Honeywell ST 3000 Installation Manual page 72

Smart transmitter
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

A.5
Configuring Smart Meter Using Pushbuttons,
Selecting Engineering
Units, continued
Table A-5
Selecting Engineering Units, continued
Step
4
Press UNITS SET button to lock in selected
code.
ATTENTION
according to the selections in Step 3, the
meter display will show an error code Er1 for
one second and then return to the previous
engineering units selection.
Goes blank for 1/2 second and returns with
reading in engineering units.
5
If selected engineering unit does not match
one of six unit indicators on meter, peel off
matching stick-on unit label from sheet
(drawing number 30756918-001) and paste it
in lower right hand corner of meter.
6
If you selected Custom or Flow engineering
units, go to Tables A-7 and A-8 to set lower
and upper display limits for smart meter
display.
62
Action
If you select an invalid code
ST 3000 Release 300 Installation Guide
Meter Display
V A R
SEL.
%
0
100
0
1
8
.
ANALOG
In H O
Use stick-on label for engineering units
without indicators on display.
V AR
SEL.
0
%
100
1
0
2
.
ANALOG
Kg/cm
Lower and upper display limits have not been
set for Custom or Flow engineering units.
VA R
SEL.
0
%
U
-
L
ANALOG
Continued
UPPE R
VALUE
UNITS
S E T
LOWER
2
VALU E
Digital reading now
in engineering units
of inches of water
UPPE R
VALUE
UNITS
S E T
LOWER
VALUE
2
Stick-on label
identifies selected
engineering units
UPPE R
VALUE
100
UNITS
S E T
FLOW
LOWER
VALUE
GPM
Continued on next page