Summary of Contents for Sartorius arium mini plus
Page 1
Presented By Dallas Ft Worth Austin Houston NicolScales.com 800.225.8181 Contact Us Nicol Scales & Measurement is an ISO Accredited Calibration Company that has provided calibration, repair and sales of all types of weighing and measurement products since 1931.
Page 2
Operating Instructions arium mini plus | arium mini ® ® H2O-MA-... | H2O-MM-... Ultrapure Water System 1000025355...
Page 6
Contents 8.7 Enabling, Disabling or Configuring Reminders for Replacing Final Filters . .78 8.8 Carrying Out Depressurization ........79 8.9 Carrying Out Venting.
About this Document About this Document Introduction ® These instructions provide all the information necessary for operation of the arium mini (hereinafter referred to as the device). These instructions must be read, understood and used by all personnel using the device. Before working with the device: Read the instructions carefully and completely.
About this Document Symbols Used As instruction and direct warning of danger, all text statements in these operating instructions to be particularly noted will be marked as follows: This symbol denotes a possible danger with medium risk that death or severe injury may result if it is not avoided.
Safety Notes Safety Notes Intended Use The device is a water purification system that produces ultrapure water for daily requirements of up to 10 liters. The device is not suitable for use in hazardous environments. The device may only be used indoors. These instructions are part of the device.
Work on and modifications to the electrical equipment of the device may only be carried out by Sartorius Service personnel. The device may only be opened by Sartorius Service personnel. Check the device regularly for defects, such as loose connections or damage to the insulation.
Device Description Device Description Device Overview ® ® 3.1.1 Front View of the arium mini plus and arium mini Fig. 3-1: Front view Description Display with touch function Water outlet Final filter Front cover Side cover ® ® Operating Instructions arium mini plus | arium mini...
Electrical Connections of the arium mini plus and arium mini Fig. 3-2: Electrical connections Description Explanation “USB” connection For Sartorius Service “Power” connection For connection of the AC adapter (power supply) ® ® 3.1.3 Water Connections on the arium mini plus and arium mini Fig.
Device Description ® Sartorius arium mini systems were designed to produce ultrapure water. ® The arium mini plus is directly connected to the potable water supply and purifies the water in two stages. In the first stage, the arium ®...
Device Description 3.2.3 Tank Filling Tubing The tank filling tubing serves to feed pretreated water from an external tank to the arium ® mini. Fig. 3-7: Tank filling tubing 3.2.4 Drain Water Tubing The drain water tubing is used to drain unpurified water and rinse water out of the device.
Operating Design Operating Design Dispensing Mode In dispensing mode, the display shows information on water quality and the level of the bag as well as buttons relevant to dispensing. 0.055 21.3 °C & 0.25 Fig. 4-1: Dispensing mode Symbol Name Description Working The following information can be displayed:...
Operating Design Message Display The device displays three types of messages: − Error messages (errors): − Dispensing is not possible. − Details on the cause of error messages and how to troubleshoot them can be found in Chapter 9.1, page 82. −...
Operating Design Messages in Dispensing Mode If several messages are active in dispensing mode, the message list can be called up (see Chapter 4.4, page 18). The message list only appears if several messages are active and at least one message cannot be displayed for conductivity or temperature of the water.
Operating Design Message List All active messages can be viewed in the message list. The messages in the message list are sorted by priority. Error messages are on top. Within the same priority level, the messages are sorted by date and time. Messages cannot be deleted manually.
Operating Design Menu All system settings and operation steps for the care of the device can be carried out ® in the menu. On the arium mini, bag filling can be started via the menu. Fig. 4-5: Menu Symbol Name Description Dispensing Opens the dispensing screen...
Operating Design Navigating Menus ® A display with touch function is used to operate the arium mini and the arium ® mini plus. If multiple menu items are available, you can scroll up and down using the touch screen to select the desired entry. Sharp or pointed instruments (such as ballpoint pens) can damage the device! When using the touch screen with the aid of objects, the touch screen can be easily damaged.
Operating Design Numeric Keypad The numeric keypad is used to enter a dispensing volume or to configure various system settings. > Fig. 4-6: Numeric keypad (example: Entering the time) Symbol Name Description Name of the Displays the name of the current dialog box. dialog box Numerical value Displays the currently entered numeric value.
Measured values Displays the current water quality of the ultrapure water stage and pre-stage. Service Shows Sartorius contact information information and the next maintenance date. Displayed Changes the display units for values water quality and water temperature.
Page 24
Operating Design Menu Menu Submenu Description Reminder Displays the times for the next required consumables changes. Care Change Starts a bag change. consumable Pretreatment Starts a pretreatment cartridge cartridge change. Ultrapure water Starts an ultrapure water cartridge cartridge change. UV lamp Starts a UV lamp change.
Operating Design Parameters of the “Settings” Menu Parameters Setting values Explanation Language English German French Italian Spanish Portuguese Polish Russian Japanese Chinese Date format DD.MM.YYYY day.month.year MM/DD/YYYY Month/Day/Year YYYY-MM-DD Year-month-day, according to ISO standard (ISO)* Time format 12 h (AM/PM) 12-hour clock notation 24 h* 24-hour clock notation...
Operating Design Parameters Setting values Explanation Key beeps Enables | disables short beeps when you touch a button. Off* Alarm beep Enables | disables the sustained beep when a warning message occurs until the warning message has been confirmed. Error beep Enables | disables the sustained beep when an error message occurs until the error message has been confirmed.
Transport Transport Transporting the Device Equipment damage due to improper transport! If the device is lifted where there are loose components, it may fall and be seriously damaged. Never lift the device by the two side covers for transport. Procedure Grip the front of the device under the display and the recess for the power supply at the back of the device and lift carefully.
Startup Startup Equipment Supplied The equipment supplied includes the following components: Qty. Qty. ® ® (arium (arium mini) mini plus) ® arium mini plus as a bench system – arium ® mini as a bench system – Feed water tubing (PE, 4" outer diameter, 2.40 m in length) with reduction connector a“...
Equipment damage due to operation with third-party equipment! The use of third-party power supplies not authorized by Sartorius may cause damage to the device. Use only the original Sartorius power supply.
Page 30
Procedure Check whether the plug design of the power cord is equivalent to your country’s standard. If required: Call the Sartorius Service or your dealer. Connect the power supply to the connection labeled “Power” on the rear of the device.
Confirm the selection with the [l] button. The dialog box “Startup” appears. Startup The demo mode of the device can be started by tapping the [ DEMO ] button. Access to demo mode is password-protected and only accessible to Sartorius employees. DEMO 6.4.2 Starting Startup Mode Procedure Tap the [w] button.
Startup 6.4.3 Setting the Date and Time Procedure 01.07. the [t] button. > Select the desired date format: − DD.MM.YYYY − MM/DD/YYYY − YYYY-MM-DD (ISO) Confirm the selection with the [l] button. > Enter the date. Confirm the selection with the [l] button. >...
Startup 6.4.4 Setting Displayed Values Procedure Tap the [t] button. > Set the desired format for the water quality: − μS/cm c (compensated to 25°C) − MOcm c (compensated to 25°C) Set the desired format for the temperature: °C − °C °F −...
Page 34
Startup Equipment damage due to impurities! If dirt or foreign objects enter the ultrapure water circulation, individual device components can potentially become clogged or wear out faster. If possible, do not touch the cartridge connections. This will prevent impurities from getting into the device. Remove the plugs from the two external cartridge connections.
Startup If required: Repeat the process with the second cartridge. Confirm the insertion of the second cartridge with the [ OK ] button. The dialog box “Put the cover back on” appears. Put the front cover on the device. Put the cover back on Confirm that the cover has been put on with the [ OK ] button.
Page 36
Startup Damage to the bag! The bag may be damaged if forced into the guide rails. Use only slight pressure when inserting the bag into the guide rail. To make it easier to reach the lower connections on the device: Pull the tray up and remove it from the device.
Page 37
Startup Reinsert the tray. Make sure that the bag does not get caught between the tray and the housing. Reattach the left cover of the device. Confirm the insertion of the bag with the [ OK ] button. The dialog box “Connect tubing” appears. Connect tubing ®...
Startup 6.4.7 Connecting Tubing on Device Rear Contamination of tank filling tubing! If the tank filling tubing of the arium ® mini is not stored properly, it may become contaminated. Sufficient quality of the pure water the bag is filled with is no longer guaranteed in this case.
