Sartorius arium mini plus Operating Instructions Manual page 84

Ultrapure water system
Hide thumbs Also See for arium mini plus:
Table of Contents

Advertisement

Error 0180
The fill level of
the bag is not
being properly
Please contact
detected.
your service
technician.
Error
The conductivity
of the ultrapure
water is outside
Conductivity
of measuring
measurement of
the ultrapure
range.
water is out of
range.
The temperature
Error
of the ultrapure
Temperature
water is outside
measurement of
of measuring
the ultrapure
range.
water is out of
range.
Error
The limit has
been exceeded
and the
Ultrapure water
dispensing lock
quality
> XX μS/cm
is preventing
further
dispensing.
t
The bag is damaged
Remove the left side cover and check
or improperly
the bag for damage. If required:
connected.
Change bag.
t
Check that the three bag hoses are
properly connected to the device
connectors. If required: Remove the bag
and reconnect it.
t
The sensor is
Using the tank outlet tubing connected
to "Bag Outlet", let about 4 of the pure
defective.
water out of the bag and check the level
change.
t
If required: Call the Sartorius Service.
t
There is air in the
Initiate venting.
ultrapure water
circulation.
t
The ultrapure water
Replace the ultrapure water cartridge.
cartridge has been
t
To dispense water for testing purposes:
used up.
Disable the dispensing lock.
t
If required: Call the Sartorius Service.
t
The temperature of
Check if the ambient temperature
the ultrapure water
complies with the device specifications
is outside of
in the technical data.
measuring range.
t
Dispense about 1 liter:
− To dispense water for testing
− Dispense and discard about 1 liter of
t
If required: Call the Sartorius Service.
t
There is air in the
Initiate venting.
ultrapure water
t
Replace the ultrapure water cartridge.
circulation.
t
To dispense water for testing purposes:
Disable the dispensing lock.
t
If required: Call the Sartorius Service.
t
The limit has
Check the limit. If required: Configure
been configured
the limit.
incorrectly.
Operating Instructions arium
purposes: Disable the dispensing lock.
water.
®
mini plus | arium
Malfunctions
8.6.1, page 66
8.6.1, page 66
7.2.4, page 58
8.9, page 80
8.6.3, page 63
11.2, page 88
8.9, page 80
8.6.3, page 63
®
mini
83

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the arium mini plus and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Arium mini

Table of Contents

Save PDF