Figure 16 Press Block Over the Manifold
3. Lock the connector block in place by moving the handle to the left as shown in Figure 17.
Figure 17 Close the Handle
WARNING Inspect the seating of the connector block to the manifold. Failure to achieve a proper
connection could result in an endoscope that is not disinfected and therefore should not be used
on a patient.
65

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ADVANTAGE PLUS and is the answer not in the manual?

Questions and answers

Ed
March 7, 2025

How do you reprint the last disinfection cycle?

1 comments:
Mr. Anderson
March 7, 2025

To reprint the last disinfection cycle log on the MEDIVATORS® ADVANTAGE PLUS® system, follow these steps:

1. Double-click the printer icon in the system tray to redisplay the current cycle log.
2. If needed, access all completed cycle logs through the Management program:
- Minimize all running programs to display the Windows desktop.
- Double-click the MEDIVATORS ADVANTAGE PLUS System ADV icon to open the folder.
- Double-click the Management icon.
- Log in to the Management system.
- Click on the "General" tab in the Management screen sidebar.
- Click on the "Cycles" icon.

This allows viewing and reprinting of previous cycle logs as required.