Startup 6.4.7.2 Connecting Drain Water Tubing (Drain) Procedure Connect the drain water tubing to the connector labeled “Drain” on the rear of the device. Water leakage! Water can escape from the drain water tubing during operation. Attach the drain water tubing to the outlet. Guide the free end of the drain water tubing to the drain and attach it.
Startup 6.4.7.3 Connecting the Tank Outlet Tubing (Bag Outlet) Procedure Connect the long tank outlet tubing to the connector labeled “Bag Outlet” on the rear of the device. Close the ball cock. Confirm that the tubing has been connected with the [ OK ] button. The prompt to open and close the ball cock is shown later in the wizard.
Startup 6.4.8 Rinsing the Ultrapure Water Cartridge During rinsing, the device fills and rinses the ultrapure water cartridge. Air is removed from the ultrapure water circulation in the process. If the bag no longer contains a sufficient amount of water or empties during the rinsing process, the bag is filled automatically.
Page 42
Startup Close the tank outlet (Bag Outlet) on the rear side of the device. Close bag outlet After the rinsing process starts, the remaining rinse time is displayed in minutes. In order to cancel the rinsing process, e.g. if a sufficiently large vessel is Rinsing unavailable: Tap the [>] button.
Startup 6.4.9 Connecting the Final Filter Procedure Attach the bell assembly to the final filter. Press the final filter into the quick connector of the water outlet. Confirm that the final filter has been connected with the [ OK ] button. The display changes to the dispensing screen.
Confirm the selection with the [l] button. The dialog box “Startup” appears. Startup The demo mode of the device can be started by pressing the [DEMO] button. Access to demo mode is password-protected and only accessible to Sartorius employees. DEMO 6.5.2 Starting Startup Mode Procedure Tap the [w] button.
Startup 6.5.3 Setting the Date and Time Procedure Tap the [t] button. > Select the desired date format: − DD.MM.YYYY − MM/DD/YYYY − YYYY-MM-DD (ISO) Confirm the selection with the [l] button. > Enter the date. Confirm the selection with the [l] button. >...
Startup 6.5.4 Setting Displayed Values Procedure Tap the [t] button. > Set the desired format for the water quality: − μS/cm c (compensated to 25°C) − MOcm c (compensated to 25°C) Set the desired format for the temperature: °C − °C °F −...
Page 47
Startup Equipment damage due to impurities! If dirt or foreign objects enter the ultrapure water circulation, individual device components can potentially become clogged or wear out faster. If possible, do not touch the cartridge connections. This will prevent impurities from getting into the device. Remove the plugs from the two external cartridge connections.
Startup The dialog box “Put the cover back on” appears. Put the front cover on the device. Put the cover back on Confirm that the cover has been put on with the [OK] button. The dialog box “Insert bag” appears. Insert bag 6.5.6 Inserting Bags...
Page 49
Startup Damage to the bag! The bag may be damaged if forced into the guide rails. Use only slight pressure when inserting the bag into the guide rail. To make it easier to reach the lower connections on the device: Pull the tray up and remove it from the device.
Page 50
Startup Reinsert the tray. Make sure that the bag does not get caught between the tray and the housing. Reattach the left cover of the device. Confirm the insertion of the bag with the [OK] button. The dialog box “Connect tubing” appears. Connect tubing ®...
Startup 6.5.7 Connecting Tubing on Device Rear Contamination of tank filling tubing! If the tank filling tubing of the arium ® mini is not stored properly, it may become contaminated. Sufficient quality of the pure water the bag is filled with is no longer guaranteed in this case.
Startup 6.5.7.2 Connecting Drain Water Tubing (Drain) Procedure Connect the drain water tubing to the connector labeled “Drain” on the rear of the device. Water leakage! Water can escape from the drain water tubing during operation. Connect the drain water tubing to the outlet. Guide the free end of the drain water tubing to the drain and attach it.
Startup The dialog box “Fill bag” appears. Change the bag (see Chapter 6.5.8, page 52). Fill bag Rinse ultrapure water cartridge (see Chapter 6.5.8, page 52). 6.5.8 Fill bag Procedure Prepare a container with a sufficient amount of pretreated water (at least 5 liters).
