Hitachi 902 Service Manual page 59

Automatic analyzer
Table of Contents

Advertisement

No.
Data Alarm
Internal standard
concentration
abnormal
Calculation test
error
Overflow
Calculation
disabled
Expected value
over
Printer
S. I/F
I. STD
D
(ISE)
The internal standard
concentration (C(IS)) is within
the following range.
Na : C(IS) < 120.0mEq/L or
K : C(IS) < 3.0mEq/L or
CL : C(IS) < 80.0mEq/L or
%
CALC?
Any data alarm other than
shown below is indicated for
the test to be used for
calculation.
Calculation disabled, test-to-
test compensation disabled,
expected value over
O
OVER
Concentration value (or activity
value) cannot be output within
the specified range of digit
count.
•In process of calculation, the
X
???
denominator becomes zero.
•An overflow occurs in
logarithmic or exponential
calculation.
•In isozyme Q-channel
concentration calculation, the
data alarm of 'calculation
disabled' is indicated for the
isozyme P-channel data or
the isozyme P channel is not
measured.
•Calculation for a calculation
test has been attempted with
the data having a data
blanking alarm.
The test result is outside the
expected value range (outside
mean value ± 2SD in case of a
control sample).
•Smaller than the lower limit
value
L
•Larger than the upper limit
H
value
Description
160.0mEq/L < C(IS)
7.0mEq/L < C(IS)
120.0mEq/L < C(IS)
2 - 47
(cont
Remedy
•Confirm monthly flow
path washing.
•Replace the diluent and
internal standard
solution.
•Check the data alarm
name and take a proper
measure.
•The data is left blank.
•Prepare appropriate
standard solution.
•The data is left blank.
•Determine the cause of
failure to color
development such as
improper kind of
standard solution, wrong
set position and
clogging of sample
probe.
•Dilute the sample or
reduce its volume and
perform analysis again.
•This alarm is not
indicated for serum
index.
•Correct setting on the
parameter registration
menu 1 screen.
d)
'