6 Dispensing a treatment
6.1 Accessory usage conditions
The accessories and the handpiece must be cleaned, disinfected and sterilised prior to each use.
Refer to the cleaning, disinfection and sterilisation instructions for accessories listed in the chapter Associated
documentation page 5.
6.2 Preparation for use
To prepare your medical device, follow the steps below:
1. Wear safety goggles and protective gloves.
2. Clean the unit with an alcohol disinfectant wipe.
3. Remove the handpiece from its sterilisation bag.
4. Remove the wrench from its sterilisation bag.
5. Remove the tip from its sterilisation bag.
6. Screw the tip onto the handpiece, first manually and finishing with the wrench.
7. Connect the handpiece to the handpiece cord socket.
8. Place the handpiece on its support. .
9. Switch on the medical device.
10. Check the irrigation parameters depending on the tip chosen.
11. After water drainage, check that the spray works correctly.
Your medical device is now ready to use.
User Manual | Newtron
®
P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 15/38