Download Print this page

Sanyo MLS-3751L Service Manual page 26

High-pressure steam sterllizer labo autoclave

Advertisement

@Processing frequency
Total frequency
In the process frequency display, the total frequency of all cycles
and the processing frequency of each cycle are displayed.
• Each time when the "TIME" button is pressed, the processing
frequency of each cycle are displayed.
R
··ALL" stands for the total frequency, and "cSl" for LlQUD STERI,
"cS2" for STERI KEEP WARM, "cS3" for MERT/KEEP WARM
and "cS4" for INSTRUMENT STERI, respectively.
C cle-wise
Processing
frequency
·When the "START" button is pressed while pressing the
·SET/CINFIRM" buttons, the total processing frequency and
operation frequency of each cycle are cleared.
C
S ,
"--r--I~Cycl~e---.J1
1"""''--'' ......
_..0.....,=1
L -_ _ _
--=' ' - ' ' ' O .....,H
~~c:~~
I
@Hlstory of errors
Error No., error code and error occurring process No. are displayed.
The latest error data is displayed in the first place.
• The error No. is changed by pressing the" ... ·or .... " button and
the history of errors can be checked.
The error No. can be recorded for 8 in number at the maximum.
If more than 9 errors occur, the oldest error is deleted and new
error is recorded instead.
E
6
t
·When the "START" button is pressed while pressing the
·SET/CONFIRM" button, the error data is deleted.
(Total errors are not deleted.)
I
Error code
I
.Error Codes.
2
+-t
Error
No.
I
04--
Error occurring
process No.
1 : Lateral thermistor (TH1) Open
2 : Bottom thermistor (TH3) Open
6 : Movable handle Unlock impossible
7 : Movable handle and lid Lock impossible
3: Lateral thermistor Abnormal temperature
8: Abnormally high pressure
4 : Object temperature sensor (TH2) Abnormal temperat 9: Low pressure
5: Bottom thermistor Abnormal temperatura
10: EEPROM communication error
<
Error occurring process Nos.>
Process No Error Occurring Process
Description
Process No
0
Stand-by
6
1
Reservation
7
2
Heating 1
Up to 90"C
8
3
Heating 2
From 90<>C
to
boiling poin1
9
4
Exhaust
10
5
Heating 3
THl
temp. rises up to
11
the preset temp.
@Boiling Poing setting
Change the boiling point temperature according to the altitude of
the place for installation by pressing the" ... ·or ..... button.
·When the 'START" button is pressed while pressing the
·SET/CONFIRM" button, the content of the change is written in
the non-volatile memory.
(The boiling temperature has been set at 100"C in the factory.)
Error Occurring Process
Description
Heating 4
TH2
temp. rises up to
the preset temp.
SterilizationlMelting
Cooling 1
Down
to
boiling point
Cooling 2
Before cooling point
Keep warm
Completion
bP
,
~OH
Set temperature
I
• When setting the boiling point, refer to the table (P __ ) of the altitude and atmospheric pressure in the
standard atmosphere and the boiling point of water.
- 25-

Advertisement

loading

This manual is also suitable for:

Mls-3751Mls-3781Mls-3781l