Transcend 365 miniCPAP Auto
Check the reservoir is at its full capacity by holding it vertically and viewing the
3.
water level. The water should be level with the max fill line (bottom of the
reservoir seal).
Repeat steps 2 and 3 as needed to fill and check the distilled water level in the reservoir.
The reservoir should be completely filled before each use.
Once the reservoir is filled completely, set it on a secure, flat surface and line up
4.
the CPAP heater cartridge with the opening in the reservoir.
place and set the CPAP onto it so that the magnets snap the pieces together
securely.
TM
User Manual
Reservoir should
be filled to the
'Max Fill' line.
Line up the heater
cartridge with the
opening in the
reservoir.
Hold the reservoir in
Page 22

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 365 miniCPAP Auto and is the answer not in the manual?

Questions and answers

Jackie Mason
April 2, 2025

I have a red water droplet and !!! Heater Error !!! in red How do I find the code?

1 comments:
Mr. Anderson
April 2, 2025

To resolve the red water droplet and "!!! Heater Error !!!" message on the Transcend 365 miniCPAP Auto:

1. Red Water Droplet:
- This indicates the humidifier is disabled due to a condition such as:
- Use of water other than distilled.
- Residue from previously used tap water.
- Corrupt water signal, likely from moisture inside the CPAP.
- Solution:
- Empty the reservoir and rinse thoroughly with distilled water.
- Refill with only distilled water.
- If issue persists, run the CPAP in Dry Mode to remove internal moisture.
- After drying, disconnect and reconnect the power source to reset the system.

2. Heater Error:
- This may appear together with the red droplet and indicates the humidifier heater is disabled.
- Solution:
- Follow the same steps as above to address possible moisture or water quality issues.
- If error continues, contact your medical equipment provider or customer service for further assistance.

Using only distilled water and keeping the system dry helps prevent these errors.