Connecting the sample lines

Note.
Sample inlet tubing is customer-supplied.
Keep sample drains as short as possible and vertical to
enable the sample to drain freely.
Where particulate matter is present (for example,
magnetite in boiler samples) it is recommended that a 60
micron sample filter is fitted to the sample inlet line.
To make external sample inlet connections:
1. Fit a shut-off valve (not supplied) at each sample inlet.
Referring to Fig. 3:
2. Connect the sample inlet tubing
1
(
/
in. ID non-flowmeter or
4
B
connector
at the base of:
C
sample inlet
(wet-section without flowmeter)
or
D
sample inlet
(wet-section with flowmeter)
3. Ensure the drain tubing
is routed to a suitable drain.
4. Ensure the 2 drain tubes
are kept as straight as possible and routed through the funnel
H
at the base of the wet-section. Drain tubes are
factory-fitted but may have moved in transit.
Fig. 3 Connecting the sample lines
CI/AHM550–EN Rev. D
A
using flexible PVC tubing
3
/
in. ID with flowmeter) to barbed
8
E
is kept as straight as possible and
F
from the constant-head unit
5
Connecting the standard solution
2 (high) and reagent tubing
Connections to the standard solution 2 (high) and reagent
containers (at the QD couplings) must be made on site. All other
internal wet-section tubing connections are factory-made.
Note. If reagent and standard solutions have not already been
prepared, proceed to Solution preparation (overleaf).
Referring to Fig. 44:
1. Fill the standard solution 2 (high) and reagent containers with
the correct solutions – refer to Solution preparation.
2. Connect the standard solution 2 (high) tubing QD coupling
A
plug
to the mating connector
standard solution 2 (high) container.
3. Connect the reagent solution tubing QD coupling plug
the mating connector
container.
G
Standard Solution 2
(High) tank – 400 ml
Fig. 4 Connecting the standard solution 2 (high) and reagent tubing
QD couplings
B
at the base of the
D
at the base of the reagent solution
Filler caps
Reagent Solution
tank – 200 ml
C
to
5