Dreamstation Heated Humidifier Specifications; Heated Tubing Specifications - Philips DreamStation User Manual

Heated humidifier
Hide thumbs Also See for DreamStation:
Table of Contents

Advertisement

DreamStation Heated Humidifier Specifications
Environmental
Operating Temperature: 5° to 35° C (41° to 95° F)
Storage Temperature: -20° to 60° C (-4° to 140° F)
Relative Humidity (operating & storage): 15 to 95% (non-condensing)
Atmospheric Pressure: 77 to 101 kPa (0 - 2286 m / 0 - 7500 ft)
Physical
Dimensions (humidifier with therapy device): 29.7 x 19.3 x 8.4 cm (11.7" L x 7.6" W x 3.3" H)
Height does not include humidifier air outlet port.
Weight (empty humidifier with therapy device and power supply): Approximately 1.98 kg (4.37 lbs.)
Service Life
The expected service life of the DreamStation Heated Humidifier is 5 years.
Maximum Recommended Pressure
25 cm H
O
2
Water Capacity
325 ml (11 oz.) at recommended water level
Standards Compliance This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Basic Safety and Essential Performance of Medical Electrical Equipment
ISO 8185 General Requirements for Humidification Systems
Electrical (When the heated humidifier is used with a Philips Respironics therapy device)
AC Power Consumption (with 80W power supply): 100 – 240 VAC, 50/60 Hz, 2.0-1.0 A
DC Power Consumption: 12 VDC, 6.67 A
Type of Protection Against Electric Shock: Class II Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water: Drip Proof, IP22
Mode of Operation: Continuous
Electromagnetic Compatibility: The device meets the requirements of EN 60601-1-2.
Heater Plate
Max Temperature: 65° C (149° F)
Pressure Drop with Humidifier
Max: 0.6 cm H
O at 60 LPM flow
2
Humidity
Humidity
Output: 10 mg H
min
Measured @ max flow, 35° C, 15% RH.
Heated Tubing Specifications
Maximum Recommended Pressure
25 cm H
O
2
Inner Diameter
15 mm (0.6 in.)
Length
1.83 m (6 ft.)
Heated Tubing Temperature Range
16° to 30º C (60° to 86° F)
Heated Tubing Temperature Cut-out
≤ 41° C (≤ 106° F)
Material
Flexible plastic and electrical components
Electrical (Heated tubing is powered by the attached heated humidifier)
Refer to "Electrical" section of DreamStation Heated Humidifier Specifications
Environmental
Refer to "Environmental" section of DreamStation Heated Humidifier Specifications
12
O/L
2

Hide quick links:

Advertisement

Table of Contents
loading