GE Moisture monitor series 3 Programming Manual page 70

Panametrics hygrometer
Hide thumbs Also See for Moisture monitor series 3:
Table of Contents

Advertisement

March 2008
Entering Moisture Curve
Data (cont.)
3-42
#
DP/°C
MH
12
+20.0
0.0000
#
DP/°C
MH
12
+20.0
1.046
Repeat the last three steps to finish entering all the information for
each data point. Enter the next data point number you want to enter
over the present data point number and press the right arrow key.
When you do this, the DP/°C and MH values will display as the last
values entered. This indicates you are ready to begin entering
information for the selected data point.
Note: To review the moisture curve data, enter the number of the
data point and press
You may now do one of the following:
Enter data for a different channel or type of sensor by pressing the
[NO]
key until you reach the Programming Menu, and then refer to
Entering Calibration Data for New Probes/Sensors on page 3-39.
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
[NO]
Exit by pressing the
IMPORTANT:
You must configure a channel AND enter the
necessary calibration data in order for a sensor to
work properly.
1
Enter the corresponding dew
[YES]
point. Press
right arrow key.
Enter the corresponding MH
1
[YES]
value. Press
right arrow key.
[YES]
.
[RUN]
key followed by the
Advanced Programming
and press the
and press the
key.