Page 54
Startup ® If the bag in the arium mini no longer contains a sufficient amount of water or empties during the rinsing process, the dialog box “Fill bag” appears. Fill bag Change the bag (see Chapter 6.5.8, page 52). Start the rinsing process again. The device continues the previous rinsing process.
Startup Requirements The filling level is too low for the dispensing volume entered during volume- controlled dispensing. The dialog box “Fill bag?” appears. Fill bag? Procedure Tap the [ YES ] button. If the bag in the no longer contains a sufficient amount of water or empties during the rinsing process, the dialog box “Fill bag”...
Startup 6.5.11 Rinsing the Final Filter Requirements The device is now in dispensing mode. Procedure Place a vessel (at least 6 liters) under the final filter. Open the final filter’s vent valve. Remove the protective cap from the bell assembly. Dispense at least 5 liters of water (the equivalent of a full bag).
Operation Operation Turning the Device On and Off When the device is turned off during normal operation, e.g. in the evening or on weekends, consistent ultrapure water quality is no longer guaranteed. Activating or deactivating standby mode (see Chapter 7.3.2, page 60). Procedure To turn the device on: Reconnect the device to the AC power.
Operation To stop dispensing: Tap the [STOP] button. 0.055 μS/cm c 21.3 °C STOP 0.90 l When dispensing is finished: Attach the protective cap to the final filter bell assembly. 7.2.2 Volume-controlled Dispensing of Ultrapure Water During volume-controlled water dispensing, a previously specified amount of water is dispensed.
Operation To cancel dispensing before it is finished: Tap the [m] button. 0.055 21.3 °C STOP Using Favorites The last selected dispensing volume is automatically stored and displayed on the Favorites button on the dispensing screen. To start another dispensing process with the previously selected dispensing volume: Tap the Favorites button.
Operation Operating the Menu Procedure To open the menu: Tap the button in dispensing mode. ® Manually fill the bag (only arium mini) (see Chapter 7.3.1, page 59). 0.055 Activate or deactivate standby mode (see Chapter 7.3.2, page 60). Open “Settings” menu (see Chapter 7.3.3, page 61). 21.3 °C Open “Care”...
Operation Requirements The bag contains water. Bag filling must be initiated using the menu. Procedure Tap the symbol [Fill bag]. The dialog box “Fill bag” (Step 1) appears. Check whether the tank filling tubing is connected to the “Inlet” connection. If required: Connect the tank filling tubing.
Operation ® If the arium mini plus is manually switched into standby mode while the bag is being automatically filled, a prompt for further action appears. The filling of the bag can be canceled immediately, which puts the devices in standby mode. If filling is continued, the device will automatically switch to standby mode.
Operation 7.3.4 Performing Volume Adjustment To dispense the most accurate amount of ultrapure water, the flow rate sensor of the device can be readjusted. For this purpose, a sample volume of approximately one half liter is dispensed. The actual volume of the extracted sample volume is measured and transmitted to the device.
Cleaning and Maintenance Cleaning and Maintenance Cleaning the Display Procedure To avoid uncontrolled changes to the settings of the device: Activate standby mode (see Chapter 7.3.2, page 60). Wipe the display gently with a soft, dry cloth. Deactivate standby mode (see Chapter 7.3.2, page 60). Cleaning the Device Housing Danger of injury caused by electric current! When cleaning the device and its components while they are connected to the...
Cleaning and Maintenance Maintenance Schedule Depending on the volume of water dispensed, it may be necessary to change the consumables more often than specified in the maintenance schedule. If, for example, sterile water is required, the final filter must be replaced regularly. It is advisable to replace different consumables in the same maintenance step.
Cleaning and Maintenance Displaying Reminders Reminders to replace certain consumables are automatically displayed as warning messages (see Chapter 9.2, page 84). Pending consumables replacements can be viewed at a glance. All reminders are automatically updated after the consumable has been replaced. Procedure Tap the menu item [ Reminder ] in the “Care”...
Cleaning and Maintenance 8.6.1 Changing the Bag Requirements In the “Change consumable” menu, the menu item [ Bag ] is activated. The dialog box “Drain bag” appears. Drain bag Procedure Check whether the tank outlet tubing is connected to the “Bag Outlet” connection.
