Message
Purpose of Message
Crit-Line needs
The CLiC device has not been
verification
verified within the past 30 days. The
Hct/BVCrit-Line graph on the "Crit-
Line" screen will be disabled.
Crit-Line: No Blood
The Tx Clock is running and the CLiC
device no longer senses blood in the
Blood Chamber.
185
Troubleshooting
Type
Action Required
Advisory
Verify the CLiC device:
message
If the message is not cleared:
If the message is still not cleared, the CLiC device cannot be used for
the dialysis treatment.
Warning
Check the Crit-Line Blood Chamber:
If the message is not cleared, the CLiC device cannot be used for the
dialysis treatment.
ALARMS
TROUBLE-
SHOOTING
Place the CLiC device on the CLiC Verification Filter, which is
attached to its USB cable.
Wait up to one minute for the message to clear. If the
message clears, the CLiC device is verified.
Select the "Test & Options" screen-button then select the
Verify Crit-Line button to start the verification process
manually.
Wait up to one minute for the message to clear.
Disconnect the CLiC device cable from the USB port.
Contact a qualified service technician.
Make sure there is proper blood flow: There should be no air
or bubbles in the chamber.
Make sure the CLiC device is properly seated on the Crit-
Line Blood Chamber: The device should be placed
perpendicularly over the chamber and cannot be rotated up
or down.
Remove the device from the Crit-Line Blood Chamber.
Disconnect the CLiC device cable from the USB port and
press the RESET key to disable CLiC device.
Contact a qualified service technician.
2008T Machine Operator's Manual P/N 490122 Rev N
Need help?
Do you have a question about the 2008T and is the answer not in the manual?
Questions and answers
Fresenius 2008 t machine with b bag options do the filters and o rings on the b bag need to be change on each annual pm and semiannual pm
What is the policy or steps if a Machine misses the weekly bleach by 1 day?
TMP low on t machine
A low TMP (Transmembrane Pressure) on a Fresenius Medical Care 2008T machine may indicate a potential issue with the balancing system or a possible leak. Adjusting TMP repeatedly can decrease the ultrafiltration (UF) removed from the patient. If TMP is rising, it should be investigated for leaks in the system.
This answer is automatically generated
WHAT IS A BIC PUMP
The Bic pump in the Fresenius Medical Care 2008T Hemodialysis Machine is used to manage the delivery of bicarbonate concentrate during treatment.
This answer is automatically generated