Message in the display
The CP42CP52 hard disk
is full!
Error during configuration
of channels x, y, z!
The following channels
have been configured:
x, y, z
This is the wrong hard
disk!
Overload
(In Status mode the red
"ERROR/WARN." LED on
the front panel of the
amplifier module is lit).
Calibration errors
(In Status mode the red
"ERROR/WARN." LED of
the front panel of the
amplifier module is lit).
MGCplus
6.2.3
Messages of AB22A/AB32
Cause
The capacity of the storage medium is
exceeded.
a) Amplifier types x, y, z do not match
the types of the configuration file (mod
ules mixed up).
b) There is no amplifier module present.
Information about successful
configuration of channels x, y, z
There is a file in the storage medium
(IdentTEDS.txt) containing the serial no.
CP42/
of a
CP52. This serial no. does
not match the serial no. of the
CP42/
CP52 that is present.
The set measuring range is exceeded
(gross, net or both).
a) No feedback lines connected and
Autocal was triggered.
b) Connection board has been changed.
c) RAM/EEPROM errors
d) Calibration line is not correct
A0534-30.0 HBM: public
Functions and symbols of the AB22A
Remedy
Insert a new storage medium or delete
data.
Check the fitting of the MGCplus device.
-
Check the assignment of the storage
CP42/
medium
CP52 and if necessary
delete the IdentTEDS.txt file (see also
the operating manual for the "MGCplus
Assistant; Serial number of
CP42/
CP52").
a) Check the values in zero or tare
memory.
b) Perform a zero balance or taring.
c) Zoom in on the measuring range.
a) Check the transducer connection and
trigger an autocalibration.
b) Perform a setup.
c) Perform a setup.
d) Check the transducer settings (trans
ducer type, excitation voltage, output
characteristics).