Page 68
Cleaning and Maintenance Damage to the bag! The bag may be damaged if forced into the guide rails. Use only slight pressure when inserting the bag into the guide rail. Using the handle, slide the bag into the guide rail in the upper portion of the device housing.
Cleaning and Maintenance The dialog box “Close bag outlet” appears. Close the ball cock of the tank outlet tubing. Close bag outlet Confirm that the tank outlet has been closed with the [ OK ] button. The arium ® mini plus automatically begins filling the bag. The dialog box “Fill bag?”...
Cleaning and Maintenance 8.6.3 Changing Cartridges To replace cartridges, the final filter must be removed and the device depressurized. Requirements In the “Change consumable” menu, the menu item [ Pretreatment cartridge ], or [ Ultrapure water cartridge ] is activated. The dialog box “Remove filter”...
Cleaning and Maintenance 8.6.3.2 Inserting Cartridges The arium ® mini plus has a pretreatment cartridge and an ultrapure water cartridge. Procedure Equipment damage due to impurities! If dirt or foreign objects enter the ultrapure water circulation, individual device components can potentially become clogged or wear out faster. If possible, do not touch the cartridge connections.
Page 72
Cleaning and Maintenance Requirements In the “Change consumable” menu of the arium ® mini, the menu item [ Ultrapure water cartridge ] is enabled. The dialog box “Change cartridge” appears and asks you to replace the second cartridge. Requirements Cartridge replacement is complete. In the “Change consumable”...
Cleaning and Maintenance 8.6.3.3 Rinsing the Device Requirements The new cartridges have been inserted. Procedure If required: Rinse the pretreatment cartridge of the arium ® mini plus (see arium ® mini plus in Chapter 6.4.7.3, page 39). Prepare rinsing: Attach the sampling tubing (see Chapter 6.4.9, page 42) Rinse ultrapure water cartridge (see Chapter 6.4.9, page 42).
Cleaning and Maintenance 8.6.4.1 Removing Cartridges Procedure Press the two protruding terminals of the cartridge together and pull the cartridge forward and out. Confirm removal of the cartridge with the [ OK ] button. If the menu item [ UV lamp ] is activated in the “Change consumable” menu, the dialog box “Change UV lamp”...
Page 75
Cleaning and Maintenance Press the cartridge firmly into the cartridge holder until it locks into the guide rails with a distinct clicking sound. Check that the cartridge has been installed securely by turning it slightly. Confirm the insertion of the cartridge with the [ OK ] button. Requirements mini, the menu item [ Ultrapure ®...
Cleaning and Maintenance 8.6.4.3 Rinsing the Device Requirements The new cartridges have been inserted. Procedure If required: Rinse the pretreatment cartridge of the arium ® mini plus (see Chapter 6.4.7.3, page 39). Prepare rinsing: Connect the dispense tube (see Chapter 6.4.9, page 42) Rinsing the Ultrapure Water Cartridge: Start rinsing (see Chapter 6.4.9, page 42).
Cleaning and Maintenance Requirements The filling level is too low for the dispensing volume entered during volume- Fill bag? controlled dispensing. The dialog box “Fill bag?” appears. Procedure Tap the [YES] button. 8.6.5 Replacing the UV Lamp If a UV lamp is installed in the device, the UV lamp is located behind the ®...
Cleaning and Maintenance 8.6.5.1 Removing the UV Lamp Procedure Disconnect the device from the AC power. The display goes out. After the power has been restored, the wizard will continue automatically. Risk of injury from electrical current and UV radiation! The UV lamp emits UV radiation and may be live.
Cleaning and Maintenance 8.6.5.2 Inserting the UV Lamp Touching UV lamps with your bare fingers will cause them to become defective! Touching the UV lamp with your bare fingers will leave fingerprints. The fingerprints can become so hot during operation that the UV lamp is destroyed. Never touch the glass of the UV lamp with your bare fingers.
Cleaning and Maintenance If no sterile water is needed: − Remove final filter (see Chapter 8.6.2, page 68). − To deactivate reminders about replacing the final filter: In the dialog box “Reminder Endfilter”, tap the [ lnactive ] button. Carrying Out Depressurization The device is under pressure during operation.
