Safety Information
Electromagnetic Immunity
The CF-1 is intended for use in the electromagnetic environment specified below.
The user of the CF-1 should assure that it is used in such an environment.
Immunity Test
Electrostatic dis-
charge (ESD)
EN IEC 61000-4-2
Electrical fast tran-
sient/burst
EN IEC 61000-4-4
Surge
EN IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
EN IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level.
T
IEC 60601 Test
Compliance
Level
Level
±(2, 4, 6) kV
±(2, 4, 6) kV
contact
contact
±(2, 4, 8) kV air
±(2, 4, 8) kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV differen-
±1 kV differen-
tial mode
tial mode
±2 kV common
±2 kV common
mode
mode
<5% U
<5% U
T
T
(>95% dip in
(>95% dip in
U
) for 0.5 cycle
U
) for 0.5 cycle
T
T
40% U
40% U
T
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
<5% U
<5% U
T
T
(>95% dip in
(>95% dip in
U
) for 5 sec
U
) for 5 sec
T
T
3 A/m
3 A/m
-8-
Electromagnetic Environment –
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
)
T
ment. If the user of the CF-1 requires
continued operation during power
mains interruptions, it is recommended
that the CF-1 be powered from an
)
T
uninterruptible power supply or a bat-
tery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.