Returning Blood To The Patient Using Assisted Reinfusion - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Chapter 4—Monitoring the Treatment

Returning Blood to the Patient Using Assisted Reinfusion

Figure 83 – Assisted Reinfusion Button on the Test & Options Screen
If the 'Assisted Reinfusion' option is set in Service Mode (functional software version 2.70 or
later), the Assisted Reinfusion button is displayed on the "Test & Options" screen. This
button becomes available when blood is sensed and is used to aid the operator in returning all
of the patient's blood in the bloodlines. Pause the Tx Clock then select the Assisted
Reinfusion button. This will stop the blood pump and display the first "Assisted Reinfusion"
screen.
To begin returning the patient's blood, select the Assisted Reinfusion button.
Note: The Assisted Reinfusion button is unavailable when either Low Volume or Single
Needle options are set to 'On'.
Note: This procedure is shown using CombiSet bloodlines P/N 03-2722-9; if a different
bloodline set is used, the clinic is responsible for providing instructions.