4.2.3 Setting the Alarm Range Error Response

A Range Error can occur when a measured value is within the capacity of the analyzer, but it exceeds the
calibration range of the probe. They can be either Under Range or Over Range errors.
The moisture.IQ indicates Range Errors with an Over Range or Under Range message.The error condition
applies to all measurements of the same mode. For example, if a dew point measurement is Over Range , then
moisture in ppMv will also be Over Range. If multiple Range Errors occur simultaneously, the moisture.IQ
responds to them in the following order: Oxygen Errors , Moisture Errors , Temperature Errors and then Pressure
Errors.
Complete the following steps to configure the range error response of the selected alarm:
1. Tap the Channel (1-6) and Alarm ( A or B ), and then tap the Range Error button to open a screen similar
to Figure 49 below.
2. Set the alarm to either Trip on Under Range errors or to Ignore Under Range errors ( Ignore is the default
setting).
3. Set the alarm to either Trip on Over Range errors or to Ignore Over Range errors ( Ignore is the default
setting).
4. Tap the Save button to keep the new settings or tap the Cancel button to keep the previous settings.
moisture.IQ User's Manual
Figure 49: Range Error Setup Screen - Alarm Response
Chapter 4. Using the Outputs, Alarms and Logger Menus
55