17. CARE AND CLEANING

To ensure accurate measurements, it is very important
to keep the infra-red sensor clean and free of scratches.
Fingerprints and dirt will affect the accuracy of the
thermometer. In order to get an accurate measurement,
please clean the infra-red sensor. Gently wipe its surface
with a cotton swab slightly moistened with alcohol and
immediately wipe dry with a clean cotton swab.
Use a soft dry cloth to clean the body of the
thermometer. Never clean the thermometer using an
abrasive cleanser, thinner or benzene, nor immerse the
thermometer in water or other liquids. After cleaning,
place the thermometer in its protective case. Store it in a
clean, dry place at room temperature. Never expose the
thermometer to extreme temperatures, humidity, direct
sunlight or shock.

17.1 CALIBRATION

The thermometer is initially calibrated during
manufacturing, before ex-factory. If the thermometer
is used in accordance with the instructions, periodic
recalibration is not required. Should you question the
accuracy of measurements at any time, contact the
retailer immediately.
Do not attempt to modify or reassemble the
thermometer.