Sanyo MLS-3751L Service Manual page 32

High-pressure steam sterllizer labo autoclave

Advertisement

Select the instrument sterilization
cycle and set the sterilization
temperature at 135'C and the
sterilization time at 30 min. Close
the cover and press the
STRAT
~.
Set the cycle, sterilization temp.
and time.
OC
and sterilization process
Reaching 135
Press the "V"or
",6."
button
while pressing the
SET lCONFIRM
button
and
adjust the pressure so that the
pressure display becomes same
as the value on the pressure
gauge of the product.
When the
START
button is
pressed While pressing the
SET /DISPLA Y
button
I
the
corrected value is written in.
Pressure adjustment
(Correction)
(See the Mght figure.)
The pressure gauge of the
product shows the pressure in
MPa and the pressure display
value in kPa.
Pressure adjustment
(WMting-in)
-31-
Nonmal display
Release.
SET/COFIRM
Press
Example of pressure
correction display
button
Pressure
Corrected
display value
value