Malfunctions
9
Malfunctions
9.1
Error Messages
If an error message is active, dispensing is canceled and locked automatically.
Error message
Fault
Measured values
Error 0105
are not being
displayed.
Error 0140
The UV lamp is
not recognized.
Check UV lamp,
replacement may
be needed.
Error 0150
The device has a
leak.
Check device
for leakage.
Please contact
your service
technician.
Error 0160
Operation of the
device is not
Front cover
possible.
is open.
Please put the
front cover back
on.
Error 0163
The pretreatment
cartridge is not
Check proper
recognized.
fit of the
pretreatment
cartridge (R).
The ultrapure
Error 0166
water cartridge is
not detected.
Check proper fit
of the ultrapure
water cartridge
(L).
82
Operating Instructions arium
Cause
There is an internal
communications
error.
The UV lamp is not
connected properly
or defective.
Water has leaked
inside the device.
The front cover
is not properly
attached.
The pretreatment
cartridge has not
been properly
inserted.
The ultrapure water
cartridge has not
been properly
inserted.
®
®
mini plus | arium
mini
Correction
t
Disconnect the device from the AC
power and wait a minute.
t
Reconnect the device to the AC power.
t
If required: Call the Sartorius Service.
t
Check whether the black plug of the UV
lamp is connected correctly.
t
Remove the UV lamp and check for
damage. If required: Change the UV
lamp.
t
If required: Call the Sartorius Service.
t
Remove the left side cover and check
that the three bag hoses are properly
connected to the device connectors.
If required: Remove the bag and
reconnect it.
t
Drain the tray at the bottom of the
device and check that the connections
are tight during operation.
t
If required: Call the Sartorius Service.
t
Correctly attach the front cover to the
device.
t
If required: Call the Sartorius Service.
t
Press the pretreatment cartridge in
firmly until you hear a distinct clicking
sound.
t
If required:
− Remove the cartridge.
− Insert cartridge.
t
If required: Call the Sartorius Service.
t
Press the ultrapure water cartridge in
firmly until you hear a distinct clicking
sound.
t
If required:
− Remove the cartridge.
− Insert cartridge.
t
If required: Call the Sartorius Service.
Chapter
7.1, page 56
8.6.5, page 76
8.6.5, page 76
8.6.1, page 66
8.6.3.1, page 69
8.6.3.2, page 70
8.6.3.1, page 69
8.6.3.2, page 70