Alarm Text
NO SIGNAL
REPLACE BATTERY
REPLACE BATTERY T
This INOP may also
Note—
be configured to display as
a Red or Yellow Technical
Alarm.
SOME ECG ALRMS OFF
SpO
T EQUIP MALF
2
SpO
T ERRATIC
2
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
Hard,
Dead battery. No
Latched
monitoring is occurring.
Soft
Some yellow arrhythmia
alarms have been turned off
for this patient.
Hard
Malfunction in the SpO
equipment
Hard
Erratic SpO
measurements, often due to
a faulty sensor or invalid
SpO
2
incorrect transducer
position
Patient is out of
range, or
No batteries in
transceiver, or
Transceiver has
failed, or
Connection to
Database Server has
failed.
2
2
measurements, or
Technical Alarms (INOPs)
What to do
Make sure that the
transceiver is in range
and has good batteries.
Replace the transceiver
if Power On Self Test
fails.
Put bed in Standby.
Contact Service.
Replace batteries.
For information only.
Call Service.
Repeat measurement,
reposition sensor on patient,
or finally, replace sensor.
Alarms
5-11