ABB Navigator 550 Commissioning Instructions page 8

Hydrazine wet-section
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Solution preparation
Reagent Solution –
5M (20% W/V) Sodium Hydroxide
Warning. Sodium Hydroxide is extremely caustic and must be
handled with great care. Wear gloves and eye protection.
The solution used to fill the reagent container is made up as shown
below – consumption is approximately 250 ml in 2 to 4 weeks.
1. Weigh out 2.5 (±0.1) g EDTA and transfer to a 500 ml (16.9 fl
oz) measuring flask (a little high purity water can be used to
help this transfer).
2. In a separate vessel, weigh out 100 (±1) g sodium hydroxide,
NaOH, pellets (analytical reagent grade) and dissolve in
approximately 300 ml (10.0 fl oz) high purity water in a plastic
container. Allow this solution to cool.
3. Transfer this solution to the measuring flask, shake well to
dissolve the EDTA and top up to the mark with more high
purity water.
Standard solution
Warning. Hydrazine sulphate is an irritant to skin and eyes.
Avoid breathing the dust. Wear gloves, eye protection and a
dust mask when handling this substance.
Choose a convenient value for the hydrazine concentration of the
standard solution – typically 50 or 80 µg kg
can be used if required.
Note. Hydrazine solutions deteriorate with time: replace the
stock solution at monthly intervals. Dilute standard solutions
must be freshly prepared.
Prepare a stock solution of 1000 mg l
1. Weigh out 4.058 (±0.001) g analytical reagent grade
hydrazine sulphate (N
H
2
approximately 800 ml (27.0 fl oz) high purity water.
2. Transfer to a 1 l (33.81 fl oz) volumetric flask and top up to
the mark with more high purity water.
3. Dilute the stock solution to provide the required standard
solution for the particular measuring range (usually 30 or
-1
80 µg kg
).
-1
. Other concentrations
-1
hydrazine as follows:
,H
SO
) and dissolve in
4
2
4
ABB Limited
Measurement & Analytics
Oldends Lane, Stonehouse
Gloucestershire,
GL10 3TA
UK
Tel: +44 (0)1453 826661
Fax: +44 (0)1453 829671
Email: instrumentation@gb.abb.com
ABB Inc.
Measurement & Analytics
125 E County Line Road
Warminster, PA 18974
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183
abb.com/measurement
We reserve the right to make technical changes or modify the contents of this document
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in
whole or in parts – is forbidden without prior written consent of ABB.
© ABB 2018
3KXA835550R4401

Advertisement

Table of Contents
loading

Table of Contents