Chapter 4. Using the Outputs, Alarms and Logger Menus

4.1.4 Setting the Output Range Error Response

A Range Error can occur when a measured value is within the capacity of the analyzer, but it exceeds the
calibration range of the probe. They can be either Under Range or Over Range errors.
The moisture.IQ indicates Range Errors with an Over Range or Under Range message.The error condition
applies to all measurements of the same mode. For example, if a dew point measurement is Over Range , then
moisture in ppMv will also be Over Range. If multiple Range Errors occur simultaneously, the moisture.IQ
responds to them in the following order: Oxygen Errors , Moisture Errors , Temperature Errors and then Pressure
Errors.
Complete the following steps to configure the range error response of the selected output:
1. Tap the desired Channel (1-6) and Output ( A or B ), and then tap the Range Error button to open a screen
similar to Figure 46 below.
2. Set the output to be driven either High on Under Range errors, Low on Under Range errors or to Ignore
Under Range errors ( Ignore is the default setting).
3. Set the output to be driven either High on Over Range errors, Low on Over Range errors or to Ignore
Over Range errors ( Ignore is the default setting).
4. Tap the Save button to keep the new settings or tap the Cancel button to keep the previous settings.
52
Figure 46: Range Error Setup Screen - Output Response
moisture.IQ User's Manual