ECG Leads
Monitored
transceiver usage. If the transceiver senses the insertion of a leadset not
matching the current configuration, the transceiver will source the ECG per the
leadset detected and issue and auditory signal to indicate the condition. An
INVALID LEADSET INOP is generated at the Information Center.
There is no cardiotach within the transceiver; cardiotach analysis resides in the
arrhythmia algorithm at the Information Center. Therefore, arrhythmia analysis
is always turned on for telemetry patients. Arrhythmia analysis is either Basic or
Enhanced, depending on the product configuration.
Depending on the leadset connected to the transceiver, a different set of
viewable leads are available at the Information Center. The transceiver can
source up to four raw ECG waves.
If you are using ...
3-wire
5-wire (Standard mode)
these leads can be selected at the
Information Center
I, II, III
Sourced (raw) waves are received as:
Channel 1 = I, II, or III
Default is II.
I, II, III, aVR, aVL, aVF, MCL and V
Sourced (raw) waves are received as:
Channel 1 = II
Channel 2 = III
Channel 3 = MCL
Defaults are II, V, III.
Measuring ECG
6-5
ECG Monitoring