Thank you for buying the INFICON D-TEK Select Refrigerant Leak Detector! With normal use, your D-TEK Select from INFICON will give you years of trouble-free service. To get the best performance from your D-TEK Select, please read this manual carefully before you start using the instrument. If you have any questions or need additional assistance, please call 800-344-3304.
Getting Started The D-TEK Select is shipped with the infrared cell and power stick installed. The nickel/metal hydride batteries in the power stick are not yet charged. To charge them, connect the cord of the AC power adapter to the power jack at the rear of the instrument, and plug the adapter into the appropriate mains outlet.
3. When the instrument detects a leak source, its yellow LEDs will illuminate and it will beep rapdily. 4. When the D-TEK Select signals a leak, pull the probe away from the leak for a moment, then bring it back to pinpoint the location.
Increases in concentrations of refrigerants will be displayed by the illumination of the yellow LEDS. For best test results, allow the D-TEK Select to warm up for 5 - 10 minutes before engaging manual zero mode.
LED and the amber battery charging LED will flash. Recharge the D-TEK Select by connecting the supplied AC power adapter, or using the cigarette lighter adapter. NOTE: The D-TEK Select does not need to be fully charged to operate nor fully discharged before recharging. W A R N I N G Do not attach any other power cords to the D-TEK select’s power...
This specialized IR cell will operate for about 1000 hours. The D-TEK Select will flash all the yellow leak LEDs when the IR cell has reached the end of its useful life. To replace the IR cell: 1.
Disengage the connector and slide the power stick out of the D-TEK Select. Slide the new stick into the D-TEK Select and push the miniature connector on the power stick into the mating half on the PCB board. Do not bend the PCB board’s connector away from the board.
Replacement Parts and Accessories Replacement parts and accessories for your D-TEK Select are available through the same dealer from whom you bought the instrument. Molded plastic storage case....... . . 712-702-G1 Headphones .
90 seconds to ready to use. see if the unit starts to beep and the lights stop scrolling. If not, contact INFICON. 2d) Battery may be 2d) See #3 below. dead. 2c) Filter cartridge 2c) Replace used filter may be clogged, cartridge with a new one.
MERCHANTABILITY or of FITNESS FOR A PARTICULAR PURPOSE or otherwise. All such other warranties are expressly disclaimed. INFICON shall have no liability in excess of the price paid to INFICON for the instrument plus return transportation charges prepaid. INFICON shall have no liability for any...
Return Materials Authorization Procedure All instruments and parts returned to INFICON for repair or credit must be properly packaged, insured, shipped transportation charges prepaid, and must have a Return Material Authorization (RMA) number issued before the material is returned. The RMA number is to be marked on all shipping labels and packing slips.