Chapter 4—Monitoring the Treatment
Returning Blood to the Patient Using Assisted Reinfusion
Figure 83 – Assisted Reinfusion Button on the Test & Options Screen
If the 'Assisted Reinfusion' option is set in Service Mode (functional software version 2.70 or
later), the Assisted Reinfusion button is displayed on the "Test & Options" screen. This
button becomes available when blood is sensed and is used to aid the operator in returning all
of the patient's blood in the bloodlines. Pause the Tx Clock then select the Assisted
Reinfusion button. This will stop the blood pump and display the first "Assisted Reinfusion"
screen.
To begin returning the patient's blood, select the Assisted Reinfusion button.
Note: The Assisted Reinfusion button is unavailable when either Low Volume or Single
Needle options are set to 'On'.
Note: This procedure is shown using CombiSet bloodlines P/N 03-2722-9; if a different
bloodline set is used, the clinic is responsible for providing instructions.
170
2008T Machine Operator's Manual P/N 490122 Rev Q