SYSTEM OPERATIONS - continued
End of Day Shutdown
1.
Power off device: Depress and hold the power button. May take up to 5 seconds to Power OFF.
2.
Discard all 24-Hour Multi-Use Tubing.
3.
Wipe outside surfaces and Temperature Probe with an EPA-approved surface disinfectant.
Data Logs
From Home Screen, arrow down, select Data Logs:
1. Select View Logs to view logs; arrow to select the desired log.
2. Select Export Logs to copy all data from the device to an USB drive. No data is deleted from the device, unless 'Remove DB after Export' is checked.
Insert USB drive, press USB Export icon, Enter SCOPE BUDDY™ PLUS Endoscope Flushing Aid Settings password,
then select which USB to export data to.
The screen will display export options. Check the desired data boxes. Press `OK'.
When the export reaches 100%, the `Select Export Destination' screen will return.
Remove USB drive
Import data into SCOPE BUDDY PLUS Endoscope Flushing Aid Dashboard to generate reports.
MAINTENANCE
Daily Maintenance
1. Install new Pump Tube, and Auxiliary Detergent Out Tube (if using clean container setup).
2. Inspect all tubing for leaks and damage to the plastic fittings. Discard tubing and replace if damaged.
3. Perform Dosing Verification. If Dosing Verfication fails, perform Dosing Calibration and Re-Verification.
4. Install new 24-Hour Multi-Use Hookup for Endoscope, If using DSD Hookups, install a new 24-Hour Multi-Use DSD Hookup Adapter.
5. At end of day, discard all 24-Hour Multi-Use Tubing.
6. Wipe down device and Temperature Probe with EPA-approved surface disinfectant.
7. Power off device by holding down Power Button for 5-7 seconds.
TROUBLESHOOTING
Troubleshooting
1. Review Troubleshooting Section in SCOPE BUDDY PLUS Endoscope Flushing Aid User Manual.
2. Contact Medivators Support 1-800-444-4729 to order supplies or for technical assistance.
Medivators Website for Logs and Product Bulletins
Medivators Customer and Technical Support Toll Free: Phone: 1-800-444-4729
SCOPE BUDDY™, INTERCEPT™ and MEDICLEAN EZ™ are trademarks of Medivators Inc.
OLYMPUS
®
is a registered trademark of Olympus Corporation.
PENTAX
®
is a registered trademark of Hoya Corporation.
FUJIFILM™ is a trademark of FUJIFILM Corporation.
© 2019 Medivators Inc. All rights reserved. 50098-1472-EN Rev B
SCOPE BUDDY™ PLUS
Endoscope Flushing Aid |
FOR FACILITY USE ONLY
www.medivators.com
Staff Name:
Training Date:
Trainer:
In-Service Guide

Advertisement

loading
Need help?

Need help?

Do you have a question about the SCOPE BUDDY PLUS and is the answer not in the manual?

Questions and answers

Summary of Contents for Medivators SCOPE BUDDY PLUS

  • Page 1 When the export reaches 100%, the `Select Export Destination' screen will return. • Remove USB drive • Import data into SCOPE BUDDY PLUS Endoscope Flushing Aid Dashboard to generate reports. MAINTENANCE Daily Maintenance 1. Install new Pump Tube, and Auxiliary Detergent Out Tube (if using clean container setup).
  • Page 2 Dosing: Ensure sink and clean container are filled with correct volume and dose detergent Wall Mounting: SCOPE BUDDY PLUS Endoscope Flushing Aid located no higher than 0.6 meters (2 feet) above the top of fluid solution in sink Brushing: If brushing step is turned ON, set desired time and press PLAY.