Preparing Mini Field Agent For Installation And Configuration; Mounting Information - GE Mini Field Agent Instructions For Installation And Configuration

Hide thumbs Also See for Mini Field Agent:
Table of Contents

Advertisement

Preparing Mini Field Agent for Installation and Configuration

Hardware required
Software required

Mounting Information

Carestation Insights Cloud information
2
Description
Carestation Insights Privacy and Security Manual 2094533-001
Carestation Insights Cloud Gateway Hardware
and Software Kit (contains necessary MFA
hardware with software installed, as well as
manuals)
Carestation Insights USB drive with software
(contains software only for MFA, used for
upgrade or fix for existing MFA hardware)
The following hardware is required for the initial set up:
One Mini Field Agent device.
One computer with a network port, WinSCP and Putty
applications installed.
Two RJ45 ethernet cables.
Note
Ethernet cables are not included in the Carestation Insights kit.
One cable is required to be left connected to the MFA. The other
cable is used for configuration only. Contact the Hospital/Facility
network team for ethernet cables, if necessary. Note that each
Aisys CS2 software version 11 with connectivity and Aisys
software version 11 upgrade kit comes with 2 ethernet cables.
The Mini Field Agent installation and configuration requires the use
of open source WinSCP and PuTTY applications. Download these
from:
https://winscp.net/eng/download.php
www.putty.org
For Mini Field Agent mounting instructions see "MFA Mounting
Instructions" in the Appendix 2 of this Manual.
Obtain the following information from the GE representative:
Carestation Insights Cloud Service information
CLOUD_URL
Part number
2094209-001
2095043-001
2094532-001 D 02 2018

Advertisement

Table of Contents
loading

Table of Contents