Operation; Selection Of Exposure Mode; Selection Of Exposure Time; Activating And Deactivating The Light - 3M elipar freelight Operating Instructions Manual

Cordless polymerization unit
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Operation

Selection of Exposure Mode

LED ON =
"Exponential"
LED OFF =
"Standard"
Press the button in the light-blue field to select one
of the two exposure mode options. During exposure
the exposure mode selection button is inactive.
- In "Exponential" exposure mode, the LED in
the light blue field is ON and the exposure time
switches to 40 sec - 4 LED's in the dark-blue
field are ON. Switching from "Exponential" to
"Standard" exposure mode, the unit automatically
switches the exposure time back to the previous
setting in "Standard" mode.
- In "Standard" exposure mode, the LED in the
light blue field is OFF. Several exposure time
options are available in this exposure mode.
Please refer to "Selection of Exposure Time".

Selection of Exposure Time

Exposure
time display
40 sec
30 sec
20 sec
10 sec
In "Exponential" exposure mode the exposure time is
automatically set to 40 sec; no other times are available
as options. Exposure times of 10, 20, 30, and 40 sec
are available options in "Standard" exposure mode.
The adequate exposure time of the product at hand
is stated in the product's Instructions for Use.
Select the exposure time by pressing the button in
the dark-blue field.
- The selected exposure time is indicated by the
- Each time the button is pressed, the setting
- While exposure is ongoing, the button for selection
Button for exposure

Activating and Deactivating the Light

mode selection
Start/Stop button
Activate the light by briefly pressing the Start button.
- Initially, the LEDs show the preset exposure time:
Button for exposure
time selection
If it is desired to turn off the light before the preset
exposure time has completely elapsed, this can be
done by again pressing the Start button.
Full Light Intensity from the Start in
"Exponential" Mode
If it is desired to have the full light intensity available
in "Exponential" Mode for the entire 40 sec of
exposure time of a single exposure, press the Start
button of the handpiece for approx. 2 sec until a
signal is emitted.This indicates that the "Exponential"
Mode is turned off for the present exposure.
4 green LEDs.
advances to the next (higher) value; from 40 sec
the exposure time setting returns to 10 sec.
Keeping the button depressed the display scrolls
through the available settings.
of the exposure time is deactivated.
4 illuminated LEDs represent 40 sec of exposure
time. After every 10 sec of exposure, one of the
LEDs is turned off, i.e. 3 LEDs correspond to
30 sec of exposure time remaining, 2 LEDs to
20 sec of exposure time remaining, etc.
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents