Testing the System

Testing the System
The system test is a comprehensive test of the system operational status. This test includes
numerous subtests. If a subtest fails, the system completes the remaining subtests. The system
test displays only a pass-fail result on the system monitor. If the system test fails, notify your
Philips service representative.
Run the system test any time a system error is displayed, or if you suspect problems with the
system. If an error message is displayed during the test, restart the system with the
Off) control.
NOTE
The system test can take several minutes to run.
1. Turn on the system.
2. Disconnect all transducers from the system.
3. Press Support.
4. In Philips SupportConnect, click the Test/Utilities tab.
5. Under Navigator, click System Test.
6. In the System Test area, click Run. The system displays a message when the test is
complete, indicating whether the test passed or failed.
7. If the test fails, contact your Philips service representative for instructions about how to
export the log files.
8. Restart the system as explained in

For Assistance

If you are unable to correct a problem, call your local Philips representative.
EPIQ 7 User Manual 4535 617 25341
To send a test pattern to a local printer, double-click a test pattern to display it full-
screen, and touch Alt Print.
To send a test pattern to a DICOM printer or archive server, click a test pattern
number to select it, click Send To, select a device, and click OK.
"Turning the System On and Off" on page
System Maintenance
(On/
113.