Cleaning and Maintenance Carrying Out Venting During venting, the device fills and rinses the ultrapure water cartridge. Air is removed from the ultrapure water circulation in the process. The ultrapure water circulation must be rinsed under the following conditions: − when the water quality displayed during operation fluctuates constantly −...
Improper repairs to the device pose risks to the user and affect the proper operation of the device. Disconnect the device from power. Ensure that all repairs are performed only by authorized and specialized personnel. If required: Call Sartorius Service or a Sartorius dealer. ® ® Operating Instructions arium mini plus | arium...
Reconnect the device to the AC power. If required: Call the Sartorius Service. Error 0140 The UV lamp is The UV lamp is not Check whether the black plug of the UV 8.6.5, page 76...
Page 84
7.2.4, page 58 to “Bag Outlet”, let about 4 of the pure defective. water out of the bag and check the level change. If required: Call the Sartorius Service. Error The conductivity There is air in the Initiate venting. 8.9, page 80...
RO water is out of range. determined. The pretreatment Replace the pretreatment cartridge. 8.6.3, page 63 cartridge has been If required: Call the Sartorius Service. used up. Conductivity The conductivity There is air in the Initiate venting. 8.9, page 80...
Sartorius Maintenance The maintenance Call the Sartorius Service. maintenance is service must be service interval has required. performed. expired. Additional Faults Fault Cause Correction...
Disposal 10 Disposal 10.1 Disposing of Packaging The packaging is made from environmental-friendly materials that can be used as secondary raw materials. Procedure Dispose of the packaging in accordance with local government regulations. 10.2 Information on Decontamination The device does not contain any hazardous materials that would necessitate special disposal measures.
The crossed-out waste bin symbol indicates that separate collection is required. In Germany and several other countries, Sartorius itself assumes responsibility for the return and legally compliant disposal of its electronic and electrical products. These products may not be placed with household waste or brought to collection centers run by local public disposal operations –...
Country-specific, 3-pin, two-sided plug Connector according to DIN EN/IEC 60320-1/C14 3-pin Other data See power supply label Device Power supply Only via Sartorius power supply 1000018304 Input supply voltage +24 (±10%) Current consumption, max. 11.2 Ambient Conditions Unit Value Environment...
Technical Specifications 11.3 Safety of Electrical Equipment Unit Value Safety requirements according to Safety requirements for electrical equipment for DIN EN/IEC 61010-1 measurement, control and laboratory use – Part 1: General requirements 11.4 Electromagnetic Compatibility Unit Value EMC requirements according to DIN EN 61326-1 Electrical equipment for measurement, control and laboratory use –...
Technical Specifications ® arium mini Unit Value Water type Ultrapure water ASTM Type 1 Production output – < 1,0 Water dispensing flow rate L/min Volume-controlled dispensing Between 0.05 and 5 (in 50-ml increments) Typical conductivity μS/cm 0.055 (compensated to 25 °C Typical resistance MO ×...
Technical Specifications ® arium mini Purified water using reverse osmosis, distillation or deionization Unit Value Inlet pressure Depressurized Temperature °C 2 to 30 Specific conductivity μS/cm < 100 (compensated to 25 °C) < 50 Turbidity < 1 pH value 4 to 10 11.7 Device Properties Unit Value...
Master Contract: 167555 Project: Date Issued: 2016-01-15 70055155 Issued to: Sartorius Lab Instruments GmbH & Co. KG 94-108 Weender Landstrasse Goettingen, 37075 GERMANY The products listed below are eligible to bear the CSA Mark shown with adjacent indicators 'C' and 'US' for Canada and US or with adjacent indicator 'US' for US only or without either indicator for Canada only.
Page 96
Project: 70055155 Date Issued: 2016-01-15 Rating: 24 Vdc; 3,0 A (max.). For use with the power supply: FSP, Model FSP150-AAAN2 (Sartorius model 1000018304): Input: 100-240Vac, 50-60Hz, 2A, Output: 24Vdc, 6.25A. Notes: 1. The above model has been evaluated for use in a Pollution Degree 2, Installation Category II.
Page 97
Sartorius Lab Instruments GmbH & Co. KG Otto-Brenner-Strasse 20 37079 Goettingen, Germany Phone: +49.551.308.0 Fax: +49.551.308.3289 www.sartorius.com The information and figures contained in these instructions correspond to the version date specified below. Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.