Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 Test
Level
3 Vrms
150 kHz to 80 MHz
6 Vrms
Amateur Radio
& ISM Bands
between 150 kHz
and 80 MHz
10 V/m
80 MHz to 2.7 GHz
Special
Telecommunication
Bands between
300 MHz and 5.6
GHz
Compliance Level
3 Vrms
150 kHz to 80
MHz
6 Vrms
Amateur Radio
& ISM Bands
between 150 kHz
and 80 MHz
10 V/m
Up to 28 V/m
10. Additional Notes
Electromagnetic
Environment - Guidance
Portable and mobile RF
communications equipment
should be used no closer
to any part of the device,
including cables, than
the recommended 30 cm
separation distance.
Interference may occur in
the vicinity of equipment
marked with the following
symbol: