 Setpoint
A2 Action EA
EB
A2 Setpoint
RTX Type
4-20
0-20
0-10
Test Retrans (%)
100.0
75.0
50.0
25.0
0.0
Range 85% -100
Range 80% -100
Alter Sec. Code
Electrical
Calibration Page
ELECTRICAL CAL
Calibrate
No
Yes
Yes
No
xxx.x
mV Zero (0.00mV)
xxx.x
mV Span (10.00mV)
Adjust RTX Zero
Adjust RTX Span
31