Quick Guide to Turbo Vap
Endpoint Selections (1)
Choose between four concentration modes:
»
Time: Concentrate for the time specified by the
Endpoint Time setting on the panel. (3)
»
Sensor: Concentrate to an end volume (0.5 or 0.75
mL depending on the style of concentrator tube).
»
Sensor and Time: Concentrate past the 0.5 or
0.75 mL setting by the amount of time specified
by the Endpoint Time setting. (3)
»
Manual: Manually control the concentration with
the Start/Stop button for individual cells. (4)
See pages 3–4 of the Operator's Manual for more information.
Bath Temperature (2)
Set water bath temperature using the pushwheel. Must be
less than 60° C if sensors are installed. The display blinks if
temperature deviates ±2 °C from 60 °C.
Endpoint Time (3)
Use this to set the time for either the Time or Sensor and Time
endpoint selections. (1)
Cell Selections (4)
Start or stop the concentrator cell. The lights indicate the status
of the cell.
»
On: concentrating
»
Off: not concentrating
»
Blinking: endpoint reached
Displayed Conditions (5)
Action (Selected)
All
Pressure
Temperature
Time Elapsed
Diagnostics
Table 1. Select and view the current status of a particular condition.
© Biotage 2013
II
®
Condition
Value
##
Pi
#.#
br
##
°C
#=
##
SE
––
––
––
TurboVap
1
TurboVap
II
®
Time (minutes)
Sensor
Endpoint
Select
Sensor and Time
Manual
Display
2
3
Condition
Value
Bath
Endpoint
Temperature
Time
Start/Stop
4
Cell One
Cell Two
Cell Four
Cell Five
Figure 1. TurboVap
II front panel.
®
Description
Display cycles through pressure, temperature,
and time.
Pressure in psi.
Pressure in bars (alternates).
Temperature in degrees centigrade.
Tube position number and time in minutes.
Display during sensor portion of Sensor and
Time concentration.
If time exceeds 99 minutes.
If concentration stopped or not started.
Varies, see Section 8 of the Operator's Manual.
II Quick Guide | Page 1
®
Concentration
Workstation
5
All
Pressure
Temperature
Time Elapsed
Diagnostics
Select
Displayed
Condition
Cell Three
Cell Six

Advertisement

loading
Need help?

Need help?

Do you have a question about the Turbo Vap II and is the answer not in the manual?

Questions and answers

Summary of Contents for Biotage Turbo Vap II

  • Page 1 Time concentration. If time exceeds 99 minutes. –– –– If concentration stopped or not started. –– Diagnostics Varies, see Section 8 of the Operator’s Manual. Table 1. Select and view the current status of a particular condition. © Biotage 2013...
  • Page 2 Part Number: UI312 © 2013 Biotage. All rights reserved. No material may be reproduced or published without the written permission of Biotage. Information in this document is subject to change without notice and does not represent any commitment from Biotage. E&OE.