The NI USB-9211 consists of two components: an NI 9211 module and an NI USB-9161 carrier, as shown in Figure 1. The NI USB-9211A consists of two components: an NI 9211 module and an NI USB-9162 carrier, as shown in Figure 1.
Dimensions Figure 2 shows the NI USB-9211/9211A device dimensions. Hi-Speed USB Carrier NI USB-9162 140.23 mm 120.68 mm (5.521 in.) (4.751 in.) 137.82 mm 118.26 mm (5.426 in.) (4.656 in.) 88.12 mm 25.34 mm (3.469 in.) (0.998 in.) Figure 2. NI USB-9211/9211A Device in Millimeters (Inches) Safety Guidelines Operate the NI USB-9211/9211A only as described in these operating instructions.
Start»All Programs»National Instruments»NI-DAQ»DAQ Getting Started Guide. The NI-DAQ Readme lists which devices are supported by this version of NI-DAQ. Select Start»All Programs»National Instruments»NI-DAQ» NI-DAQ Readme. The NI-DAQmx Help contains general information about measurement concepts, key NI-DAQmx concepts, and common applications that are applicable to all programming environments.
LabVIEW If you are a new user, use the Getting Started with LabVIEW manual to familiarize yourself with the LabVIEW graphical programming environment and the basic LabVIEW features you use to build data acquisition and instrument control applications. Open the Getting Started with LabVIEW manual by selecting Start»All Programs»National Instruments»LabVIEW»LabVIEW Manuals or by navigating to the directory and opening...
Select Start»All Programs»National Instruments»NI-DAQ»NI-DAQmx Help. The NI-DAQmx C Reference Help describes the NI-DAQmx Library functions, which you can use with National Instruments data acquisition devices to develop instrumentation, acquisition, and control applications. Select Start»All Programs»National Instruments»NI-DAQ» NI-DAQmx C Reference Help.
The NI-DAQmx Base Getting Started Guide describes how to install your NI-DAQmx Base software, your NI-DAQmx Base-supported DAQ device, and how to confirm that your device is operating properly. In Windows, select Start»All Programs»National Instruments»NI-DAQmx Base» Documentation»Getting Started Guide. Getting Started with NI-DAQmx Base for Linux and Mac Users describes...
Example Programs The NI-DAQmx CD contains example programs that you can use to get started programming with the NI USB-9211A. Refer to the NI-DAQmx for USB Devices Getting Started Guide that shipped with your device, and is also accessible from Start»All Programs»National Instruments»...
Mounting the NI USB-9211/9211A to a Panel Threaded inserts are located in the NI USB-9211/9211A for mounting it to a panel. Refer to Figure 5 for dimensions. 85.7 mm (3.37 in.) 72.2 mm (2.84 in.) Threaded Insert M3 x 0.58.5 mm (0.34 in.) Max Depth 76.1 mm (3.00 in.)
Page 13
Shield Thermocouple NI USB-9211/9211A TC– Figure 6. Connecting a Thermocouple Input Signal to the NI USB-9211/9211A Refer to Table 3 for the terminal assignments for each channel. Table 3. Terminal Assignments Module Terminal Signal TC0+ TC0– TC1+ TC1– TC2+ TC2– TC3+ TC3–...
NI USB-9211/9211A Circuitry The NI USB-9211/9211A channels share a common ground that is isolated from the chassis and the host computer. Each channel has an impedance between the TC+ and COM terminals and between the TC– and COM terminals. Each channel is filtered and then sampled by a 24-bit analog-to-digital converter (ADC).
15 °C or more than 35 °C, use this channel to read the offset error. The NI USB-9211 always has this channel enabled. Use NI-DAQmx to configure the behavior of the autozero channel for the NI USB-9211A. Refer to the LabVIEW Help for 8.0 or later, accessible from Start»All Programs»National Instruments»LabVIEW 8.5»LabVIEW Help, and...
Page 17
Max over Temp Range with Autozero Max at Room Temp Range without Autozero Typ over Temp Range with Autozero 1000 1200 1400 1600 1800 Measured Temperature (°C) Figure 10. Type B Errors Max over Temp Range with Autozero Max at Room Temp Range without Autozero Typ over Temp Range with Autozero –200 1000...
Max over Temp Range with Autozero Max at Room Temp Range without Autozero Typ over Temp Range with Autozero –200 1000 1200 1400 1600 Measured Temperature (°C) Figure 14. Type R and S Errors Cold-Junction Temperature Measurement Accuracy Heat from other nearby heat sources can cause errors in thermocouple measurements by heating up the terminals so that they are at a different temperature than the cold-junction compensation sensor used to measure the cold junction.
Max sampling rate (Hz) Type of Measurement Number of Temperature Temperature Channels in with CJC and with Analog Scan List Autozero Autozero Input Note: NI-DAQmx Base always applies autozero. Input bandwidth (–3 dB)......15 Hz Noise rejection ........85 dB min at 50/60 Hz Overvoltage protection ......±30 V between any input and common Differential input impedance ....20 MΩ...
Bus Interface USB specification NI USB-9211 ........USB 2.0 Full-Speed NI USB-9211A ....... USB 2.0 Hi-Speed Physical Characteristics Dimensions ..........14 cm × 8.6 cm × 2.5 cm (5.51 in. × 3.37 in. × 0.99 in.) Weight ............ Approx. 350 g (12.3 oz) Screw-terminal wiring......
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet (for example, 115 V for U.S. or 230 V for Europe). Safety Standards This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control,...
Waste Electrical and Electronic Equipment (WEEE) At the end of their life cycle, all products must be sent to a WEEE recycling EU Customers center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm RoHS...
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents.