Operatiton (Sample Disk); Definition Of Control Variable; Rotation Control System - Hitachi 902 Service Manual

Automatic analyzer
Table of Contents

Advertisement

Operatiton (sample disk)

Outer track :
................................ . .......
Inner track : 25 positions ............... For control serum, standard solution, detergent (W1, W2, W3)
(1)

Definition of Control Variable

Represented by optional position POS # (x, y) on the S. disk.
x:
Track variable (1: outer track, 2: inner track)
y:
Position variable (1 to 35: outer track, 1 to 25: inner track)
(2)

Rotation Control System

In rotation from POS #1 (x, y) to POS #2 (x', y'), the sample disk stops at position POS2-1
(x'-1, y'-1), is fed 1 pitch (back) and stops the next position (sampling position).
(a)
x = x
'
(b)
x = x'
θ is an angle between the current position and one position before the
NOTE:
destination.
For rotation and stop in one cycle, refer to the following page.
(sample disk stop position at reset)
35 positions For routine and stat samples
(position-free except detergent)
θ (back) ≥ θ (fore)
fore :
θ (back) < θ (fore)
back :
The sample disk rotates in order of POS #N (x, y) → POS #N (n, m) (nearby
intermediary position in direction determined in (1)) → POS #2 (x'-1, y'-1) →
POS #2 (x', y').
The direction of rotation is determined in the same manner as in (1). Note
that the sample disk moves back when POS #N (n, m) is ahead of POS #2-
1 (x'-1, y'-1) in the above direction of rotation.
Sampling position
Crosslink (5 points)
7 - 32
S. barcode

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents