Two-Stage I.v. Pole Assembly Patient Left - 6500-211-000; Three-Stage I.v. Pole Assembly Patient Left - 6500-216-000 - Stryker Power-PRO XT 6500 Operation & Maintenance Manual

Ambulance cot
Hide thumbs Also See for Power-PRO XT 6500:
Table of Contents

Advertisement

Two-Stage I.V. Pole assembly Patient Left - 6500-211-000

Three-Stage I.V. Pole assembly Patient Left - 6500-216-000

0004-592-000
Optional 2-Stage I.V. Pole assembly, Left - 6500-211-000 (for Reference Only)
Item
A
B
Optional 3-Stage I.V Pole assembly, Left - 6500-216-000 (for Reference Only)
Item
A
B
www.stryker.com
Part No.
Part Name
6100-115-060
Clip, I.V. Pole
6500-001-042
I.V. Pole Assembly, 2-stage, Left
(pg. 112)
Part No.
Part Name
6100-115-060
Clip, I.V. Pole
6500-001-044
I.V. Pole Assembly, 3-stage, Left
(pg. 113)
6500-309-001 REV A
qty.
1
1
qty.
1
1
Return To Table of Contents
111

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Power-PRO XT 6500 and is the answer not in the manual?

Questions and answers

Iván Garcia
January 24, 2025

Como inyectarle aceite hidráulico a la camilla? Mi camilla Stryker 6500, al parecer necesita cambio de aceite. Últimamente no sube ni baja presionando + o - . Lo hace muy esporádicamente. De ahí deduZco UE le falta aceite hidráulico.

1 comments:
Mr. Anderson
February 10, 2025

To inject hydraulic oil into the Stryker Power-PRO XT 6500 stretcher, follow these steps:

1. Raise the cot to the full up position.
2. Ensure the fill port is horizontal and aligned with the hole in the motor mount.
3. Remove the port plug using a 3/16" Allen wrench.
4. Fill the reservoir up to the bottom of the fill port using Mobil Mercon® V Synthetic Blend Oil (6500-001-293).
5. Replace the plug and run the cot up and down a few times to circulate the oil.

This ensures proper hydraulic function.

This answer is automatically generated