St Alarm Settings - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Establishing
ST Reference
Beats
(Baseline)
ST Alarm
Settings
Step
Action
3
Adjust the J point, if necessary, by placing the cursor over the J-
point button to access the adjustment arrows. Then use the arrows to
position the bar at the end of the QRS complex and the beginning of
the ST segment.
4
Adjust the ST point, if necessary, by using the J point as an
"anchor" and placing the bar at the midpoint of the ST segment.
Choices are J+0, J+20, J+40, J+60, or J+80.
After adjusting the measurement points, you can establish baseline reference
beats for all available leads in the ST Review window at the IntelliVue
Information Center. Reference beats enable you to compare waveform changes,
for example from admission, or prior to or after treatment. The reference
continues to be saved beyond the 24 hour review window, but you can update it
to any beat within the last 24 hours. Please refer to the IntelliVue Information
Center Instructions for Use or Online Help for directions.
All IntelliVue Information Center alarm settings (limits and on/off status) have
unit default settings. The IntelliVue Information Center however, lets you set the
high and low ST alarm limits for individual patients based on:
Your assessment of the patient's clinical condition.
Unit protocols.
R
P
Q
S
J point
R
P
Q
S
ST point
ST/AR Arrhythmia & ST Segment Monitoring
ST/AR ST Segment Algorithm
T
T
7-11